Influence of spectral shaping and tube voltage modulation in ultralow-dose computed tomography of the abdomen

verfasst von: Philipp Feldle, Jan-Peter Grunz, Andreas Steven Kunz, Pauline Pannenbecker, Theresa Sophie Patzer, Svenja Pichlmeier, Stephanie Tina Sauer, Robin Hendel, Süleyman Ergün, Thorsten Alexander Bley, Henner Huflage

Erschienen in: BMC Medical Imaging | Ausgabe 1/2024

Abstract

Purpose

Unenhanced abdominal CT constitutes the diagnostic standard of care in suspected urolithiasis. Aiming to identify potential for radiation dose reduction in this frequent imaging task, this experimental study compares the effect of spectral shaping and tube voltage modulation on image quality.

Methods

Using a third-generation dual-source CT, eight cadaveric specimens were scanned with varying tube voltage settings with and without tin filter application (Sn 150, Sn 100, 120, 100, and 80 kVp) at three dose levels (3 mGy: standard; 1 mGy: low; 0.5 mGy: ultralow). Image quality was assessed quantitatively by calculation of signal-to-noise ratios (SNR) for various tissues (spleen, kidney, trabecular bone, fat) and subjectively by three independent radiologists based on a seven-point rating scale (7 = excellent; 1 = very poor).

Results

Irrespective of dose level, Sn 100 kVp resulted in the highest SNR of all tube voltage settings. In direct comparison to Sn 150 kVp, superior SNR was ascertained for spleen (p ≤ 0.004) and kidney tissue (p ≤ 0.009). In ultralow-dose scans, subjective image quality of Sn 100 kVp (median score 3; interquartile range 3–3) was higher compared with conventional imaging at 120 kVp (2; 2–2), 100 kVp (1; 1–2), and 80 kVp (1; 1–1) (all p < 0.001). Indicated by an intraclass correlation coefficient of 0.945 (95% confidence interval: 0.927–0.960), interrater reliability was excellent.

Conclusions

In abdominal CT with maximised dose reduction, tin prefiltration at 100 kVp allows for superior image quality over Sn 150 kVp and conventional imaging without spectral shaping.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Prevalence of kidney stone disease has increased globally over the last decades, even among children [13]. Its high rate of reoccurrence is well documented, often requiring repeated imaging by computed tomography (CT), which is nowadays considered the diagnostic modality of choice [4, 5]. Although CT outperforms ultrasound with regards to sensitivity and specificity, focus has shifted towards concerns about radiation dose and possible long term side effects for years [69].
While the detection of urolithiasis constitutes a high-contrast imaging task, abdominal imaging in general is challenged by the low-contrast properties of parenchymal organs. For CT angiography, low tube voltage imaging is an established approach to increase contrast due to the photoelectric effect [10]. Notably, some studies suggest a similar advantage at low tube voltages in non-contrast examinations [11]. While the relationship between radiation dose and tube current is known to be linear, i.e., doubling the tube current results in doubling the dose, the relationship between dose and tube voltage is proportional to the square of the voltage. Thus, a tube voltage reduction from 120 to 100 kVp is associated with a 33% reduction in radiation dose, and a reduction to 80 kVp with a decrease by 65% [12]. However, a disadvantage of lower tube voltage lies in the reduced tissue penetration of photons and resulting increase of image noise and higher dependency on patient size [12, 13]. Whether lower tube currents allow for maintaining sufficient image quality particularly depends on the diagnostic task [12, 13].
In the last decade, the application of tin filters has attracted increasing attention as spectral shaping allows for considerable dose savings, particularly for imaging of the lungs und paranasal sinus [1416]. Quantifying the dose reduction potential of spectral shaping remains difficult, however, as the results of the individual studies vary considerably [1719]. Since the use of tin filters is also becoming increasingly common for abdominal CT examinations in recent years [18, 20, 21], a direct comparison between different dose saving techniques is warranted.
As excessive radiation reduction may render an examination diagnostically insufficient, this cadaveric study aimed to determine the most efficient ultralow-dose scan protocol for unenhanced abdominal CT. In order to achieve that, low-kV imaging was compared with tin filtration protocols at two different tube voltages on a third-generation dual-source CT system.

Methods

For this experimental investigation, scans were performed on eight formalin-fixed cadaveric specimens obtained from the local university’s anatomical institute. As body donors had volunteered their corpses for scientific and educational purposes during their lifetime, no further written informed consent was required. Approval for the study was granted by the local ethics committee.

Scan and image parameters

All examinations were performed on a commercially available third-generation dual-source CT scanner (Somatom Force, Siemens Healthcare GmbH, Forchheim, Germany) with an energy-integrating detector system. Specimens were examined in supine position with elevated arms. Scouts were acquired with identical dose settings in anterior-posterior orientation. Each specimen was scanned with five different tube settings, of which two employed spectral shaping with a 0.6 mm tin filter (Sn 100 kVp, Sn 150 kVp). The other settings used tube voltages of 120, 100 and 80 kVp. With each tube setting used for scans at three different dose levels, a total of 15 CT examinations were performed on each specimen. Target volume computed tomography dose indices (CTDIvol) were 3 mGy for standard-dose, 1 mGy for low-dose, and 0.5 mGy for ultralow-dose examinations. Automatic dose modulation was activated for all scans, as is mandatory in clinical routine. Without repositioning of specimens, scan ranges were set with identical length in all 15 consecutive scans. All datasets were reconstructed with a field of view of 350 mm, a slice thickness of 3 mm and a regular soft tissue body kernel (Br36) employing iterative reconstruction at a strength level of 3. (Admire, Siemens Healthcare GmbH). Window settings were selected at 400 (width) / 50 (centre) Hounsfield units, while readers were allowed to adjust according to personal preferences.

Objective image quality

Standardized regions of interest were placed in the spleen, renal cortex, vertebral body and fat tissue at kidney level, each, on three consecutive slices. Mean density in Hounsfield units and standard deviation was measured to calculate a signal-to-noise ratio (SNR) for each organ according to the following formula:
$$ SNR=\frac{mean\, attenuation \left(organ\, tissue\right) }{standard\, deviation \left(organ\, tissue\right)}$$

Subjective image quality

All datasets were independently analysed by three radiologists with at least five years of experience in CT imaging using a commercially available PACS system (Merlin, Phönix-PACS, Freiburg, Germany) and diagnostic monitors (30-inch diameter, RadiForce RX660, EIZO, Hakusan, Japan). For each dataset, readers assessed image quality according to a seven-point scale: 1 = very poor; 2 = poor; 3 = fair; 4 = satisfactory; 5 = good; 6 = very good; 7 = excellent.

Statistics

The statistical analysis was conducted using specialized software (IBM SPSS Statistics, Armonk, USA). Kolmogorov-Smirnov tests were used to examine the normal distribution of cardinal variables. If variables were normally distributed, they are presented as mean ± standard deviation. Nonparametric variables are reported as absolute and relative frequencies, along with median values and interquartile ranges. Friedman tests were employed to compare ordinal-scaled variables, while one-way repeated measures ANOVA was used for comparing continuous data. Pairwise post-hoc tests were conducted with Bonferroni adjustment to account for multiple comparisons. The intraclass correlation coefficient (ICC) was calculated based on absolute agreement of single measures using a two-way random effects model. The ICC values were categorized according to established guidelines [22]. Statistical significance was determined at p-values of less than 0.05.

Results

Radiation dose

With automatic exposure control activated, the standard-dose protocols were associated with a mean CTDIvol of 2.93 mGy ± 0.49 mGy across the performed 40 scans. While the low-dose protocols resulted in a mean CTDIvol of 0.93 mGy ± 0.19 mGy, the ultralow-dose protocols resulted in a mean CTDIvol of 0.47 ± 0.09 mGy. No substantial difference in radiation dose was ascertained for the five individual tube voltage settings on each dose level (all p > 0.05). Table 1 provides a detailed display of the radiation dose administered in each specimen.
Table 1
Radiation dose for each cadaveric specimen with different acquisition protocols employing automatic exposure control
Tube voltage
Sn 100 kVp
Sn 150 kVp
120 kVp
100 kVp
80 kVp
Dose level
Standard
Low
Ultralow
Standard
Low
Ultralow
Standard
Low
Ultralow
Standard
Low
Ultralow
Standard
Low
Ultralow
Specimen 1
3.40
1.12
0.54
3.69
1.25
0.61
3.91
1.27
0.60
3.71
1.26
0.65
3.56
1.34
0.7
Specimen 2
3.00
0.99
0.47
3.28
1.06
0.5
3.51
1.10
0.47
3.39
1.08
0.49
2.79
0.94
0.47
Specimen 3
3.18
1.05
0.47
3.07
1.10
0.53
3.48
1.10
0.51
3.29
1.18
0.53
2.9
1.00
0.48
Specimen 4
2.61
0.80
0.39
2.66
0.76
0.37
2.91
0.94
0.40
2.87
0.90
0.36
2.62
0.77
0.54
Specimen 5
3.12
1.05
0.48
3.35
1.14
0.57
3.38
0.94
0.53
3.37
1.10
0.6
3.40
1.29
0.55
Specimen 6
2.75
0.95
0.41
2.87
0.98
0.48
2.81
0.77
0.47
2.79
0.98
0.48
2.72
1.08
0.53
Specimen 7
2.24
0.76
0.34
1.97
0.72
0.36
2.17
0.64
0.38
2.19
0.72
0.37
2.16
0.73
0.36
Specimen 8
2.40
0.82
0.35
2.37
0.79
0.39
2.34
0.67
0.38
2.39
0.82
0.39
2.44
0.83
0.41
CTDIvol[mGy]
2.84 ± 0.38
0.94 ± 0.13
0.43 ± 0.07
2.91 ± 0.52
0.98 ± 0.18
0.48 ± 0.09
3.06 ± 0.57
0.93 ± 0.21
0.47 ± 0.07
3.00 ± 0.50
1.01 ± 0.17
0.48 ± 0.10
2.82 ± 0.44
1.00 ± 0.21
0.51 ± 0.10
Note.– CTDIvol– volume computed tomography dose index

Objective image quality

While Sn 100 kVp provided the highest SNR within each dose level, SNR decreased with lower applied CTDIvol. In direct comparison to Sn 150 kVp, superior SNR was ascertained for Sn 100 kVp in spleen (p ≤ 0.004) and kidney tissue (p ≤ 0.009) at all three dose levels. In contrast, no significant difference could be ascertained among the tin prefiltration protocols for bone and fat tissue. Detailed SNR comparisons between all protocols in every tissue are provided in Fig. 1 for standard dose, in Fig. 2 for low-dose, and in Fig. 3 for ultralow-dose imaging. A comprehensive overview of SNR values is provided in Table 2.
Table 2
Signal-to-noise ratios were calculated for all scan protocols and tissues. Results are presented as mean values ± standard deviations
Tissue
Tube voltage
Standard-dose protocol
Low-dose protocol
Ultralow-dose protocol
Spleen
Sn 100 kVp
4.74 ± 1.32
3.04 ± 0.92
2.33 ± 0.77
Sn 150 kVp
3.76 ± 1.22
2.54 ± 0.83
1.98 ± 0.55
120 kVp
3.88 ± 1.26
2.68 ± 0.84
1.73 ± 0.52
100 kVp
4.04 ± 1.14
2.54 ± 0.89
1.78 ± 0.60
80 kVp
3.42 ± 1.25
2.25 ± 1.04
1.33 ± 0.47
Kidney
Sn 100 kVp
4.45 ± 1.22
2.79 ± 0.74
2.20 ± 0.82
Sn 150 kVp
3.57 ± 1.13
2.17 ± 0.72
1.66 ± 0.53
120 kVp
3.67 ± 1.35
1.96 ± 0.63
1.57 ± 0.64
100 kVp
3.84 ± 1.43
2.37 ± 0.87
1.47 ± 0.41
80 kVp
3.45 ± 1.30
2.36 ± 1.11
1.22 ± 0.46
Bone
Sn 100 kVp
4.61 ± 2.36
3.88 ± 1.97
3.26 ± 1.49
Sn 150 kVp
3.97 ± 2.24
3.34 ± 1.95
2.85 ± 1.66
120 kVp
4.28 ± 2.18
3.40 ± 1.58
3.03 ± 1.76
100 kVp
4.24 ± 2.04
3.54 ± 1.57
2.91 ± 1.37
80 kVp
4.05 ± 1.83
3.43 ± 1.64
2.98 ± 1.37
Fat
Sn 100 kVp
10.19 ± 6.19
6.44 ± 4.10
5.00 ± 4.41
Sn 150 kVp
8.24 ± 7.70
4.62 ± 2.17
3.51 ± 2.35
120 kVp
8.08 ± 3.74
5.90 ± 4.54
3.61 ± 1.97
100 kVp
9.47 ± 5.91
6.12 ± 4.48
4.21 ± 4.08
80 kVp
9.09 ± 4.96
5.67 ± 3.15
3.45 ± 1.94

Subjective image quality

Higher radiation dose resulted in superior image quality ratings regardless of tube voltage setting. For each dose level, Friedmans rank-based ANOVA revealed significant differences between the five individual scan settings (all p < 0.008). On a standard dose level, pairwise post-hoc analyses demonstrated superior image quality of Sn 100 kVp (median score 7; interquartile range 6–7; p = 0.001) and Sn 150 kVp (6; 6–7; p = 0.019) compared to 80 kVp imaging (6; 6–6). Only Sn 100 kVp was found to be superior to 100 kVp (p = 0.009), however. In ultralow-dose scans, subjective image quality of Sn 100 kVp (3; 3–3) was deemed higher compared to conventional imaging at 120 kVp (2; 2–2), 100 kVp (1; 1–2), and 80 kVp (1; 1–1) (all p < 0.001). Sn 150 kVp (3; 2–3) was considered superior to 100 kVp and 80 kVp imaging (both p < 0.001), albeit equal to 120 kVp (p = 0.071). Among the three tube settings without spectral shaping, no significant difference was ascertained at any dose level (p ≥ 0.156). Detailed results of the subjective image quality analyses are given in Table 3. Figure 4 provides representative CT images for the five different tube settings at an ultralow-dose level of 0.5 mGy. An ICC of 0.945 (95% confidence interval: 0.927–0.960) indicated excellent interrater reliability.
Table 3
Pooled image quality assessment three radiologists. Results are displayed as absolute frequencies and percentages in parentheses. Median values are provided with interquartile ranges
Tube voltage
Sn 100 kVp
Sn 150 kVp
120 kVp
100 kVp
80 kVp
Dose level
Standard
Low
Ultralow
Standard
Low
Ultralow
Standard
Low
Ultralow
Standard
Low
Ultralow
Standard
Low
Ultralow
7
13 (54.2)
0 (0)
0 (0)
9 (37.5)
0 (0)
0 (0)
6 (25.0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
6
11 (45.8)
0 (0)
0 (0)
15 (62.5)
0 (0)
0 (0)
18 (75.0)
0 (0)
0 (0)
21 (87.5)
0 (0)
0 (0)
18 (75.0)
0 (0)
0 (0)
5
0 (0)
20 (83.3)
0 (0)
0 (0)
18 (75.0)
0 (0)
0 (0)
17 (70.8)
0 (0)
3 (12.5)
14 (58.3)
0 (0)
6 (25.0)
9 (37.5)
0 (0)
4
0 (0)
4 (16.7)
5 (20.8)
0 (0)
6 (25.0)
0 (0)
0 (0)
7 (29.2)
0 (0)
0 (0)
10 (41.7)
0 (0)
0 (0)
15 (62.5)
0 (0)
3
0 (0)
0 (0)
19 (79.2)
0 (0)
0 (0)
15 (62.5)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
2
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
9 (37.5)
0 (0)
0 (0)
20 (83.3)
0 (0)
0 (0)
7 (29.2)
0 (0)
0 (0)
6 (25.0)
1
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
4 (16.7)
0 (0)
0 (0)
17 (70.8)
0 (0)
0 (0)
18 (75.0)
Median (IQR)
7 (6–7)
5 (5–5)
3 (3–3)
6 (6–7)
5 (5–5)
3 (2–3)
6 (6–6)
5 (4–5)
2 (2–2)
6 (6–6)
5 (4–5)
1 (1–2)
6 (6–6)
4 (4–5)
1 (1–1)
Note.– IQR– interquartile range

Discussion

This experimental study aimed to evaluate the benefits of spectral shaping versus conventional and low-kV imaging for radiation dose reduction in unenhanced abdominal CT. We were able to ascertain the superiority of tin-filtered scans at 100 kVp over all other voltage settings, with advantages being most pronounced at the ultralow-dose level of 0.5 mGy. Notably, while the radiation dose of all investigated protocols was far below the national diagnostic reference standard for abdominal CT, i.e. 15 mGy [23], the dedicated ultralow-dose protocols realized a thirtyfold decrease, yet still providing sufficient image quality with spectral shaping at 100 kVp.
Although the results of this study concur with other publications that advocate spectral shaping for abdominal low-dose CT imaging [18, 21, 24, 25], protocols without tin filtration are still commonly applied by many radiologists as standard of care [26, 27]. Our findings are generally in line with an experimental study by Karmazyn et al., who reported that merely decreasing the tube voltage is insufficient for adult sized phantoms, especially in ultralow-dose imaging [28]. While this investigation lacked the assessment of additional tin prefiltration, our study included two different voltages settings with spectral shaping. Results indicate that the use of a tin filter at 150 kVp bears no decisive advantage over conventional or low-kV imaging. Therefore, in synopsis with the current literature, we postulate that this setting should rather be considered for obese patients or in cases where artefacts are to be expected due to the presence of metal implants [2931]. Similarly, our findings do not support the superiority of 100 kVp versus 120 kVp, which has been suggested by some authors [27, 32].
On the other side of the spectrum, it is known that lowering the tube voltage leads to a redistribution of the effective radiation exposure with a higher surface dose and increased peripheral absorption of x-ray beams [33]. In an earlier patient study comparing a 120 kVp with a 90 kVp protocol, Nakayama et al. suggested a cut-off value of 70 kg for the latter to prevent a significant drop in SNR [34]. In contrast to similar studies comprising either Sn 150 kVp or Sn 100 kVp settings, our investigation bears the advantage of comprehensive comparisons between all relevant voltage settings at different dose levels for various tissues [24, 35]. Taking into account that none of the cadaveric samples was of obese stature, the reported advantages of Sn 100 kVp protocols regarding SNR and subjective image quality at all dose levels are in line with current literature [25, 27, 36]. As the detection of urinary calculi represents a high-contrast imaging task, suchlike examinations are ideally combined with dedicated low-dose and ultralow-dose scan protocols. In this context, prior studies have suggested dose saving potentials ranging from 30% to as much as 50% with regards to kidney stone imaging and by as much as 90% in chest imaging when combined with automatic tube current modulation and iterative reconstruction algorithms [25, 36, 37].
Some limitations must be considered for the present study: First, the body donors examined were of normal constitution, neither being morbidly obese nor cachectic. Since the selection of cadavers in this study was performed at random, the presence of urinary calculi could not be controlled. Despite this limitation, our study provides valuable insights into potential radiation dose reduction strategies in unenhanced abdominal CT, offering a basis for further investigations with larger patient samples. Second, the minimal applicable tube current with the employed dual-source scanner in a 120 kVp setting is 5 mAs. Although this setting was not selected in the current study, the limited tube current lowering potential could lead to a dose disadvantage in slim patients, while rendering a dose-comparable study at 150 kVp technically impossible. In clinical practice, however, it can be assumed that the automatic tube voltage selection prevents this constellation. Third, the dual-source scanner hardware and tin prefiltration technique are exclusive to one particular vendor. However, other manufacturers also rely on pre-patient beam filtration for hardening of the x-ray spectrum (albeit with different materials), enhancing the relevance and generalizability of our findings to a broader context. Finally, the present study was limited to investigations on energy-integrating detector technology. Novel photon-counting detectors may provide additional advantages when combined with dedicated low-dose protocols for abdominal imaging [35].

Conclusion

Ultralow-dose scans of the abdomen, e.g. for the detection of urinary calculi, benefit from the application of spectral shaping with tin filtration at 100 kVp. Compared with low-kV or conventional imaging, both subjective and quantitative image quality were superior with Sn 100 kVp. In contrast, employing tin filtration at 150 kVp did not provide a similar advantage in the investigated sample of cadaveric specimens with normal body constitution.

Acknowledgements

Not applicable.

Declarations

The anatomical institute of our university received general approval from the local ethics committee to use the cadaveric specimens that are donated to them for study and research purposes. For the present study, additional specific permission was obtained from the institutional review board of the University of Würzburg, Germany (approval number 20210623 01). The need for written informed consent was waived, since body donors had volunteered their corpses for scientific and educational purposes during their lifetime. Medical history and sociodemographic data of body donors were not accessible for our study; therefore all image material was anonymized before analysis. All procedures were carried out in accordance with the ethical standards of the institutional and national research committee and with the 1975 Declaration of Helsinki.
Not applicable.

Competing interests

Three authors (J.-P.G, A.S.K., T.A.B.) have received speaker honoraria from Siemens Healthineers unrelated to the present investigation within the last three years. The other authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
2.
Zurück zum Zitat Chewcharat A, Curhan G. Trends in the prevalence of kidney stones in the United States from 2007 to 2016. Urolithiasis. 2021;49:27–39.CrossRefPubMed Chewcharat A, Curhan G. Trends in the prevalence of kidney stones in the United States from 2007 to 2016. Urolithiasis. 2021;49:27–39.CrossRefPubMed
3.
Zurück zum Zitat Paraboschi I, Gnech M, De Marco EA, Minoli DG, Bebi C, Zanetti SP, et al. Pediatric Urolithiasis: current Surgical strategies and future perspectives. Front Pediatr. 2022;10:1–7.CrossRef Paraboschi I, Gnech M, De Marco EA, Minoli DG, Bebi C, Zanetti SP, et al. Pediatric Urolithiasis: current Surgical strategies and future perspectives. Front Pediatr. 2022;10:1–7.CrossRef
4.
Zurück zum Zitat Benson DA, Maxwell RM, Poeter E, Ibrahim H, Dean A, Revielle J, et al. EAU guidelines on urolithiasis. Eur Assoc Urol. 2019;49:1–88. Benson DA, Maxwell RM, Poeter E, Ibrahim H, Dean A, Revielle J, et al. EAU guidelines on urolithiasis. Eur Assoc Urol. 2019;49:1–88.
5.
Zurück zum Zitat Coursey CA, Casalino DD, Remer EM, Arellano RS, Bishoff JT, Dighe M, et al. ACR appropriateness Criteria® Acute Onset Flank Pain–Suspicion of Stone Disease. Ultrasound Q. 2012;28:227–33.CrossRefPubMed Coursey CA, Casalino DD, Remer EM, Arellano RS, Bishoff JT, Dighe M, et al. ACR appropriateness Criteria® Acute Onset Flank Pain–Suspicion of Stone Disease. Ultrasound Q. 2012;28:227–33.CrossRefPubMed
6.
Zurück zum Zitat Hendee WR, O’Connor MK. Radiation risks of medical imaging: separating fact from fantasy. Radiology. 2012;264:312–21.CrossRefPubMed Hendee WR, O’Connor MK. Radiation risks of medical imaging: separating fact from fantasy. Radiology. 2012;264:312–21.CrossRefPubMed
7.
Zurück zum Zitat Berrington de González A. Projected Cancer risks from computed Tomographic Scans Performed in the United States in 2007. Arch Intern Med. 2009;169:2071.CrossRefPubMed Berrington de González A. Projected Cancer risks from computed Tomographic Scans Performed in the United States in 2007. Arch Intern Med. 2009;169:2071.CrossRefPubMed
8.
Zurück zum Zitat Ulusan S, Koc Z, Tokmak N. Accuracy of sonography for detecting renal stone: comparison with CT. J Clin Ultrasound. 2007;35:256–61.CrossRefPubMed Ulusan S, Koc Z, Tokmak N. Accuracy of sonography for detecting renal stone: comparison with CT. J Clin Ultrasound. 2007;35:256–61.CrossRefPubMed
9.
Zurück zum Zitat Fowler KAB, Locken JA, Duchesne JH, Williamson MR. US for detecting Renal Calculi with Nonenhanced CT as a reference Standard. Radiology. 2002;222:109–13.CrossRefPubMed Fowler KAB, Locken JA, Duchesne JH, Williamson MR. US for detecting Renal Calculi with Nonenhanced CT as a reference Standard. Radiology. 2002;222:109–13.CrossRefPubMed
10.
Zurück zum Zitat Patino M, Prochowski A, Agrawal MD, Simeone FJ, Gupta R, Hahn PF, et al. Material separation using dual-energy CT: current and emerging applications. Radiographics. 2016;36:1087–105.CrossRefPubMed Patino M, Prochowski A, Agrawal MD, Simeone FJ, Gupta R, Hahn PF, et al. Material separation using dual-energy CT: current and emerging applications. Radiographics. 2016;36:1087–105.CrossRefPubMed
11.
Zurück zum Zitat Nagayama Y, Oda S, Nakaura T, Tsuji A, Urata J, Furusawa M, et al. Radiation dose reduction at pediatric CT: use of low tube voltage and iterative reconstruction. Radiographics. 2018;38:1421–40.CrossRefPubMed Nagayama Y, Oda S, Nakaura T, Tsuji A, Urata J, Furusawa M, et al. Radiation dose reduction at pediatric CT: use of low tube voltage and iterative reconstruction. Radiographics. 2018;38:1421–40.CrossRefPubMed
12.
Zurück zum Zitat Seyal AR, Arslanoglu A, Abboud SF, Sahin A, Horowitz JM, Yaghmai V. CT of the abdomen with reduced tube voltage in adults: a practical approach. Radiographics. 2015;35:1922–39.CrossRefPubMed Seyal AR, Arslanoglu A, Abboud SF, Sahin A, Horowitz JM, Yaghmai V. CT of the abdomen with reduced tube voltage in adults: a practical approach. Radiographics. 2015;35:1922–39.CrossRefPubMed
13.
Zurück zum Zitat Modica MJ, Kanal KM, Gunn ML. The obese emergency patient: imaging challenges and solutions. Radiographics. 2011;31:811–24.CrossRefPubMed Modica MJ, Kanal KM, Gunn ML. The obese emergency patient: imaging challenges and solutions. Radiographics. 2011;31:811–24.CrossRefPubMed
14.
Zurück zum Zitat Grunz J-P, Heidenreich JF, Lennartz S, Weighardt JP, Bley TA, Ergün S, et al. Spectral shaping Via Tin Prefiltration in Ultra-high-resolution Photon-counting and energy-integrating detector CT of the temporal bone. Invest Radiol. 2022;57:819–25.CrossRefPubMed Grunz J-P, Heidenreich JF, Lennartz S, Weighardt JP, Bley TA, Ergün S, et al. Spectral shaping Via Tin Prefiltration in Ultra-high-resolution Photon-counting and energy-integrating detector CT of the temporal bone. Invest Radiol. 2022;57:819–25.CrossRefPubMed
15.
Zurück zum Zitat Petritsch B, Kosmala A, Weng AM, Bley TA. Tin-filtered 100 kV ultra-low-dose CT of the paranasal sinus: initial clinical results. PLoS ONE. 2019;14:e0216295.CrossRefPubMedPubMedCentral Petritsch B, Kosmala A, Weng AM, Bley TA. Tin-filtered 100 kV ultra-low-dose CT of the paranasal sinus: initial clinical results. PLoS ONE. 2019;14:e0216295.CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Vonder M, Dorrius MD, Vliegenthart R. Latest CT technologies in lung cancer screening: protocols and radiation dose reduction. Transl Lung Cancer Res. 2021;10:1154–64.CrossRefPubMedPubMedCentral Vonder M, Dorrius MD, Vliegenthart R. Latest CT technologies in lung cancer screening: protocols and radiation dose reduction. Transl Lung Cancer Res. 2021;10:1154–64.CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Hasegawa A, Ichikawa K, Morioka Y, Kawashima H. A tin filter’s dose reduction effect revisited: using the detectability index in low-dose computed tomography for the chest. Phys Med. 2022;99:61–7.CrossRefPubMed Hasegawa A, Ichikawa K, Morioka Y, Kawashima H. A tin filter’s dose reduction effect revisited: using the detectability index in low-dose computed tomography for the chest. Phys Med. 2022;99:61–7.CrossRefPubMed
18.
Zurück zum Zitat Kimura K, Fujioka T, Mori M, Adachi T, Hiraishi T, Hada H, et al. Dose reduction and diagnostic performance of Tin Filter–Based Spectral shaping CT in patients with colorectal Cancer. Tomography. 2022;8:1079–89.CrossRefPubMedPubMedCentral Kimura K, Fujioka T, Mori M, Adachi T, Hiraishi T, Hada H, et al. Dose reduction and diagnostic performance of Tin Filter–Based Spectral shaping CT in patients with colorectal Cancer. Tomography. 2022;8:1079–89.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Suntharalingam S, Allmendinger T, Blex S, Al-Bayati M, Nassenstein K, Schweiger B, et al. Spectral Beam shaping in unenhanced chest CT examinations: a Phantom Study on Dose Reduction and Image Quality. Acad Radiol. 2018;25:153–8.CrossRefPubMed Suntharalingam S, Allmendinger T, Blex S, Al-Bayati M, Nassenstein K, Schweiger B, et al. Spectral Beam shaping in unenhanced chest CT examinations: a Phantom Study on Dose Reduction and Image Quality. Acad Radiol. 2018;25:153–8.CrossRefPubMed
20.
Zurück zum Zitat Grunz J-P, Halt D, Schüle S, Beer M, Hackenbroch C. Thermoluminescence Dosimetry in Abdominal CT for urinary stone detection. Invest Radiol. 2023;58:231–8.CrossRefPubMed Grunz J-P, Halt D, Schüle S, Beer M, Hackenbroch C. Thermoluminescence Dosimetry in Abdominal CT for urinary stone detection. Invest Radiol. 2023;58:231–8.CrossRefPubMed
21.
Zurück zum Zitat Kunz AS, Grunz J, Halt D, Kalogirou C, Luetkens KS, Patzer TS, et al. Tin-filtered 100 kV ultra-low-dose abdominal CT for Calculi Detection in the urinary tract: a comparative study of 510 cases. Acad Radiol. 2022;30:1033–8.CrossRefPubMed Kunz AS, Grunz J, Halt D, Kalogirou C, Luetkens KS, Patzer TS, et al. Tin-filtered 100 kV ultra-low-dose abdominal CT for Calculi Detection in the urinary tract: a comparative study of 510 cases. Acad Radiol. 2022;30:1033–8.CrossRefPubMed
22.
Zurück zum Zitat Koo TK, Li MY. A Guideline of selecting and reporting Intraclass correlation coefficients for Reliability Research. J Chiropr Med. 2016;15:155–63.CrossRefPubMedPubMedCentral Koo TK, Li MY. A Guideline of selecting and reporting Intraclass correlation coefficients for Reliability Research. J Chiropr Med. 2016;15:155–63.CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Schegerer A, Loose R, Heuser LJ, Brix G. Diagnostic reference levels for diagnostic and interventional X-Ray procedures in Germany: Update and Handling. Rofo. 2019;191:739–51.CrossRefPubMed Schegerer A, Loose R, Heuser LJ, Brix G. Diagnostic reference levels for diagnostic and interventional X-Ray procedures in Germany: Update and Handling. Rofo. 2019;191:739–51.CrossRefPubMed
24.
Zurück zum Zitat Hackenbroch C, Feilhuber M, Halt D, Riesner HJ, Beer M, Wunderlich A. Low-dose CT in pelvic imaging: comparing dose and image quality in relation to clinical value in a phantom study. Am J Roentgenol. 2021;216:453–63.CrossRef Hackenbroch C, Feilhuber M, Halt D, Riesner HJ, Beer M, Wunderlich A. Low-dose CT in pelvic imaging: comparing dose and image quality in relation to clinical value in a phantom study. Am J Roentgenol. 2021;216:453–63.CrossRef
25.
Zurück zum Zitat Zhang G-M-YMY, Shi B, Sun H, Xue H-DD, Wang Y, Liang J-XX, et al. High-pitch low-dose abdominopelvic CT with tin-filtration technique for detecting urinary stones. Abdom Radiol. 2017;42:2127–34.CrossRef Zhang G-M-YMY, Shi B, Sun H, Xue H-DD, Wang Y, Liang J-XX, et al. High-pitch low-dose abdominopelvic CT with tin-filtration technique for detecting urinary stones. Abdom Radiol. 2017;42:2127–34.CrossRef
26.
Zurück zum Zitat Mussmann B, Hardy M, Jung H, Ding M, Osther PJ, Fransen ML, et al. Renal stone detection using a low kilo-voltage paediatric CT protocol - a porcine phantom study. J Med Radiat Sci. 2021;68:342–8.CrossRefPubMedPubMedCentral Mussmann B, Hardy M, Jung H, Ding M, Osther PJ, Fransen ML, et al. Renal stone detection using a low kilo-voltage paediatric CT protocol - a porcine phantom study. J Med Radiat Sci. 2021;68:342–8.CrossRefPubMedPubMedCentral
27.
Zurück zum Zitat Mozaffary A, Trabzonlu TA, Kim D, Yaghmai V. Comparison of tin filter–based spectral shaping CT and low-dose protocol for detection of urinary calculi. Am J Roentgenol. 2019;212:808–14.CrossRef Mozaffary A, Trabzonlu TA, Kim D, Yaghmai V. Comparison of tin filter–based spectral shaping CT and low-dose protocol for detection of urinary calculi. Am J Roentgenol. 2019;212:808–14.CrossRef
28.
Zurück zum Zitat Karmazyn B, Liang Y, Klahr P, Jennings SG. Effect of tube voltage on ct noise levels in different phantom sizes. Am J Roentgenol. 2013;200:1001–5.CrossRef Karmazyn B, Liang Y, Klahr P, Jennings SG. Effect of tube voltage on ct noise levels in different phantom sizes. Am J Roentgenol. 2013;200:1001–5.CrossRef
29.
Zurück zum Zitat Li J, Mai Z, Zhang Z, Cui J, Yang M, Ma X, et al. Chest CT screening in patients with overweight or obesity using spectral shaping at 150 kVp: compared with 120 kVp protocol and spectral shaping at 100 kVp protocol. Jpn J Radiol. 2020;38:451–7.CrossRefPubMed Li J, Mai Z, Zhang Z, Cui J, Yang M, Ma X, et al. Chest CT screening in patients with overweight or obesity using spectral shaping at 150 kVp: compared with 120 kVp protocol and spectral shaping at 100 kVp protocol. Jpn J Radiol. 2020;38:451–7.CrossRefPubMed
30.
Zurück zum Zitat Huflage H, Grunz JP, Hackenbroch C, Halt D, Luetkens KS, Alfred Schmidt AM, et al. Metal artefact reduction in low-dose computed tomography: benefits of tin prefiltration versus postprocessing of dual-energy datasets over conventional CT imaging. Radiography. 2022;28:690–6.CrossRefPubMed Huflage H, Grunz JP, Hackenbroch C, Halt D, Luetkens KS, Alfred Schmidt AM, et al. Metal artefact reduction in low-dose computed tomography: benefits of tin prefiltration versus postprocessing of dual-energy datasets over conventional CT imaging. Radiography. 2022;28:690–6.CrossRefPubMed
31.
Zurück zum Zitat Hackenbroch C, Schüle S, Halt D, Zengerle L, Beer M. Metal Artifact Reduction with Tin Prefiltration in Computed Tomography: a cadaver study for comparison with other Novel techniques. Invest Radiol. 2022;57:194–203.CrossRefPubMed Hackenbroch C, Schüle S, Halt D, Zengerle L, Beer M. Metal Artifact Reduction with Tin Prefiltration in Computed Tomography: a cadaver study for comparison with other Novel techniques. Invest Radiol. 2022;57:194–203.CrossRefPubMed
32.
Zurück zum Zitat Chang KJ, Caovan DB, Grand DJ, Huda W, Mayo-Smith WW. Reducing radiation dose at CT colonography: decreasing tube voltage to 100 kVp. Radiology. 2013;266:791–800.CrossRefPubMed Chang KJ, Caovan DB, Grand DJ, Huda W, Mayo-Smith WW. Reducing radiation dose at CT colonography: decreasing tube voltage to 100 kVp. Radiology. 2013;266:791–800.CrossRefPubMed
33.
Zurück zum Zitat Matsunaga Y, Haba T, Kobayashi M, Suzuki S, Asada Y, Chida K. Fetal radiation dose of four tube voltages in abdominal CT examinations during pregnancy: a phantom study. J Appl Clin Med Phys. 2021;22:178–84.CrossRefPubMedPubMedCentral Matsunaga Y, Haba T, Kobayashi M, Suzuki S, Asada Y, Chida K. Fetal radiation dose of four tube voltages in abdominal CT examinations during pregnancy: a phantom study. J Appl Clin Med Phys. 2021;22:178–84.CrossRefPubMedPubMedCentral
34.
Zurück zum Zitat Nakayama Y, Awai K, Funama Y, Liu D, Nakaura T, Tamura Y, et al. Lower tube voltage reduces contrast material and radiation doses on 16-MDCT aortography. AJR Am J Roentgenol. 2006;187:490–7.CrossRef Nakayama Y, Awai K, Funama Y, Liu D, Nakaura T, Tamura Y, et al. Lower tube voltage reduces contrast material and radiation doses on 16-MDCT aortography. AJR Am J Roentgenol. 2006;187:490–7.CrossRef
35.
Zurück zum Zitat Huflage H, Grunz JP, Patzer TS, Pannenbecker P, Feldle P, Sauer ST, et al. Potential of Unenhanced Ultra-low-dose Abdominal Photon-counting CT with tin filtration: a cadaveric study. Diagnostics. 2023;13(4):603.CrossRefPubMedPubMedCentral Huflage H, Grunz JP, Patzer TS, Pannenbecker P, Feldle P, Sauer ST, et al. Potential of Unenhanced Ultra-low-dose Abdominal Photon-counting CT with tin filtration: a cadaveric study. Diagnostics. 2023;13(4):603.CrossRefPubMedPubMedCentral
36.
Zurück zum Zitat Dewes P, Frellesen C, Scholtz J-E, Fischer S, Vogl TJ, Bauer RW, et al. Low-dose abdominal computed tomography for detection of urinary stone disease– impact of additional spectral shaping of the X-ray beam on image quality and dose parameters. Eur J Radiol. 2016;85:1058–62.CrossRefPubMed Dewes P, Frellesen C, Scholtz J-E, Fischer S, Vogl TJ, Bauer RW, et al. Low-dose abdominal computed tomography for detection of urinary stone disease– impact of additional spectral shaping of the X-ray beam on image quality and dose parameters. Eur J Radiol. 2016;85:1058–62.CrossRefPubMed
37.
Zurück zum Zitat Haubenreisser H, Meyer M, Sudarski S, Allmendinger T, Schoenberg SO, Henzler T. Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100 kVp. Eur J Radiol. 2015;84:1608–13.CrossRefPubMed Haubenreisser H, Meyer M, Sudarski S, Allmendinger T, Schoenberg SO, Henzler T. Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100 kVp. Eur J Radiol. 2015;84:1608–13.CrossRefPubMed
Metadaten
Titel
Influence of spectral shaping and tube voltage modulation in ultralow-dose computed tomography of the abdomen
verfasst von
Philipp Feldle
Jan-Peter Grunz
Andreas Steven Kunz
Pauline Pannenbecker
Theresa Sophie Patzer
Svenja Pichlmeier
Stephanie Tina Sauer
Robin Hendel
Süleyman Ergün
Thorsten Alexander Bley
Henner Huflage
Publikationsdatum
01.12.2024
Verlag
BioMed Central
Erschienen in
BMC Medical Imaging / Ausgabe 1/2024
Elektronische ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-024-01228-1

Weitere Artikel der Ausgabe 1/2024

BMC Medical Imaging 1/2024 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.