Erschienen in:
14.08.2023 | Original Article
Total-body [68 Ga]Ga-PSMA-11 PET/CT improves detection rate compared with conventional [68 Ga]Ga-PSMA-11 PET/CT in patients with biochemical recurrent prostate cancer
verfasst von:
Yining Wang, Zijun Chen, Yinjie Zhu, Haitao Zhao, Lianghua Li, Gang Huang, Wei Xue, Ruohua Chen, Jianjun Liu
Erschienen in:
European Journal of Nuclear Medicine and Molecular Imaging
|
Ausgabe 13/2023
Einloggen, um Zugang zu erhalten
Abstract
Purpose
The purpose of this study was to assess whether total-body [68 Ga]Ga-PSMA-11 PET/CT could improve the detection rate compared with conventional [68 Ga]Ga-PSMA-11 PET/CT in patients with biochemical recurrent prostate cancer.
Methods
Two hundred biochemical recurrent prostate cancer patients with similar clinicopathological characteristics were included, of whom 100 patients underwent early total-body [68 Ga]Ga-PSMA-11 PET/CT and diuretic-delayed total-body [68 Ga]Ga-PSMA-11 PET/CT, and the other 100 patients received early conventional [68 Ga]Ga-PSMA-11 PET/CT and diuretic-delayed conventional [68 Ga]Ga-PSMA-11 PET/CT. The detection rates of total-body [68 Ga]Ga-PSMA-11 PET/CT and conventional [68 Ga]Ga-PSMA-11 PET/CT were compared using a chi-square test and stratified analysis. The image quality of total-body [68 Ga]Ga-PSMA PET/CT and conventional [68 Ga]Ga-PSMA-11 PET/CT was compared based on subjective scoring and objective parameters. Subjective scoring was conducted from background noise and lesion prominence using a 5-point scale. Objective parameters were evaluated by SUVmax, SUVmean, the standard deviation (SD) of SUV, and the signal-to-noise ratio (SNR) of liver and gluteus maximus. The SUVmax of the recurrent lesions was also measured.
Results
The liver SD of the total-body [68 Ga]Ga-PSMA-11 PET/CT was significantly lower than that of conventional [68 Ga]Ga-PSMA-11 PET/CT, the SNR was significantly higher than that of conventional [68 Ga]Ga-PSMA-11 PET/CT, and the subjective evaluation was significantly better than that of conventional [68 Ga]Ga-PSMA-11 PET/CT. The detection rate of total-body [68 Ga]Ga-PSMA PET/CT for biochemical recurrence of prostate cancer was significantly higher than that of conventional [68 Ga]Ga-PSMA-11 PET/CT (91.0% vs. 74.0%, P = 0.003). Total-body [68 Ga]Ga-PSMA-11 PET/CT had better detection efficiency for patients with a Gleason score ≤ 8 or PSA ≤ 2 ng/ml. The advantages of diuretic-delayed total-body [68 Ga]Ga-PSMA-11 PET/CT were more obvious.
Conclusion
Total-body [68 Ga]Ga-PSMA-11 PET/CT could significantly improve the detection rate compared with conventional [68 Ga]Ga-PSMA-11 PET/CT in patients with biochemical recurrent prostate cancer.