The reproducibility of measuring maximum abdominal aortic aneurysm diameter from ultrasound images

verfasst von: Evan O. Matthews, Jenna Pinchbeck, Kylie Elmore, Rhondda E. Jones, Joseph V. Moxon, Jonathan Golledge

Erschienen in: The Ultrasound Journal | Ausgabe 1/2021

Abstract

Background

Accurate repeat assessment of the diameter of an abdominal aortic aneurysm (AAA) is important. This study investigated the reproducibility of different methods of measuring AAA diameter from ultrasound images.

Methods

Fifty AAA patients were assessed by ultrasound. Maximum AAA diameter was measured independently by three trained observers on two separate occasions using a standardised protocol. Five diameters were measured from each scan, three in the anterior–posterior (AP) and two in the transverse (TV) plane, including inner-to-inner (ITI), outer-to-outer (OTO) and leading edge-to-leading edge (LETLE). Intra- and inter-observer reproducibility were reported as reproducibility coefficients. Statistical comparison of methods was performed using linear mixed effects models.

Results

Intra-observer reproducibility coefficients (AP LETLE 2.2 mm; AP ITI 2.4 mm; AP OTO 2.6 mm) were smaller than inter-observer reproducibility coefficients (AP LETLE 4.6 mm: AP ITI 4.5; and AP OTO 4.8 mm). There was no statistically significant difference in intra-observer reproducibility of three types of measurements performed in the AP plane. Measurements obtained in the TV plane had statistically significant worse intra-observer reproducibility than those performed in the AP plane.

Conclusions

This study suggests that the comparison of maximum AAA diameter between repeat images is most reproducibly performed by a single trained observer measuring diameters in the AP plane.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
AAA
Abdominal aortic aneurysm
AP
Anterior–posterior anatomical plane of the body
TV
Transverse plane of the body
ITI
Inner-to-inner walls of the aorta
OTO
Outer-to-outer walls of the aorta
LETLE
Leading edge-to-leading edge within the walls of the aortas
LOA
Limits of agreement

Background

Approximately 2% of men and 0.5% of women aged > 65 years develop an abdominal aortic aneurysm (AAA) [1, 2]. Maximum AAA diameter is the most established predictor of AAA growth and rupture and is used in clinical practice to guide decision-making [35]. Current guidelines recommend considering AAA repair when the maximum diameter is ≥ 55 mm in men and ≥ 50 mm in women [6, 7]. Management of smaller asymptomatic AAAs is by surveillance with repeat imaging performed at intervals to monitor maximum AAA diameter. Furthermore, there is growing interest in the identification of drug therapies to slow AAA growth [8, 9]. Trials testing potential drugs usually assess outcome by monitoring maximum AAA diameter growth over time [10]. Reproducible methods to measure AAA diameter are therefore of both clinical and research importance.
In clinical practice and most previous clinical trials ultrasound imaging has been used to estimate maximum AAA diameter [6, 10]. Despite the importance of accurate determination of AAA diameter, measurement protocols are often incompletely reported and vary in both plane of acquisition and calliper placement [11]. The United Kingdom Small Aneurysm Trial reported maximal anterior–posterior outer-to-outer diameter (OTO) from ultrasound [12, 13]. AAA screening programmes have used a variety of different methods to measure AAA diameter on ultrasound, including the inner-to-inner (ITI) or leading edge-to-leading-edge (LETLE) methods of calliper placement [14, 15]. Disparate methods of calliper placement has been reported to cause differences of up to 5 mm in maximal AAA diameter with implications for decision-making regarding surgical repair and surveillance intervals which could impact on patient care [16, 17]. A number of theoretical principles have been cited as justification for different calliper selection, but no standardised method has been ubiquitously adopted. Three recent studies have compared these three methods of calliper placement, but had inconsistent findings [1618].
A key aspect in choosing a method of measurement is its repeatability or reproducibility. This is particularly important for AAA diameter as this is commonly remeasured at intervals during which only small changes occur. This study aimed to compare the inter- and intra-observer repeatability of five different methods of measuring AAA diameter using distinct measurement planes and calliper placement.

Methods

Study design and participant recruitment

Participants were recruited from The Department of Vascular and Endovascular Surgery at Townsville University Hospital [19]. All participants provided written informed consent for inclusion and ethics approval was obtained from The Townsville Hospital and Health Service Human Research Ethics Committee (13/QPCH/16). A convenience sample size of 50 participants was used in concordance with previous similar studies [16, 17, 20]. Participants met the following inclusion criteria: (1) a small infra-renal AAA measuring 30–55 mm in maximal diameter and (2) an ultrasound performed by one of three experienced vascular sonographers within the Townsville University Hospital Vascular Laboratory using a standardised protocol. Patients with a history of previous abdominal aortic surgery were excluded. At recruitment an interview and physical examination was conducted to collect relevant medical history and clinical measurements. Ischaemic heart disease, hypertension and diabetes were defined as a previous diagnosis or treatment of these conditions by a qualified medical physician. Stroke was defined as a documented history of ischaemic or haemorrhagic stroke. The presence of aneurysms at other sites was determined through documented history or relevant medical imaging. Brachial blood pressure, waist circumference and body mass index were measured as previously described [21].

Ultrasound imaging

All ultrasound scans were performed using a Phillips iu22 machine (Phillips Medical Systems, United States) with a C5-1 MHz general purpose curvilinear abdominal transducer by one of three experienced sonographers between December 2013 and December 2015. Each sonographer was a formally trained and registered sonographer with specialised experience in vascular ultrasound. Each participant was fasted for 12 h prior to their ultrasound scan to minimise interference from bowel gas. All participants were scanned in the supine position. The abdominal aorta was examined in both the transverse and sagittal planes to identify infra-renal landmarks, vessel tortuosity and obliqueness. Static ultrasound images in the transverse plane acquired in systole were obtained at the point of maximal dilation of the infra-renal abdominal aorta perpendicular to the central vessel line. The images were then centrally stored in accordance with health service policy.

Measurement protocol

Three observers were trained to measure maximal AAA diameter on static ultrasound images using a predefined protocol. Observers were selected based on pre-existing knowledge of aortic anatomy, cardiovascular physiology and imaging. Observer one was a qualified vascular sonographer with extensive experience in acquiring and interpreting ultrasound imaging of the aorta. Observer two was a clinical medical student with previous experience measuring AAA growth on computerised tomographic angiography scans. Observer three was a research worker and exercise physiologist with extensive experience in interpreting static ultrasound images of the abdominal aorta. A set measurement protocol was developed in consultation between a vascular sonographer, vascular surgeon and researcher. Five measurements were performed to assess AAA diameter. Three in the anterior–posterior (OTO, ITI and LETLE) and two (OTO and ITI) in the transverse plane (Fig. 1). Observer training involved both theoretical discussion and practical demonstration of the measurement protocol. Observers then measured a separate series of ten static ultrasound scan images independently before a consensus discussion was conducted on calliper placement between observers. Static ultrasound images were imported as DICOM images to the OsiriX Lite 32-bit version (Pixmeo, Geneva, Switzerland) software for analysis. To avoid bias, only static images where the sonographer placed measurement callipers had been omitted were included in this study. Prior to each measurement, each observer measured a 10-mm marked interval on a 100-mm scale to ensure accurate calibration of callipers. Each observer independently measured identical static images from the 50 participants and were blinded to the other observers’ measurements. Each observer repeated measurements 1 week later blinded to their earlier results.

Data analysis

Data analyses were performed in SPSS Version 23.0 (IBM SPSS Inc., Chicago, IL, United States) and R (Foundation for Statistical Computing, Vienna, Austria) with assistance from an expert statistician (REJ). Demographic data were reported as count (%) for dichotomous data and median (inter-quartile range 25th to 75th centile) for continuous data.

Intra- and inter-observer reproducibility

The mean and standard deviation (SD) of the differences between measurement one and two were calculated for each observer. The SD of the differences were then multiplied by 1.96 to obtain the reproducibility coefficient, or 95% limits of agreement (LOA), for each individual observer. The MethComp package in R was used to combine data from all three observers to calculate the pooled reproducibility coefficient for each individual method. Each scan–observer combination was treated as an individual item. Linear mixed effects models were used to formally test for a significant difference in repeatability between each individual method by comparing squared mean differences from a reference method (anterior–posterior ITI). Observers and subjects were both treated as random effects and mean difference and calliper placement method as fixed effects. The overall mean difference between ITI, LETLE and OTO for each scan was also calculated.

Results

The risk factors of the included patients are reported in Table 1. Median maximum diameter of the included patients varied by up to 8.3 mm depending on the measurement method used (Table 1).
Table 1
Patient characteristics
Characteristic
Number (%)
Male
38 (74%)
Age
72 (68–77)
AAA AP OTO diameter (mm)
41.8 (37.4–44.9)
AAA AP LETLE diameter (mm)
38.2 (34.4–41.7)
AAA AP ITI diameter (mm)
35.8 (31.9–38.8)
AAA TV OTO diameter (mm)
44.1 (38.8–48.4)
AAA TV ITI diameter (mm)
37.3 (33.0–41.3)
Systolic blood pressure*
138 (127–148)
Diastolic blood pressure*
77 (70–82)
Hypertension
42 (84%)
Diabetes
11 (22%)
Ischaemic heart disease
34 (68%)
Stroke
8 (16%)
Aneurysm at another site
11 (22%)
Waist circumference (cm)
105.5 (97–114)
Body mass index (kg.m−2)
28.5 (26.1–32.9)
Continuous variables are presented as median (inter-quartile range)
Categorical variables are presented as count (percent)
AP: anterior–posterior; TV: transverse; OTO: outer-to-outer; LETE: leading edge-to-leading edge; ITI: inner-to-inner
*Four values missing

Intra-observer reproducibility

The intra-observer reproducibility coefficients for each individual observer are shown in Table 2. There was notable variation in reproducibility coefficient between each individual method and each individual observer. Scans obtained in the anterior–posterior plane and with LETLE calliper placement had the lowest overall intra-observer reproducibility coefficient (± 2.2 mm), but this was not statistically significantly different from ITI and OTO calliper placement in the same plane. All three observers consistently measured images in the transverse plane with a poorer repeatability as highlighted by the poorer overall intra-observer reproducibility coefficient for both ITI (± 4.9 mm) and OTO (± 5.6 mm) calliper placement. Measurements in the transverse plane (ITI P = 0.001; OTO P < 0.001) were significantly less reproducible than those measured in the AP plane (Table 2).
Table 2
Intra-observer reproducibility coefficients
Anatomical plane
Calliper placement
Reproducibility coefficient
Comparison between methods
Observer 1 (± mm)
Observer 2 (± mm)
Observer 3 (± mm)
Overall (± mm)
Mean difference squared
Standard error
P-value
AP
ITI
2.22
1.68
2.65
2.35
1.38
1.19
Reference
AP
LETLE
2.17
1.89
2.37
2.16
-0.17
1.43
0.906
AP
OTO
2.31
2.53
2.50
2.60
0.31
1.43
0.830
Transverse
ITI
4.76
4.91
4.83
4.90
4.63
1.44
0.001
Transverse
OTO
5.62
4.74
5.51
5.62
6.49
1.44
 < 0.001
ITI: Inner-to-inner; OTO: outer-to-outer; LETLE: leading edge-to-leading edge; AP: anterior–posterior

Inter-observer reproducibility

The inter-observer reproducibility coefficient for each method is shown in Table 3. The inter-observer reproducibility coefficients were poorer than the corresponding intra-observer reproducibility coefficient for each method. The inter-observer reproducibility was poorest for images measured transversely (Table 3).
Table 3
Inter-observer reproducibility coefficients
Patient position
Anatomical direction
Method
Reproducibility coefficient (± mm)
Mean diameter difference from reference method (mm)
Supine
AP
ITI
4.47
Reference
Supine
AP
LETLE
4.59
2.69
Supine
AP
OTO
4.82
5.52
Supine
Transverse
ITI
6.02
Reference
Supine
Transverse
OTO
6.22
6.40
ITI: inner-to-inner; OTO: outer-to-outer; LETLE: leading edge-to-leading edge; AP: anterior–posterior

Discussion

This study examined the influence of alternative methods of measurement of AAA diameter under conditions typically required in clinical trials. There was no statistically significant difference between alternative methods of measurement where calliper placement was in line with probe positioning (anterior–posterior for supine position). Measurements obtained perpendicular to the probe (transverse in the supine position) have been reported to be less repeatable due to poorer resolution of the lateral vessel walls [22]. This study supports this finding with statistically significant worse intra-observer reproducibility in both ITI and OTO measurements obtained in the transverse direction compared to those measured anterior–posterior.
Sixteen previous studies reporting the reproducibility of abdominal aortic measurements with ultrasound were identified [14, 1618, 20, 2333]. There was marked variation in reproducibility coefficients for both intra-observer (range ± 0.9 mm to ± 4.0 mm) and inter-observer (range ± 1.7 mm to ± 12.6 mm) repeatability. Gurtelschmid et al. reported better inter-observer reproducibility coefficients in anterior–posterior LETLE (± 4.0 mm) and anterior–posterior ITI (± 4.6 mm) calliper placement when compared with anterior–posterior OTO (± 5.3 mm) calliper placement. Borgbjerg et al. reported similar findings with better inter-observer reproducibility coefficients with anterior–posterior LETLE (± 3.8 mm) and anterior–posterior ITI (± 3.9 mm) calliper placement compared with anterior–posterior OTO (± 5.2 mm) [16, 18]. These findings are in contrast to those of Chui et al. who reported no statistical difference in reproducibility coefficients between these three methods (anterior–posterior LETLE ± 3.5 mm; anterior–posterior ITI ± 4.8 mm; anterior–posterior OTO ± 3.4 mm) [17]. The current study found no statistically significant differences between different methods of calliper placement when only measurements obtained in the same plane as the ultrasound probe are considered. The overall intra-observer reproducibility found in the current study are similar to those previously reported [16, 18].
The mean difference between AAA diameter measured by the ITI, LETLE and OTO methods were comparable to those previously reported and relate to vessel wall thickness [16, 17]. These differences highlight the importance of having clearly defined methods of calliper placement that are consistently used in both clinical practice and research. Multiple studies have looked at the influence of using alternative methods of calliper placement on the recruitment of patients into surveillance programmes. ITI measurements underestimate AAA size and lead to reduced sensitivity when used as a screening tool. A previous study analysed the influence of calliper placement on AAA prevalence in a cohort of 18,698 patients and found that it led to a significant difference in AAA diagnosis and subsequent recruitment into surveillance programmes (AAA prevalence ITI = 3.3%, LETLE = 4.0% and OTO = 5.9%) [16].
This study suggests that the ITI, OTO and LETE calliper placement methods can be equally well reproduced when placed in the same plane as the US probe, i.e. anterior–posterior. The measurement of transverse ITI or OTO diameter is not as reproducible. These findings suggest that measurement in the anterior–posterior plane should be used in clinical practice and clinical trials. Since the repeatability of measurements is much better within rather than between individuals it is also preferable for measurements to be performed by the same observer.
The current study used modern ultrasound technology and standardised methodology to directly compare the three leading methods of calliper placement. Of the 16 previous reproducibility studies identified seven [23, 27, 2933] were published prior to 2000 and only three [1618] reported the inter-observer reproducibility for all three methods of calliper placement and two [16, 17] the intra-observer reproducibility. This study examined the variation in measurements introduced by different methods of calliper placement on static AAA images obtained at a single time point. Measurement error introduced during the acquisition of scans was not assessed and therefore in clinical practice the reproducibility coefficients are likely larger.

Conclusions

In conclusion, AAA diameter measurements obtained perpendicular to the orientation of the ultrasound probe (anterior–posterior) can be performed more reproducibly than those performed in the transverse plane. Measurements performed by the same observer also have better repeatability than those performed by different observers. The findings suggest that measurements of AAA size should be performed in the anterior–posterior plane and compared between different time periods by the same observer, particularly for situations such as clinical trials where high precision is required to sensitively detect changes in AAA diameter over time.

Acknowledgements

Not applicable.
All participants provided written informed consent for inclusion and ethics approval was obtained from The Townsville Hospital and Health Service Human Research Ethics Committee (13/QPCH/16).
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
13.
Zurück zum Zitat Participants TU (1998) Mortality results for randomised controlled trial of early elective surgery or ultrasonographic surveillance for small abdominal aortic aneurysms. The UK Small Aneurysm Trial Participants. Lancet 352(9141):1649–1655 (Epub 1998/12/16)CrossRef Participants TU (1998) Mortality results for randomised controlled trial of early elective surgery or ultrasonographic surveillance for small abdominal aortic aneurysms. The UK Small Aneurysm Trial Participants. Lancet 352(9141):1649–1655 (Epub 1998/12/16)CrossRef
19.
Zurück zum Zitat Pinchbeck JL, Moxon JV, Rowbotham SE, Bourke M, Lazzaroni S, Morton SK, et al. Randomized Placebo-Controlled Trial Assessing the Effect of 24-Week Fenofibrate Therapy on Circulating Markers of Abdominal Aortic Aneurysm: Outcomes From the FAME-2 Trial. J Am Heart Assoc. 2018;7(19):e009866. doi: https://doi.org/10.1161/JAHA.118.009866. Epub 2018/10/30 Pinchbeck JL, Moxon JV, Rowbotham SE, Bourke M, Lazzaroni S, Morton SK, et al. Randomized Placebo-Controlled Trial Assessing the Effect of 24-Week Fenofibrate Therapy on Circulating Markers of Abdominal Aortic Aneurysm: Outcomes From the FAME-2 Trial. J Am Heart Assoc. 2018;7(19):e009866. doi: https://​doi.​org/​10.​1161/​JAHA.​118.​009866. Epub 2018/10/30
Metadaten
Titel
The reproducibility of measuring maximum abdominal aortic aneurysm diameter from ultrasound images
verfasst von
Evan O. Matthews
Jenna Pinchbeck
Kylie Elmore
Rhondda E. Jones
Joseph V. Moxon
Jonathan Golledge
Publikationsdatum
01.12.2021
Verlag
Springer International Publishing
Erschienen in
The Ultrasound Journal / Ausgabe 1/2021
Print ISSN: 2036-3176
Elektronische ISSN: 2524-8987
DOI
https://doi.org/10.1186/s13089-021-00211-z

Weitere Artikel der Ausgabe 1/2021

The Ultrasound Journal 1/2021 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Bei seelischem Stress sind Checkpoint-Hemmer weniger wirksam

03.06.2024 NSCLC Nachrichten

Wie stark Menschen mit fortgeschrittenem NSCLC von einer Therapie mit Immun-Checkpoint-Hemmern profitieren, hängt offenbar auch davon ab, wie sehr die Diagnose ihre psychische Verfassung erschüttert

Antikörper mobilisiert Neutrophile gegen Krebs

03.06.2024 Onkologische Immuntherapie Nachrichten

Ein bispezifischer Antikörper formiert gezielt eine Armee neutrophiler Granulozyten gegen Krebszellen. An den Antikörper gekoppeltes TNF-alpha soll die Zellen zudem tief in solide Tumoren hineinführen.

Erhebliches Risiko für Kehlkopfkrebs bei mäßiger Dysplasie

29.05.2024 Larynxkarzinom Nachrichten

Fast ein Viertel der Personen mit mäßig dysplastischen Stimmlippenläsionen entwickelt einen Kehlkopftumor. Solche Personen benötigen daher eine besonders enge ärztliche Überwachung.

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.