Simultaneous venous–arterial Doppler during preload augmentation: illustrating the Doppler Starling curve

verfasst von: Jon-Émile S. Kenny, Stanley O. Gibbs, Joseph K. Eibl, Andrew M. Eibl, Zhen Yang, Delaney Johnston, Chelsea E. Munding, Mai Elfarnawany, Vivian C. Lau, Benjamin O. Kemp, Bhanu Nalla, Rony Atoui

Erschienen in: The Ultrasound Journal | Ausgabe 1/2023

Abstract

Providing intravenous (IV) fluids to a patient with signs or symptoms of hypoperfusion is common. However, evaluating the IV fluid ‘dose–response’ curve of the heart is elusive. Two patients were studied in the emergency department with a wireless, wearable Doppler ultrasound system. Change in the common carotid arterial and internal jugular Doppler spectrograms were simultaneously obtained as surrogates of left ventricular stroke volume (SV) and central venous pressure (CVP), respectively. Both patients initially had low CVP jugular venous Doppler spectrograms. With preload augmentation, only one patient had arterial Doppler measures indicative of significant SV augmentation (i.e., ‘fluid responsive’). The other patient manifested diminishing arterial response, suggesting depressed SV (i.e., ‘fluid unresponsive’) with evidence of ventricular asynchrony. In this short communication, we describe how a wireless, wearable Doppler ultrasound simultaneously tracks surrogates of cardiac preload and output within a ‘Doppler Starling curve’ framework; implications for IV fluid dosing are discussed.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
IV
Intravenous
SV
Stroke volume
SV
Change in stroke volume
CVP
Central venous pressure
PLR
Passive leg raise
ED
Emergency department
FDA
Food and Drug Administration
ccFT
Corrected carotid flow time
ccFT
Change in corrected carotid flow time
VTIc
Velocity time integral of the carotid artery
VTIc
Change in the velocity time integral of the carotid artery
VTI
Velocity time integral
FT
Flow time
RFC
Rapid fluid challenge
mL/min
Milliliters per minute
mL
Milliliters
ms
Milliseconds
LVEF
Left ventricular ejection fraction
Ivd
Interventricular delay
RV
Right ventricular
LV
Left ventricular
IVC
Inferior vena cava
VExUS
Venous excess ultrasound score
TTE
Transthoracic echocardiography

Main text

Introduction

The prescription of intravenous (IV) fluid is a clinical decision most often triggered by signs and symptoms of organ hypoperfusion and guided by traditional vital signs [1, 2]. However, the intended physiological effect of IV fluid (i.e., to augment stroke volume (SV) [3]) is rarely measured, especially in the emergency department [4]. Importantly, widely employed clinical markers such as urine output and traditional vital signs do not reliably indicate blood flow response [5]; therefore, without directly quantifying SV, the intended effect of cardiac preload is inscrutable.
In early sepsis and septic shock, a clinically significant proportion of patients receiving IV fluids do not exhibit SV augmentation and, as resuscitation progresses, more than 90% of patients cease having the anticipated SV-enhancing effect of a crystalloid infusion [68]. Importantly, withholding IV fluids in patients who fail to have the desired physiological effect is not harmful [6]. In fact, in a separate randomized and controlled trial, guiding IV fluids by changing SV (SV) led to significantly less IV fluid administered and improved patient-centered outcomes [9].
We have described a novel, wearable Doppler ultrasound that measures and displays common carotid artery and internal jugular Doppler spectrograms [1016]. These real-time data afford synchronously acquired surrogates of changing SV and right atrial pressure, respectively [12, 13, 1719]. Indeed, to our knowledge we were the first to describe simultaneous venous and arterial Doppler assessments during a passive leg raise (PLR) maneuver in a critically ill patient [12]. Further, a recent physiological framework based on this data elaborated the link between real-time venous and arterial Doppler [20]. In this short communication, we describe two patients receiving preload augmentation in an emergency department (ED) while monitored by the wearable Doppler; clinical and physiological implications are considered within the framework of a ‘Doppler Starling curve’.
Written and informed consent was obtained for both patients for publication of this report and accompanying images; the study was approved by the Peoria Institutional Review Board. The study was performed in accordance with the ethical standards as laid down in the 1964 Declaration of Helsinki and its later amendments. Both patients were recruited from a community emergency department and were deemed in need of IV fluid resuscitation by the treating clinician who was blinded to the results of the wearable ultrasound. The choice for preload augmentation was at the discretion of the treating clinician.

Wearable Doppler ultrasound

The ultrasound patch (Flosonics Medical, Sudbury, ON. Canada) is a wearable, wireless, FDA-cleared, continuous-wave 4 MHz ultrasound. Adhesive straps fix the transducer angle relative to the direction of carotid blood flow [21]. The wearable ultrasound displays real-time carotid corrected flow time (ccFT) as well as the common carotid velocity time integral (VTIc) (Fig. 1).

Doppler analysis

The Doppler spectrograms were analyzed for the absolute and % change in ccFT (ccFT) and % change in VTI (VTIc). The number of cardiac cycles averaged before and during preload augmentation was dictated by the coefficient of variation of each particular measure to ensure change could be detected with statistical confidence [23]. The assessment windows showing the largest change between baseline and preload augmentation were considered for analysis. The primary determinant of whether or not a patient was deemed ‘preload responsive’ was the threshold reported by Barjaktarevic and colleagues, that is, + 7 ms ccFT [24]. We also considered, as secondary determinants, the thresholds that we have identified during simulated severe hypovolemia and blood transfusion in healthy volunteers, that is +4% ccFT and +18% VTIc [14, 15].
The venous Doppler spectrograms were analyzed qualitatively based on the framework put forth by us [25] and Iida and colleagues [26]. These venous spectrogram changes have been previously described in the jugular vein [2729], superior vena cava [30], inferior vena cava [31], hepatic veins [32] and even the femoral vein [33]. Use of the hepatic, portal and intra-renal venous Doppler have been incorporated into the recently-described venous excess ‘VExUS’ score [34] (see Figs. 2, 3). Venous Doppler data were evaluated every 30 s before, during and after the preload challenge to assess for morphological changes consistent with rising right atrial (or central venous) pressure.

Preload augmentation

Both patients had baseline measures recorded in the semi-Fowler position at 45 degrees. Prior to preload augmentation, at least 30 s of continuous Doppler spectrograms were acquired with the patient instructed to remain motionless and breathe quietly. Patient 1 received a PLR by a clinical assistant lifting the legs to increase cardiac preload; the Doppler spectrograms were monitored continuously while the patient was supine with legs raised for an additional 60 s. For patient 2, the first 250 mL of each liter ordered comprised a ‘rapid fluid challenge’ (RFC) assessment. The 250 mL RFC was administered at a rate of 100 mL/min while the patient remained in semi-Fowler [35, 36]. This rate of fluid infusion was achieved using the LifeFlow (Durham, N.C., U.S.A.) device which can provide IV bolus rates upwards of 250 mL/min through a peripheral IV. During the RFC assessment, Doppler spectrograms were monitored continuously for an additional 180 s, thus recording the entire 250 mL of rapid saline infusion plus an additional 30 s of monitoring after the rapid infusion ended. The remaining 750 mL of each liter ordered were infused at the discretion of the treating clinician.

Patient 1

A 72-year-old man presented to the emergency department with altered mental status, back pain, hypotension, leukocytosis, acute kidney injury and a working diagnosis of sepsis due to lumbar osteomyelitis versus ascending urinary tract infection. He had a known history of tricuspid and mitral band valvuloplasties and diminished left ventricular ejection fraction (LVEF) of 40% consequent to chronic hypertension and ethanol use. His outpatient medications were lisinopril, furosemide and carvedilol. The patient’s intake vital signs were a heart rate of 83 beats per minute, blood pressure of 69/52 mmHg, afebrile with normal respiratory rate and oxygen saturation. His ECG showed a possible ectopic atrial rhythm with incomplete left bundle block pattern. In the emergency department, a PLR was performed prior to IV fluids and the results presented in Table 1 and Fig. 2.
Table 1
Carotid artery Doppler measures
Carotid Doppler measure
ccFT coefficient of variation
Cardiac cycle sample window
Absolute ccFT
% ccFT
% VTIc
Patient 1
3.3%
19 cardiac cycles
− 4.31 ms
− 2.0%
− 26.0%
Patient 2
4.1%
28 cardiac cycles
 + 16.3 ms
 + 5.0%
 + 11.0%
The coefficient of variation is the standard deviation divided by the mean of the baseline section, this was used to determine the number of cardiac cycles in the baseline and intervention periods needed to detect change with statistical confidence. The ccFT and VTIc for patient 2 but not patient 1 are consistent with a clinically significant increase in stroke volume [15, 24]
Patient 1 was treated with broad spectrum antibiotics and required norepinephrine infusion to maintain a mean arterial pressure above 60 mmHg. He was transferred to the intensive care unit where trans-thoracic echocardiography showed abnormal septal motion with systolic and diastolic flattening consistent with right ventricular (RV) pressure and volume overload, dilated RV size and reduced RV systolic function with RV hypertrophy. The patient had no significant valvular abnormalities, indeterminate left ventricular (LV) diastolic function, and a left ventricular ejection fraction (LVEF) of 50% while on norepinephrine.

Patient 2

A 54-year-old woman presented with shortness of breath and acute hypercapnic and hypoxemic respiratory failure secondary to sepsis from an infected sacral ulcer versus urinary source. She had history of traumatic quadriplegia with chronic sacral wounds complicated by osteomyelitis, indwelling suprapubic catheter, active cigarette smoker, COPD on home oxygen, diastolic dysfunction with congestive heart failure, obstructive sleep apnea and type II diabetes mellitus. She was febrile on presentation with a heart rate of 75 beats per minute, blood pressure 87/51 mmHg, respiratory rate 25 breaths per minute and 93% on 6 L nasal cannula. In the ED she received broad spectrum antibiotics. The patient had a Doppler assessment performed for the first 250 mL of each of the two liters received in the ED. Given her hypotension, the first and second liters were given in rapid succession (Table 1, Fig. 3).
This patient was briefly initiated on norepinephrine after the second liter and admitted to the intensive care unit. Trans-thoracic echocardiography in the ICU revealed a dilated right ventricle (RV) and inferior vena cava (IVC) with normal RV systolic function and no valvular abnormalities; the LVEF was 60–65% with normal diastolic function.

Discussion

The baseline, low power, non-pulsatile, high velocity venous Doppler spectrograms observed in both patients strongly suggested jugular vein collapse, indicating low CVP (i.e., < 5 mmHg) [25, 37, 38]. Nevertheless, the response of the arterial Doppler spectrogram for patient 1 during PLR revealed a state of ‘fluid unresponsiveness’, or perhaps even a detrimental response to preload (Fig. 4). Accordingly, patient 1 began in ‘quadrant 3’ [20] where his low, baseline CVP is potentially misleading with regard to IV fluid provision. That is to say, this patient disclosed preload intolerance only when a dynamic maneuver (e.g., PLR) was executed; thus, the ‘quadrant 3’ physiology illustrated is a state of dynamic fluid intolerance [20].
Recognizing a state of dynamic fluid intolerance is important clinically because many of these patients might receive physiologically unhelpful preload based upon markers of right heart filling. For example, a CVP of less than 5 mmHg was observed in 25% of fluid unresponsive patients [39], while in a meta-analysis, a CVP of less than 8 mmHg was observed in upwards of 40% of fluid unresponsive patients [40]. Similar physiology plays out with assessments of inferior vena cava size and collapsibility [41]. As might be anticipated, holding IV fluids in unresponsive patients (e.g., patient 1) does not have adverse effects in sepsis and septic shock [6, 42]. Furthermore, when IV fluid is guided by measures of ‘fluid responsiveness’, one multicenter, randomized and controlled trial in sepsis and septic shock showed that patients received less IV fluid and had improved outcomes [9]. Given that hundreds of millions of liters of IV fluid are administered every year in the United States alone [43], holding physiologically ineffective IV fluids might lead to significant cost-savings at the population level [16, 42, 44].
That the arterial Doppler response of patient 1 implied falling SV during the PLR deserves some elaboration, albeit speculative. Given the results of his trans-thoracic echocardiogram (TTE) performed within hours after his assessment by the wearable Doppler ultrasound, both diastolic and systolic ventricular interdependence likely contributed to impaired LV output. That is, with RV volume overload noted on the TTE, RV diastolic filling during the PLR diminished LV preload via cross-ventricular diastolic stiffening [45]. Further, given RV pressure overload, systolic ventricular interaction may have additionally impaired the LV. This is especially noteworthy given the dynamic, mechanical asynchrony observed by the wearable Doppler induced by the PLR. With increased preload, the onset of RV systole (i.e., venous ‘s’ wave) occurred significantly before the carotid upstroke. As the carotid waveform temporally reflects LV systole, the interval between the venous ‘s’ wave and carotid upstroke illustrates a dynamic, mechanical interventricular conduction delay (i.e., ‘ivd’, Fig. 2D)—consistent with the patient’s known incomplete left bundle branch block. When this occurs, RV systolic ejection stiffens the interventricular septum prior to LV ejection and generates unfavorable mechanics (including worsened functional mitral regurgitation) for the delayed LV contraction [46, 47]. Indeed, this pathophysiology is the rationale for ventricular resynchronization with biventricular pacing [48]. To our knowledge, this report is the first demonstrating dynamic, mechanical ventricular asynchrony using simultaneous venous and arterial Doppler.
The primary limitation of this short communication is the small, clinical sample size, though this report is not meant to change local clinical practice or protocols. Our objective is to highlight bedside physiology with a novel and potentially helpful clinical biosensor. In both patients neither CVP nor SV was measured, so the data from the wearable Doppler are inferences. Nevertheless, data supporting ccFT as a surrogate for SV have good clinical evidence [49] and we have shown that VTIc correlates strongly with SV in the largest-known Doppler data set making this comparison [14]. Furthermore, in a proof-of-principle description of two patients monitored with trans-esophageal echocardiography, the optimal VTIc for detecting significant SV mirrored that in our healthy volunteer studies [10]. Nevertheless, carotid artery distention with IV fluids could dissociate flow from VTI and carotid flow itself is a surrogate for left ventricular output only when the ratio of total body impedance-to-downstream carotid impedance remains relatively constant [15, 50].

Conclusions

In this short communication, we described the clinical application of a novel, wearable, Doppler biosensor in two patients deemed to need IV fluids by clinical examination. Simultaneous venous and arterial Doppler spectrograms are plotted using a novel framework best characterized as a “Doppler Starling curve”. Both patients initially demonstrated venous Doppler waveforms consistent with low CVP, which increased during preload augmentation. However, only patient 2 exhibited simultaneously acquired arterial Doppler changes indicative of rising stroke volume. By contrast, patient 1 displayed dynamic fluid intolerance; evidence from large, randomized controlled trials suggests that this phenotype may be managed by withholding additional IV fluids, ostensibly preventing downstream complication and cost.

Acknowledgements

Not applicable.

Declarations

Written and informed consent was obtained for both patients included in this report, and the study was approved by the Peoria Institutional Review Board. The study was performed in accordance with the ethical standards as laid down in the 1964 Declaration of Helsinki and its later amendments.
Written and informed consent was obtained for both patients for publication of this report and accompanying images.

Competing interests

JESK, SOG, JKE, AME, ZY, DJ, CM, ME work for Flosonics Medical, the start-up building the wearable Doppler ultrasound. VCL, BOK, RA and BN declare no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Chaudhuri D, Herritt B, Lewis K, Diaz-Gomez JL, Fox-Robichaud A, Ball I et al (2021) Dosing fluids in early septic shock. Chest 159(4):1493–1502PubMedCrossRef Chaudhuri D, Herritt B, Lewis K, Diaz-Gomez JL, Fox-Robichaud A, Ball I et al (2021) Dosing fluids in early septic shock. Chest 159(4):1493–1502PubMedCrossRef
2.
Zurück zum Zitat Cecconi M, Hernandez G, Dunser M, Antonelli M, Baker T, Bakker J et al (2019) Fluid administration for acute circulatory dysfunction using basic monitoring: narrative review and expert panel recommendations from an ESICM task force. Intensive Care Med 45(1):21–32PubMedCrossRef Cecconi M, Hernandez G, Dunser M, Antonelli M, Baker T, Bakker J et al (2019) Fluid administration for acute circulatory dysfunction using basic monitoring: narrative review and expert panel recommendations from an ESICM task force. Intensive Care Med 45(1):21–32PubMedCrossRef
4.
Zurück zum Zitat McGregor D, Sharma S, Gupta S, Ahmad S, Godec T, Harris T (2019) Emergency department non-invasive cardiac output study (EDNICO): a feasibility and repeatability study. Scand J Trauma Resusc Emerg Med 27(1):30PubMedPubMedCentralCrossRef McGregor D, Sharma S, Gupta S, Ahmad S, Godec T, Harris T (2019) Emergency department non-invasive cardiac output study (EDNICO): a feasibility and repeatability study. Scand J Trauma Resusc Emerg Med 27(1):30PubMedPubMedCentralCrossRef
5.
Zurück zum Zitat Marik PE (2016) Fluid responsiveness and the six guiding principles of fluid resuscitation. Crit Care Med 44(10):1920–1922PubMedCrossRef Marik PE (2016) Fluid responsiveness and the six guiding principles of fluid resuscitation. Crit Care Med 44(10):1920–1922PubMedCrossRef
6.
Zurück zum Zitat Kattan E, Ospina-Tascón GA, Teboul J-L, Castro R, Cecconi M, Ferri G et al (2020) Systematic assessment of fluid responsiveness during early septic shock resuscitation: secondary analysis of the ANDROMEDA-SHOCK trial. Crit Care 24(1):23PubMedPubMedCentralCrossRef Kattan E, Ospina-Tascón GA, Teboul J-L, Castro R, Cecconi M, Ferri G et al (2020) Systematic assessment of fluid responsiveness during early septic shock resuscitation: secondary analysis of the ANDROMEDA-SHOCK trial. Crit Care 24(1):23PubMedPubMedCentralCrossRef
7.
Zurück zum Zitat Hernández G, Ospina-Tascón GA, Damiani LP, Estenssoro E, Dubin A, Hurtado J et al (2019) Effect of a resuscitation strategy targeting peripheral perfusion status vs serum lactate levels on 28-day mortality among patients with septic shock: the ANDROMEDA-SHOCK randomized clinical trial. JAMA 321(7):654–664PubMedPubMedCentralCrossRef Hernández G, Ospina-Tascón GA, Damiani LP, Estenssoro E, Dubin A, Hurtado J et al (2019) Effect of a resuscitation strategy targeting peripheral perfusion status vs serum lactate levels on 28-day mortality among patients with septic shock: the ANDROMEDA-SHOCK randomized clinical trial. JAMA 321(7):654–664PubMedPubMedCentralCrossRef
8.
Zurück zum Zitat Kenny JS (2022) Ask not liberal or conservative intravenous fluids in septic shock: ask rather why and when. Scand J Trauma Resusc Emerg Med 30(1):63PubMedPubMedCentralCrossRef Kenny JS (2022) Ask not liberal or conservative intravenous fluids in septic shock: ask rather why and when. Scand J Trauma Resusc Emerg Med 30(1):63PubMedPubMedCentralCrossRef
9.
Zurück zum Zitat Douglas IS, Alapat PM, Corl KA, Exline MC, Forni LG, Holder AL et al (2020) Fluid response evaluation in sepsis hypotension and shock: a randomized clinical trial. Chest 158(4):1431–1445PubMedPubMedCentralCrossRef Douglas IS, Alapat PM, Corl KA, Exline MC, Forni LG, Holder AL et al (2020) Fluid response evaluation in sepsis hypotension and shock: a randomized clinical trial. Chest 158(4):1431–1445PubMedPubMedCentralCrossRef
10.
Zurück zum Zitat Kenny JS, Clarke G, Myers M, Elfarnawany M, Eibl AM, Eibl JK et al (2021) A wireless wearable doppler ultrasound detects changing stroke volume: proof-of-principle comparison with trans-esophageal echocardiography during coronary bypass surgery. Bioengineering (Basel) 8(12), (2021). https://doi.org/10.3390/bioengineering8120203https://doi.org/10.3390/bioengineering8120203. PMC8698882 Kenny JS, Clarke G, Myers M, Elfarnawany M, Eibl AM, Eibl JK et al (2021) A wireless wearable doppler ultrasound detects changing stroke volume: proof-of-principle comparison with trans-esophageal echocardiography during coronary bypass surgery. Bioengineering (Basel) 8(12), (2021). https://​doi.​org/​10.​3390/​bioengineering81​20203https://doi.org/10.3390/bioengineering8120203. PMC8698882
11.
Zurück zum Zitat Kenny J-ÉS, Munding CE, Eibl JK, Eibl AM, Long BF, Boyes A et al (2021) A novel, hands-free ultrasound patch for continuous monitoring of quantitative Doppler in the carotid artery. Sci Rep 11(1):1–11CrossRef Kenny J-ÉS, Munding CE, Eibl JK, Eibl AM, Long BF, Boyes A et al (2021) A novel, hands-free ultrasound patch for continuous monitoring of quantitative Doppler in the carotid artery. Sci Rep 11(1):1–11CrossRef
13.
Zurück zum Zitat Kenny J-ÉS (2021) Functional hemodynamic monitoring with a wireless ultrasound patch. J Cardiothorac Vasc Anesth 35(5):1509–1515PubMedCrossRef Kenny J-ÉS (2021) Functional hemodynamic monitoring with a wireless ultrasound patch. J Cardiothorac Vasc Anesth 35(5):1509–1515PubMedCrossRef
14.
Zurück zum Zitat Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Elfarnawany M, Yang Z, Eibl AM et al (2021) Carotid Doppler ultrasonography correlates with stroke volume in a human model of hypovolaemia and resuscitation: analysis of 48 570 cardiac cycles. Br J Anaesth 127(2):e60–e63PubMedCrossRef Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Elfarnawany M, Yang Z, Eibl AM et al (2021) Carotid Doppler ultrasonography correlates with stroke volume in a human model of hypovolaemia and resuscitation: analysis of 48 570 cardiac cycles. Br J Anaesth 127(2):e60–e63PubMedCrossRef
15.
Zurück zum Zitat Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Elfarnawany M, Yang Z, Eibl AM et al (2022) Carotid artery velocity time integral and corrected flow time measured by a wearable Doppler ultrasound detect stroke volume rise from simulated hemorrhage to transfusion. BMC Res Notes 15(1):7PubMedPubMedCentralCrossRef Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Elfarnawany M, Yang Z, Eibl AM et al (2022) Carotid artery velocity time integral and corrected flow time measured by a wearable Doppler ultrasound detect stroke volume rise from simulated hemorrhage to transfusion. BMC Res Notes 15(1):7PubMedPubMedCentralCrossRef
16.
Zurück zum Zitat Kenny JS, Gibbs SO, Johnston D, Yang Z, Hofer LM, Elfarnawany M et al (2023) The time cost of physiologically ineffective intravenous fluids in the emergency department: an observational pilot study employing wearable Doppler ultrasound. J Intensive Care 11(1):7PubMedPubMedCentralCrossRef Kenny JS, Gibbs SO, Johnston D, Yang Z, Hofer LM, Elfarnawany M et al (2023) The time cost of physiologically ineffective intravenous fluids in the emergency department: an observational pilot study employing wearable Doppler ultrasound. J Intensive Care 11(1):7PubMedPubMedCentralCrossRef
17.
Zurück zum Zitat Kenny J-ES, Eibl JK, Mackenzie DC, Barjaktarevic I (2021) Guidance of intravenous fluid by ultrasound will improve with technology. Chest 161(2):132–133CrossRef Kenny J-ES, Eibl JK, Mackenzie DC, Barjaktarevic I (2021) Guidance of intravenous fluid by ultrasound will improve with technology. Chest 161(2):132–133CrossRef
18.
19.
Zurück zum Zitat Kenny JS, Gibbs SO, Johnston D, Hofer LM, Rae E, Clarke G et al (2023) Continuous venous-arterial Doppler ultrasound during a preload challenge. J Vis Exp 10.3791/64410 (191) Kenny JS, Gibbs SO, Johnston D, Hofer LM, Rae E, Clarke G et al (2023) Continuous venous-arterial Doppler ultrasound during a preload challenge. J Vis Exp 10.3791/64410 (191) 
20.
Zurück zum Zitat Kenny J-ES (2022) Assessing fluid intolerance with Doppler ultrasonography: a physiological framework. Med Sci 10(1):12 Kenny J-ES (2022) Assessing fluid intolerance with Doppler ultrasonography: a physiological framework. Med Sci 10(1):12
22.
Zurück zum Zitat Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Eibl AM, Parrotta M, Long BF et al (2020) Diagnostic characteristics of 11 formulae for calculating corrected flow time as measured by a wearable Doppler patch. Intensive Care Med Exp 8(1):1–11CrossRef Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Eibl AM, Parrotta M, Long BF et al (2020) Diagnostic characteristics of 11 formulae for calculating corrected flow time as measured by a wearable Doppler patch. Intensive Care Med Exp 8(1):1–11CrossRef
23.
Zurück zum Zitat Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Elfarnawany M, Math ZYB, Eibl AM et al (2021) Carotid Doppler measurement variability in functional hemodynamic monitoring: an analysis of 17,822 cardiac cycles. Crit Care Explor 3(6):e0439PubMedPubMedCentralCrossRef Kenny J-ÉS, Barjaktarevic I, Mackenzie DC, Elfarnawany M, Math ZYB, Eibl AM et al (2021) Carotid Doppler measurement variability in functional hemodynamic monitoring: an analysis of 17,822 cardiac cycles. Crit Care Explor 3(6):e0439PubMedPubMedCentralCrossRef
24.
Zurück zum Zitat Barjaktarevic I, Toppen WE, Hu S, Montoya EA, Ong S, Buhr R et al (2018) Ultrasound assessment of the change in carotid corrected flow time in fluid responsiveness in undifferentiated shock. Crit Care Med 11:1040–1046CrossRef Barjaktarevic I, Toppen WE, Hu S, Montoya EA, Ong S, Buhr R et al (2018) Ultrasound assessment of the change in carotid corrected flow time in fluid responsiveness in undifferentiated shock. Crit Care Med 11:1040–1046CrossRef
25.
Zurück zum Zitat Kenny J-ES, Prager R, Rola P, McCulloch G, Eibl JK, Haycock K (2023) The effect of gravity-induced preload change on the venous excess ultrasound (VExUS) score and internal jugular vein Doppler in healthy volunteers. Intensive Care Med Exp 11(1):19PubMedPubMedCentralCrossRef Kenny J-ES, Prager R, Rola P, McCulloch G, Eibl JK, Haycock K (2023) The effect of gravity-induced preload change on the venous excess ultrasound (VExUS) score and internal jugular vein Doppler in healthy volunteers. Intensive Care Med Exp 11(1):19PubMedPubMedCentralCrossRef
26.
Zurück zum Zitat Tang WW, Kitai T (2016) Intrarenal venous flow: a window into the congestive kidney failure phenotype of heart failure? JACC Heart Fail 4(8):683–686PubMedCrossRef Tang WW, Kitai T (2016) Intrarenal venous flow: a window into the congestive kidney failure phenotype of heart failure? JACC Heart Fail 4(8):683–686PubMedCrossRef
27.
Zurück zum Zitat Sivaciyan V, Ranganathan N (1978) Transcutaneous doppler jugular venous flow velocity recording. Circulation 57(5):930–939PubMedCrossRef Sivaciyan V, Ranganathan N (1978) Transcutaneous doppler jugular venous flow velocity recording. Circulation 57(5):930–939PubMedCrossRef
28.
Zurück zum Zitat Ranganathan N, Sivaciyan V, Pryszlak M, Freeman MR (1989) Changes in jugular venous flow velocity after coronary artery bypass grafting. Am J Cardiol 63(11):725–729PubMedCrossRef Ranganathan N, Sivaciyan V, Pryszlak M, Freeman MR (1989) Changes in jugular venous flow velocity after coronary artery bypass grafting. Am J Cardiol 63(11):725–729PubMedCrossRef
30.
Zurück zum Zitat Appleton CP, Hatle LK, Popp RL (1987) Superior vena cava and hepatic vein Doppler echocardiography in healthy adults. J Am Coll Cardiol 10(5):1032–1039PubMedCrossRef Appleton CP, Hatle LK, Popp RL (1987) Superior vena cava and hepatic vein Doppler echocardiography in healthy adults. J Am Coll Cardiol 10(5):1032–1039PubMedCrossRef
31.
Zurück zum Zitat Reynolds T, Appleton CP (1991) Doppler flow velocity patterns of the superior vena cava, inferior vena cava, hepatic vein, coronary sinus, and atrial septal defect: a guide for the echocardiographer. J Am Soc Echocardiogr 4(5):503–512PubMedCrossRef Reynolds T, Appleton CP (1991) Doppler flow velocity patterns of the superior vena cava, inferior vena cava, hepatic vein, coronary sinus, and atrial septal defect: a guide for the echocardiographer. J Am Soc Echocardiogr 4(5):503–512PubMedCrossRef
32.
Zurück zum Zitat Abu-Yousef MM (1992) Normal and respiratory variations of the hepatic and portal venous duplex Doppler waveforms with simultaneous electrocardiographic correlation. J Ultrasound Med 11(6):263–268PubMedCrossRef Abu-Yousef MM (1992) Normal and respiratory variations of the hepatic and portal venous duplex Doppler waveforms with simultaneous electrocardiographic correlation. J Ultrasound Med 11(6):263–268PubMedCrossRef
33.
Zurück zum Zitat Abu-Yousef MM, Kakish M, Mufid M (1996) Pulsatile venous Doppler flow in lower limbs: highly indicative of elevated right atrium pressure. AJR Am J Roentgenol 167(4):977–980PubMedCrossRef Abu-Yousef MM, Kakish M, Mufid M (1996) Pulsatile venous Doppler flow in lower limbs: highly indicative of elevated right atrium pressure. AJR Am J Roentgenol 167(4):977–980PubMedCrossRef
34.
Zurück zum Zitat Beaubien-Souligny W, Rola P, Haycock K, Bouchard J, Lamarche Y, Spiegel R et al (2020) Quantifying systemic congestion with Point-Of-Care ultrasound: development of the venous excess ultrasound grading system. Ultrasound J 12(1):1–12CrossRef Beaubien-Souligny W, Rola P, Haycock K, Bouchard J, Lamarche Y, Spiegel R et al (2020) Quantifying systemic congestion with Point-Of-Care ultrasound: development of the venous excess ultrasound grading system. Ultrasound J 12(1):1–12CrossRef
35.
Zurück zum Zitat Muller L, Toumi M, Bousquet PJ, Riu-Poulenc B, Louart G, Candela D et al (2011) An increase in aortic blood flow after an infusion of 100 ml colloid over 1 minute can predict fluid responsiveness: the mini-fluid challenge study. Anesthesiology 115(3):541–547PubMedCrossRef Muller L, Toumi M, Bousquet PJ, Riu-Poulenc B, Louart G, Candela D et al (2011) An increase in aortic blood flow after an infusion of 100 ml colloid over 1 minute can predict fluid responsiveness: the mini-fluid challenge study. Anesthesiology 115(3):541–547PubMedCrossRef
36.
Zurück zum Zitat Barthélémy R, Kindermans M, Delval P, Collet M, Gaugain S, Cecconi M et al (2021) Accuracy of cumulative volumes of fluid challenge to assess fluid responsiveness in critically ill patients with acute circulatory failure: a pharmacodynamic approach. Br J Anaesth 2022 Feb;128(2):236-243. https://doi.org/10.1016/j.bja.2021.10.049. Epub 2021 Dec 8. PMID: 34895718. Barthélémy R, Kindermans M, Delval P, Collet M, Gaugain S, Cecconi M et al (2021) Accuracy of cumulative volumes of fluid challenge to assess fluid responsiveness in critically ill patients with acute circulatory failure: a pharmacodynamic approach. Br J Anaesth 2022 Feb;128(2):236-243. https://​doi.​org/​10.​1016/​j.​bja.​2021.​10.​049. Epub 2021 Dec 8. PMID: 34895718.
37.
Zurück zum Zitat Donahue SP, Wood JP, Patel BM, Quinn JV (2009) Correlation of sonographic measurements of the internal jugular vein with central venous pressure. Am J Emerg Med 27(7):851–855PubMedCrossRef Donahue SP, Wood JP, Patel BM, Quinn JV (2009) Correlation of sonographic measurements of the internal jugular vein with central venous pressure. Am J Emerg Med 27(7):851–855PubMedCrossRef
38.
Zurück zum Zitat Iida N, Seo Y, Sai S, Machino-Ohtsuka T, Yamamoto M, Ishizu T et al (2016) Clinical implications of intrarenal hemodynamic evaluation by Doppler ultrasonography in heart failure. JACC Heart Fail 4(8):674–682PubMedCrossRef Iida N, Seo Y, Sai S, Machino-Ohtsuka T, Yamamoto M, Ishizu T et al (2016) Clinical implications of intrarenal hemodynamic evaluation by Doppler ultrasonography in heart failure. JACC Heart Fail 4(8):674–682PubMedCrossRef
39.
Zurück zum Zitat Magder S, Bafaqeeh F (2007) The clinical role of central venous pressure measurements. J Intensive Care Med 22(1):44–51PubMedCrossRef Magder S, Bafaqeeh F (2007) The clinical role of central venous pressure measurements. J Intensive Care Med 22(1):44–51PubMedCrossRef
40.
Zurück zum Zitat Eskesen T, Wetterslev M, Perner A (2016) Systematic review including re-analyses of 1148 individual data sets of central venous pressure as a predictor of fluid responsiveness. Intensive Care Med 42(3):324–332PubMedCrossRef Eskesen T, Wetterslev M, Perner A (2016) Systematic review including re-analyses of 1148 individual data sets of central venous pressure as a predictor of fluid responsiveness. Intensive Care Med 42(3):324–332PubMedCrossRef
41.
Zurück zum Zitat Muller L, Bobbia X, Toumi M, Louart G, Molinari N, Ragonnet B et al (2012) Respiratory variations of inferior vena cava diameter to predict fluid responsiveness in spontaneously breathing patients with acute circulatory failure: need for a cautious use. Crit Care 16(5):R188PubMedPubMedCentralCrossRef Muller L, Bobbia X, Toumi M, Louart G, Molinari N, Ragonnet B et al (2012) Respiratory variations of inferior vena cava diameter to predict fluid responsiveness in spontaneously breathing patients with acute circulatory failure: need for a cautious use. Crit Care 16(5):R188PubMedPubMedCentralCrossRef
42.
Zurück zum Zitat Satterwhite L, Latham H (2020) Fluid management in sepsis hypotension and septic shock: time to transition the conversation from fluid responsive to fluid refractory? Chest 158(4):1319–1320PubMedCrossRef Satterwhite L, Latham H (2020) Fluid management in sepsis hypotension and septic shock: time to transition the conversation from fluid responsive to fluid refractory? Chest 158(4):1319–1320PubMedCrossRef
43.
44.
Zurück zum Zitat Latham HE, Bengtson CD, Satterwhite L, Stites M, Subramaniam DP, Chen GJ et al (2017) Stroke volume guided resuscitation in severe sepsis and septic shock improves outcomes. J Crit Care 42:42–46PubMedCrossRef Latham HE, Bengtson CD, Satterwhite L, Stites M, Subramaniam DP, Chen GJ et al (2017) Stroke volume guided resuscitation in severe sepsis and septic shock improves outcomes. J Crit Care 42:42–46PubMedCrossRef
45.
Zurück zum Zitat Naeije R, Badagliacca R (2017) The overloaded right heart and ventricular interdependence. Cardiovasc Res 113(12):1474–1485PubMedCrossRef Naeije R, Badagliacca R (2017) The overloaded right heart and ventricular interdependence. Cardiovasc Res 113(12):1474–1485PubMedCrossRef
46.
Zurück zum Zitat Lumens J, Delhaas T, Kirn B, Arts T (2009) Three-wall segment (TriSeg) model describing mechanics and hemodynamics of ventricular interaction. Ann Biomed Eng 37(11):2234–2255PubMedPubMedCentralCrossRef Lumens J, Delhaas T, Kirn B, Arts T (2009) Three-wall segment (TriSeg) model describing mechanics and hemodynamics of ventricular interaction. Ann Biomed Eng 37(11):2234–2255PubMedPubMedCentralCrossRef
47.
Zurück zum Zitat Smiseth OA, Aalen JM (2019) Mechanism of harm from left bundle branch block. Trends Cardiovasc Med 29(6):335–342PubMedCrossRef Smiseth OA, Aalen JM (2019) Mechanism of harm from left bundle branch block. Trends Cardiovasc Med 29(6):335–342PubMedCrossRef
48.
Zurück zum Zitat Lumens J, Ploux S, Strik M, Gorcsan J, Cochet H, Derval N et al (2013) Comparative electromechanical and hemodynamic effects of left ventricular and biventricular pacing in dyssynchronous heart failure: electrical resynchronization versus left–right ventricular interaction. J Am Coll Cardiol 62(25):2395–2403PubMedPubMedCentralCrossRef Lumens J, Ploux S, Strik M, Gorcsan J, Cochet H, Derval N et al (2013) Comparative electromechanical and hemodynamic effects of left ventricular and biventricular pacing in dyssynchronous heart failure: electrical resynchronization versus left–right ventricular interaction. J Am Coll Cardiol 62(25):2395–2403PubMedPubMedCentralCrossRef
49.
Zurück zum Zitat Beier L, Davis J, Esener D, Grant C, Fields JM (2020) Carotid ultrasound to predict fluid responsiveness: a systematic review. J Ultrasound Med 39(10):1965–1976PubMedCrossRef Beier L, Davis J, Esener D, Grant C, Fields JM (2020) Carotid ultrasound to predict fluid responsiveness: a systematic review. J Ultrasound Med 39(10):1965–1976PubMedCrossRef
Metadaten
Titel
Simultaneous venous–arterial Doppler during preload augmentation: illustrating the Doppler Starling curve
verfasst von
Jon-Émile S. Kenny
Stanley O. Gibbs
Joseph K. Eibl
Andrew M. Eibl
Zhen Yang
Delaney Johnston
Chelsea E. Munding
Mai Elfarnawany
Vivian C. Lau
Benjamin O. Kemp
Bhanu Nalla
Rony Atoui
Publikationsdatum
01.12.2023
Verlag
Springer International Publishing
Erschienen in
The Ultrasound Journal / Ausgabe 1/2023
Print ISSN: 2036-3176
Elektronische ISSN: 2524-8987
DOI
https://doi.org/10.1186/s13089-023-00330-9

Weitere Artikel der Ausgabe 1/2023

The Ultrasound Journal 1/2023 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Erhebliches Risiko für Kehlkopfkrebs bei mäßiger Dysplasie

29.05.2024 Larynxkarzinom Nachrichten

Fast ein Viertel der Personen mit mäßig dysplastischen Stimmlippenläsionen entwickelt einen Kehlkopftumor. Solche Personen benötigen daher eine besonders enge ärztliche Überwachung.

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

15% bedauern gewählte Blasenkrebs-Therapie

29.05.2024 Urothelkarzinom Nachrichten

Ob Patienten und Patientinnen mit neu diagnostiziertem Blasenkrebs ein Jahr später Bedauern über die Therapieentscheidung empfinden, wird einer Studie aus England zufolge von der Radikalität und dem Erfolg des Eingriffs beeinflusst.

Costims – das nächste heiße Ding in der Krebstherapie?

28.05.2024 Onkologische Immuntherapie Nachrichten

„Kalte“ Tumoren werden heiß – CD28-kostimulatorische Antikörper sollen dies ermöglichen. Am besten könnten diese in Kombination mit BiTEs und Checkpointhemmern wirken. Erste klinische Studien laufen bereits.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.