Safety and image quality of cardiovascular magnetic resonance imaging in patients with retained epicardial pacing wires after heart transplantation

verfasst von: Constantin Gatterer, Marie-Elisabeth Stelzmüller, Andreas Kammerlander, Andreas Zuckermann, Martin Krššák, Christian Loewe, Dietrich Beitzke

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2021

Abstract

Background

Temporary epicardial pacing wires, implemented in patients during heart transplantation, are routinely removed before discharge. However, in some cases, these wires may remain in situ and are often considered as a contraindication for cardiovascular magnetic resonance (CMR) imaging in the future. Therefore, we aimed to provide data about safety and image quality of CMR in these patients.

Methods

This is a report on a subpopulation out of 88 patients after heart transplantation that were included in a prospective cohort study and underwent multiple CMR in their post-transplant course. During CMR, patients were monitored by electrocardiogram and all examinations were observed by a physician to document potential adverse events. Additionally, image quality was assessed by an imaging specialist.

Results

Nineteen of 88 patients included had temporary pacing wires in situ. These patients underwent a total of 51 CMR studies. No major adverse event and only one single, mild sensory event could be documented. All CMR studies showed preserved diagnostic image quality. Temporary pacing wires were visible in 100% of HASTE and cine sequences. In less than 50% of the examinations, temporary pacing wires were also visible in T1 and T2 mapping, short tau inversion recovery (STIR), and late gadolinium enhancement (LGE) sequences, without any impairment of image quality.

Conclusions

With a low event rate of only one mild adverse event during 51 CMR examinations (2%), CMR appears to be safe in patients with retained temporary epicardial pacing wires after heart transplantation. Moreover, image quality was not impaired by the presence of pacing wires.
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
2Ch
Two chamber
3Ch
Three chamber
4Ch
Four chamber
CMR
Cardiovascular magnetic resonance
CT
Computed tomography
ECG
Electrocardiogram
HASTE
Half Fourier single-shot turbo spin-echo
ICD
Implanted cardiodefibrillator
LGE
Late gadolinium enhancement
MRI
Magnetic resonance imaging
PM
Pacemaker
PPM
Permanent pacemaker
RA
Right atrium/right atrial
RF
Radiofrequency
RV
Right ventricle/right ventricular
SAR
Specific absorption rate
SAx
Short axis
STIR
Short tau inversion recovery

Introduction

Patients after heart transplantation are at high risk for post-operative bradyarrhythmia, including atrioventricular block and sinus node dysfunction [1]. Thus, temporary pacing wires are routinely inserted in the right atrium (RA) and right ventricle (RV) after successful implantation of the donor’s heart. Furthermore, intermittent temporal pacing can facilitate the patients’ weaning process from cardiopulmonary bypass [2]. During the postoperative course, these wires are intended to be completely removed. However, in some cases, the wires have to be fixed at the myocardium (e.g. fixation by a suture to stop bleeding at the time of the insertion) or cannot be removed due to pericardial adhesion or overlap with sutures of the fascia. In these cases, pacing wires are usually cut below the skin level.
With these wires attached to the heart, patients are often excluded from magnetic resonance imaging (MRI) in the future [3]. As MRI utilization continuously increases and has a substantial impact on patient care, excluding these patients from MRI might potentially hinder a diagnostic work-up in the future [4, 5].
Causes for concern relate to potential neuromuscular voltage induction that could result in arrhythmias and movement, or heating of pacing wires with consequent tissue damage [68]. Preclinical data indicates that the greatest risk arises from the radiofrequency (RF) field due to the circularly polarized magnetic field associated with MRI devices. Higher amounts of energy may be deposited in the patient’s body by RF via abandoned pacemaker (PM) leads or pacing wires, resulting in heat burns and potentially inducing arrhythmic events [8, 9]. However, even using higher systemic absorption rates ex vivo, no myocardial damage could be observed so far [10]. Also there are several studies on patients undergoing MRI with retained temporary pacing wires published the that do not report any adverse events [3, 1114]. However, the MR systems and scan regions in these reports were quite heterogeneous, and thus, did not allow a systematic comparison on this topic. To date, data about the safety of cardiovascular magnetic resonance (CMR) have been described only in a case report [12].
MRI and CMR with perrmanent pacemaker (PPM) systems including subcutaneous PPM generators is already a well-established procedure and as nowadays most of these systems are approved for CMR/MRI under specified conditions by the vendors [15]. In contrary to permanent systems temporary wires did not undergo systematic testing for being MR conditional as these “temporary” pacing wires are not intended to remain in the patient. Additionally, temporary pacing wires are not implanted in a standardized way, and therefore the shape and the length of the abandoned wires differ from patient to patient making it impossible to predict potential CMR interaction in the further course.
Another important aspect of CMR in contrast to other imaging regions is image quality when dealing with temporary pacing wires and conventional PPM remnants, as leads can significantly alter image quality due to artifacts [16]. Especially in CMR, these artifacts may conceal pathologies and could potentially influence diagnostic image quality, as well as quantitative values of mapping techniques.
Therefore, the aim of this study was to provide further evidence about the safety and image quality in a patient cohort with temporary pacing wires who were referred to multiparametric CMR in a prospective setting.

Methods

Patients:
This subgroup analysis is part of an ongoing, prospective CMR cohort study of 100 patients after heart transplantation. It complies with the declaration of Helsinki and its amendments and was approved by the ethics committee of the Medical University of Vienna (IRB Nr.: 2012/2015). The inclusion criteria were: patients 18–99 years; and recent cardiac transplantation. Exclusion criteria were defined as the presence of abandoned implanted cardiodefibrilator (ICD) leads, claustrophobia, and pregnancy. Written, informed consent was obtained from each participating patient prior to study inclusion. The study was performed at a tertiary referral center that offered the entire spectrum of cardiovascular imaging and an organ transplant program, as part of EuroTransplant, which performs approximately 50 to 60 heart transplantations annually. Within this prospective cohort study, patients are scheduled for CMR 3, 12, 24, 36, 48 and 60 months after heart transplantation.
During heart transplantation surgery, temporary pacing wires (TME bifurcated, OSYPKA AG, Rheinfelden, Germany) were implanted at the RA and RV. The leads were then loosely folded to a loop and led intrapericardially to the xiphoid where they penetrated the skin to be connected to the extracorporeal PM. Epicardial wires that could not be removed were left in situ after being cut below the skin level. Patients who participated in this prospective cohort study and had retained temporary pacing wires were briefed about reporting any skin perceptions or heat sensations during the CMR examination.
Using routinely performed chest X-ray, computed tomography, and fluoroscopy, retained epicardial pacing wires were categorized according to their shape into “straight”, “C-shaped,” or “looped (antenna).”
All CMR examinations were performed on a 1.5T CMR scanner (Avanto Fit, Siemens Healthineers, Erlangen, Germany) with a maximum gradient strength of 45 mT/m and a maximum slew rate of 200 T/m/s. Specific absorption rate (SAR) limits for normal operation mode were 2w/kg and 4w/kg for first level mode, respectively. Two receive only coils (18-channel body array/32-channel integrated coil) were used. A standardized multiparametric CMR protocol was used to assess myocardial inflammation. The protocol included left ventricular structure and function, advanced tissue characterization by T1 and T2 mapping, conventional edema imaging, and scar assessment by late gadolinium enhancement (LGE) [17]. Details on the CMR protocol are listed in Table 1.
Table 1
Cardiovascular Magnetic Resonance (CMR) sequence pre-settings at 1.5T
Sequence
TI (ms)
TE/TR (ms)
Triggerpulse
Echo spacing (ms)
Flip angle (°)
Resolution (mm)
Layer-thickness (mm)
Readout bandwith (Hz/Pixel)
bSSFP Cine
(SAx, 4Ch, 2Ch)
 
1.21/52.2
 
2.9
64
1.4 × 1.4
6
930
T2 HASTE
 
43/467
x2
3.32
160
1.6 × 1.6
5
781
T1-mapping (SAx,4ch)
RR > 700ms)
180
1.13/279.84
 
2.7
35
1.4 × 1.4
8
1085
T1-mapping (SAx/4Ch)
*RR < 700ms
180
1.01/263.55
 
2.4
35
1.9 × 1.9
8
1085
T2-mapping
 
27/193.27
 
2.5
70
1.9 × 1.9
8
1184
STIR
150
47/760
x2
6.74
180
1.5 × 1.5
8
235
Rest Perfusion (SAx)
115
1.07/193.72
 
2.2
    
LGE (SAx,4Ch,3Ch, 2Ch)
(250)
variable
1.19/700
 
2.9
50
1.5 × 1.5
8
781
Post-contrast T1-mapping (SAx,4Ch)
*RR > 700
260
1.13/359.84
 
2.68
35
1.4 × 1.4
8
1085
Post-contrast T1-mapping (SAx,4Ch)
RR < 700
260
1.01/241.12
 
2.43
35
1.9 × 1.9
8
1085
*Heart rate adapted; bSSFP, balanced steady state free precession; SAx, short axis; 4Ch, four chamber; 3Ch, three chamber; 2Ch, two chamber; RR, RR-Interval (ECG); STIR, short tau inversion recovery; LGE, late gadolinium enhancement; TI, inversion time; TE, repetition time; TR, repetition time
All examinations were supervised by a physician and possible adverse reactions were documented. Patients were monitored using the CMR integrated electrocardiogram (ECG) gating device. Severe adverse events were defined as arrhythmias, self-reported heat sensations, hemodynamic instability, or any other possibly life-threatening complication. Mild adverse events were defined as every other event that did not meet the criteria for a severe adverse event and was clinically associated with a potential lead/CMR interaction. Furthermore, each CMR examination was assessed for image quality by two CMR specialists.
Statistics:
Descriptive statistics were performed for patient characteristics, image quality, and events. Continuous variables were reported as mean ± standard deviation if normally distributed, or as median ± interquartile range for non-normally distributed continuous variables. Categorical variables were described as counts and percentages. Cohen’s kappa was used to report interrater variability of image quality and visibility between two blinded readers.

Results

Patient characteristics

Overall, 88 patients were included in this ongoing prospective cohort study by July 2020. Nineteen of these 88 patients (21.6%) had retained temporary pacing wires in situ (42% female/ 58% male). Eleven patients had both the RA and the RV pacing wires in situ, eight patients only had the RA wire in situ. The reasons for heart transplantation in this sub-group were ischemic cardiomyopathy (n = 7), dilative cardiomyopathy (n = 9), Anderson-Fabry disease (n = 1), amyloidosis (n = 1), and valvular disease (n = 1). Most patients had repetitive CMR examinations (5 patient had four, 6 patients had three and 5 had two examinations respectively). One patient only had a single CMR examination at the time of analysis. Overall, a total of 51 CMR examinations were analyzed for safety and image quality. The median time from heart transplantation until the first CMR was 96 days (interquartile range: 88–119 days).
In one patient, the short tau inversion recovery (STIR) sequence was not added to the protocol. In another patient, LGE and post-contrast T1 mapping sequences were interrupted due to the onset of an allergic reaction to the contrast agent. All other scans were completed according to the study protocol.

Visibility of temporary pacing wires and CMR image quality

Using chest X-ray, temporary pacing wires leads were visible in only 10 of 19 patients (53%). In eight patients, the leads were detected by chest or cardiac computed tomography (CT) scans, while, in one patient, the existence of retained wires was only self-reported, but could not be verified on any radiological images prior to CMR.
The most frequent lead form, derived from other imaging modalities, was a loop (47%), followed by C-shaped (32%), and straight (21%). An overview of lead forms is displayed in Table 2. Figure 1 shows a representative cases of a C-shaped and looped retained pacing wire, hardly visible on chest X-Ray.
Table 2
Overview of patient characteristics, frequency and form of retained temporary pacemakers leads, and events
Characteristic
n = 19 (%)
Retained temporary PM leads, n (%)
19/88* (22)
Age, year
54 ± 15
Male patients, n (%)
11 (58)
CMR Studies
51 (91)
Form
Loop
9 (47)
C-shaped
6 (32)
Straight
4 (21)
Adverse reactions
Severe
0 (0)
Mild
1 (5)
 Event rate
CMR, n (%)
1/51 (2)
Numbers are given as mean ± standard deviation or count and (%)
*Total heart transplantation CMR study cohort; PM, pacemaker
Image quality was of diagnostic quality in all CMR studies. Although retained pacing wires were visible in all half Fourier single-shot turbo spin-echo (HASTE) sequences and 78% of cine sequences during CMR, diagnostic image quality was not degraded. Image quality in T1/T2 mapping sequences, and LGE images was also not impaired, as the wires could be detected in only 45% and 40%, respectively. A detailed overview of image quality and artifacts can be found in Table 3. Examples of image quality and artifacts are shown in Fig. 2.
Table 3
Visibility and image quality of temporary Pacemaker leads in CMR
Grading of image quality
n = 51(%)
n = 51(%)
n = 51(%)
n = 50(%)
n = 48(%)
HASTE
Cine
T1 /T2 mapping
STIR
LGE
Not visible
0 (0)
11 (22)
23 (45)
18 (36)
14 (29)
Visible, no image quality impairment
51 (100)
40 (78)
23 (45)
23 (46)
19 (40)
Visible, impaired image quality
0 (0)
0 (0)
0 (0)
0 (0)
0 (0)
Not in field of view
0 (0)
0 (0)
5 (10)
9 (18)
15 (31)
Numbers are given as count and (%). HASTE, half-Fourier single shot turbo spin echo, LGE, late gadolinium enhancement; STIR, short tau inversion recovery
A good interrater agreement was found for all measures, ranging from 68% for mapping sequences to 82% for cine imaging (kappa from 0.46 to 0.57, p < 0.001 for all). Both readers agreed that leads did not impair image quality in any of the cases. An overview of agreement and kappa can be found in Table 4.
Table 4
Cohen’s Kappa for interrater variability of image quality 
Measure
Agreement
Kappa
p
Cine
82%
0.48
< 0.001
Maps
68%
0.46
< 0.001
STIR
80%
0.57
< 0.001
LGE
71%
0.52
< 0.001

Safety/adverse events

During 51 CMR examinations, no serious adverse event, such as arrhythmias, self-reported heating of the pacing wires, or severe pain, could be observed. One patient, with a C-shaped temporary pacing wire in situ, described a sensory event near the subcutaneous end of the retained lead during the second CMR, 12 months after heart transplantation. This event occurred during the HASTE sequence. The effect was reproducible, as it immediately disappeared once the sequence was discontinued, but returned as soon as it was restarted. No arrhythmic event or signs of skin irritation were observed. The CMR scan was stopped and the patient was transferred out of the CMR unit. The further course was uneventful; however, the patient was excluded from the study. All other CMR studies were completed uneventfully. With only one event of 51 CMR, the event rate was 2%.

Discussion

The main finding of this study is that multiparametric CMR at 1.5T in patients with retained temporary pacing wires is associated with a low incidence rate of 2% mild adverse events. During a total number of 51 CMR scans in 19 patients, not a single serious adverse event was observed. Furthermore, the image quality of CMR is not altered by retained temporary pacing wires.
MRI and, in particular, CMR utilization is continuously rising, with an increase in use and an increasing number of clinical indications for all patient groups, not only in patients after cardiac transplantation [18, 19]. The use of MRI is reported to have an important impact on clinical decision-making, with added value for patient management [4].
Patients after cardiac surgery and, especially after cardiac transplantation, often present with metallic remnants, either after incomplete ICD lead extraction or with temporary pacing wires as in our cohort [20]. As a consequence, patients with retained wires are often prohibited from receiving MRI, thus possibly facing worse outcomes [3, 21]. Retained temporary pacing wires do not represent a regular medical implant, as the leads are intended to be removed after the operation. Therefore, these pacing wires are not tested for MR compatibility in a defined MR environment. In addition, the form of the wires within their epicardial course varies between patients, making it difficult to exactly predict MR-related complications. A further cause of concern may be the positioning of the patient with the wire in the center of the gradient and radiofrequency field as necessary in CMR. Thus the induction of heating and current might be more likely as if compared to other scan regions [22].
Overall, potential risks arise from interactions with multiple components of the MR system that should be taken into account separately prior to imaging in these patients. The static magnetic field (B0) may interfere with ferromagnetic implants by inducing translational force and torque. However according to the manufacturer the wires used in our cohort do not contain ferromagnetic components [23].
Relevant heating may be caused by the RF field resulting in thermal injury [23]. Kappus et al. showed that, in contrast to transvenous ICD and pacemaker leads, for which MR related heating up to 53.6°C was described ex vivo, no relevant heating of looped and/or straight temporary pacing wires could be found by in vitro testing using an infrared camera [9, 24]. Furthermore, an ex vivo CMR study of 12 swine hearts with implanted pacing wires provided no evidence of thermal damage to the myocardium, as evaluated in the histological examination after the CMR examination [10]. In this context it is also important that only a receive coil is used for standard CMR as the potential for heat induction over the wire is higher using a transmit/receive coil located close to the imaging center [23].
The time varying gradient magnetic field may be associated with (minor) heating of implants and neuromuscular excitation that may present as arrhythmic cardiac events [23]. Lead length and configuration, amplitude and phase of the electric field along the lead influence the current, which is potentially induced into the wires and eventually could lead to current induction, as well as additional heat elevation of the retained leads [8, 25, 26]. This theoretical risk of arrhythmia induction might even be higher if the retained wire is still connected to the RV and/or in case of a cardiac structural abnormality like a scar formation [23]. However, there are no reports available on arrhythmia induction by MR in the presence of retained pacing wires. Additionally, it may also be debatably whether a routine CMR protocol is able to induce voltages beyond the pacing thresholds for the atrial and ventricular leads respectively [27, 28]. These pacing thresholds are reported to increase over the time and therefore a (C)MR far after implantation may pose a lower risk for the patient [27].
Our findings are in line with previous reports that MRI and CMR with temporary pacing wires does not lead to serious lead related complications, thus supporting the hypothesis that CMR is safe in patients after heart surgery [3, 1114, 29].
To date, only one case of an adverse reaction during CMR in this context has been reported in the literature. The patient, who experienced anginal symptoms during stress CMR, fully recovered and his symptoms were thought to be primarily caused by his underlying ischemic cardiomyopathy [13]. Therefore, and based on the data from this cohort, MRI, and particularly CMR, can safely be performed despite the presence of temporary right atrial and right ventricular pacing wires, with very low risk for these patients [30].
There are multiple case studies published, that describe non-MRI-related migration of the wires into various regions, including the right coronary artery, which caused an infarct, and into the ascending aorta [31, 32]. In addition, infectious complications, such as breast abscess formations, have also been reported [33].
Although various approaches have been pursued by different centers to detect retained wires before an MRI/CMR scan, including screening of previous radiological studies, or reviewing the patient’s history, in our cohort, multiple chest x-ray images in the anterior posterior view were obtained during the patients’ treatment at the intensive care unit and during routine follow up. However, in 53%, retained wires could be detected only on CT or fluoroscopy images.
When analyzing image quality, the impact of retained pacing wires on CMR image quality was negligible. No relevant artifacts could be observed, as the wires neither concealed any pathology nor hampered the use of mapping techniques or post-contrast images. In contrast to residual transvenous PPM and ICD leads, retained temporary pacing wires did not result in any relevant impairment of image quality [34, 35]. Interrater agreement was good, an excellent agreement was probably not reached as the artifacts caused by the pacing wires could often hardly be detected.
Limitations:
There are limitations of our study that hinder the transition of these results to general radiology. As all CMR examinations were performed on a 1.5T according to a standardized protocol, results cannot be easily extrapolated to other protocols that include sequences with a higher risk of wire-heating and voltage induction (e.g., diffusion-weighted imaging, prolonged turbo spin echo sequences), other scan areas with different coil settings, or other field strengths. Other limitations were the relatively small sample size and the fact that ECG rhythms were observed for arrhythmias during CMR, but could not be systematically analyzed subsequently. However, most of the patients in this analysis underwent multiple uneventful CMR.

Conclusions

Repetitive CMR at 1.5T in patients with retained temporary right atrial and right ventricular pacing wires after heart transplantation or cardiac surgery appears to be safe and delivers preserved image quality. Therefore, CMR at 1.5T should be offered to these patients without restrictions.

Acknowledgements

Not applicable.
The study was approved by the institutional review board of the Medical University of Vienna (IRB Nr.: 2012/2015). All the subjects included gave written informed consent prior to study inclusion.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Reade MC. Temporary epicardial pacing after cardiac surgery: a practical review: Part 2: Selection of epicardial pacing modes and troubleshooting. Anaesthesia. 2007;62:364–73.CrossRef Reade MC. Temporary epicardial pacing after cardiac surgery: a practical review: Part 2: Selection of epicardial pacing modes and troubleshooting. Anaesthesia. 2007;62:364–73.CrossRef
2.
Zurück zum Zitat Elmistekawy E, Gee YY, Une D, Lemay M, Stolarik A, Rubens FD. Clinical and mechanical factors associated with the removal of temporary epicardial pacemaker wires after cardiac surgery. J Cardiothorac Surg. 2016;11:4–8.CrossRef Elmistekawy E, Gee YY, Une D, Lemay M, Stolarik A, Rubens FD. Clinical and mechanical factors associated with the removal of temporary epicardial pacemaker wires after cardiac surgery. J Cardiothorac Surg. 2016;11:4–8.CrossRef
3.
Zurück zum Zitat Ferris NJ, Kavnoudias H, Thiel C, Stuckey S. The 2005 Australian MRI safety survey. Am J Roentgenol. 2007;188:1388–94.CrossRef Ferris NJ, Kavnoudias H, Thiel C, Stuckey S. The 2005 Australian MRI safety survey. Am J Roentgenol. 2007;188:1388–94.CrossRef
4.
Zurück zum Zitat Samar H, Yamrozik JA, Williams RB, Doyle M, Shah M, Bonnet CA, et al. Diagnostic value of MRI in patients with implanted pacemakers and implantable cardioverter-defibrillators across a cross population: does the benefit justify the risk? A proof of concept study. JACC Clin Electrophysiol. 2017;3:991–1002.CrossRef Samar H, Yamrozik JA, Williams RB, Doyle M, Shah M, Bonnet CA, et al. Diagnostic value of MRI in patients with implanted pacemakers and implantable cardioverter-defibrillators across a cross population: does the benefit justify the risk? A proof of concept study. JACC Clin Electrophysiol. 2017;3:991–1002.CrossRef
5.
Zurück zum Zitat Leiner T, Bogaert J, Friedrich MG, Mohiaddin R, Muthurangu V, Myerson S, et al. (2020) on clinical indications for cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2020;22:76.CrossRef Leiner T, Bogaert J, Friedrich MG, Mohiaddin R, Muthurangu V, Myerson S, et al. (2020) on clinical indications for cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2020;22:76.CrossRef
6.
Zurück zum Zitat Erlebacher JA, Cahill PT, Pannizzo F, Knowles RJR. Effect of magnetic resonance imaging on DDD pacemakers. Am J Cardiol. 1986;12:8. Erlebacher JA, Cahill PT, Pannizzo F, Knowles RJR. Effect of magnetic resonance imaging on DDD pacemakers. Am J Cardiol. 1986;12:8.
7.
Zurück zum Zitat Kanal E. Safety of MR imaging in patients with retained epicardial pacer wires. Am J Roentgenol. 1998;170:213–4.CrossRef Kanal E. Safety of MR imaging in patients with retained epicardial pacer wires. Am J Roentgenol. 1998;170:213–4.CrossRef
8.
Zurück zum Zitat Luechinger R, Zeijlemaker VA, Pedersen EM, Mortensen P, Falk E, Duru F, et al. In vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J. 2005;26:376–83.CrossRef Luechinger R, Zeijlemaker VA, Pedersen EM, Mortensen P, Falk E, Duru F, et al. In vivo heating of pacemaker leads during magnetic resonance imaging. Eur Heart J. 2005;26:376–83.CrossRef
9.
Zurück zum Zitat Langman DA, Goldberg IB, Finn JP, Ennis DB. Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 1.5 tesla MRI. J Magn Reson Imaging. 2011;33:426–31.CrossRef Langman DA, Goldberg IB, Finn JP, Ennis DB. Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 1.5 tesla MRI. J Magn Reson Imaging. 2011;33:426–31.CrossRef
10.
Zurück zum Zitat Pfeil A, Drobnik S, Rzanny R, Aboud A, Böttcher J, Schmidt P, et al. Compatibility of temporary pacemaker myocardial pacing leads with magnetic resonance imaging: an ex vivo tissue study. Int J Cardiovasc Imaging Springer. 2012;28:317–26.CrossRef Pfeil A, Drobnik S, Rzanny R, Aboud A, Böttcher J, Schmidt P, et al. Compatibility of temporary pacemaker myocardial pacing leads with magnetic resonance imaging: an ex vivo tissue study. Int J Cardiovasc Imaging Springer. 2012;28:317–26.CrossRef
11.
Zurück zum Zitat Blissett S, Chetrit M, Kovacina B, Mardigyan V, Afilalo J. Performing cardiac magnetic resonance imaging in patients with cardiac implantable electronic devices: a contemporary review. Can J Cardiol Canadian Cardiovascular Society. 2018;34:1682–6.CrossRef Blissett S, Chetrit M, Kovacina B, Mardigyan V, Afilalo J. Performing cardiac magnetic resonance imaging in patients with cardiac implantable electronic devices: a contemporary review. Can J Cardiol Canadian Cardiovascular Society. 2018;34:1682–6.CrossRef
12.
Zurück zum Zitat Luetkens JA, Isaak A, Naehle CP, Dabir D, Thomas D. Extended cardiac magnetic resonance imaging with retained temporary transmyocardial pacing lead. Int J Cardiovasc Imaging Springer Netherlands. 2019;35:663–4.CrossRef Luetkens JA, Isaak A, Naehle CP, Dabir D, Thomas D. Extended cardiac magnetic resonance imaging with retained temporary transmyocardial pacing lead. Int J Cardiovasc Imaging Springer Netherlands. 2019;35:663–4.CrossRef
13.
Zurück zum Zitat Hartnell GG, Spence L, Hughes LA, Cohen MC, Saouaf R, Buff B. Safety of MR imaging in patients who have retained metallic materials after cardiac surgery. Am J Roentgenol. 1997;168:1157–9.CrossRef Hartnell GG, Spence L, Hughes LA, Cohen MC, Saouaf R, Buff B. Safety of MR imaging in patients who have retained metallic materials after cardiac surgery. Am J Roentgenol. 1997;168:1157–9.CrossRef
14.
Zurück zum Zitat Austin CO, Landolfo K, Parikh PP, Patel PC, Venkatachalam KL, Kusumoto FM. Retained cardiac implantable electronic device fragments are not associated with magnetic resonance imaging safety issues, morbidity, or mortality after orthotopic heart transplant. Am Heart J. 2017;190:46–53.CrossRef Austin CO, Landolfo K, Parikh PP, Patel PC, Venkatachalam KL, Kusumoto FM. Retained cardiac implantable electronic device fragments are not associated with magnetic resonance imaging safety issues, morbidity, or mortality after orthotopic heart transplant. Am Heart J. 2017;190:46–53.CrossRef
15.
Zurück zum Zitat Kalb B, Indik JH, Ott P, Martin DR. MRI of patients with implanted cardiac devices. J Magn Reson Imaging. 2018;47:595–603.CrossRef Kalb B, Indik JH, Ott P, Martin DR. MRI of patients with implanted cardiac devices. J Magn Reson Imaging. 2018;47:595–603.CrossRef
16.
Zurück zum Zitat Rajiah P, Kay F, Bolen M, Patel AR, Landeras L. Cardiac magnetic resonance in patients with cardiac implantable electronic devices: challenges and solutions. J Thorac Imaging. 2020;35:W1-17.CrossRef Rajiah P, Kay F, Bolen M, Patel AR, Landeras L. Cardiac magnetic resonance in patients with cardiac implantable electronic devices: challenges and solutions. J Thorac Imaging. 2020;35:W1-17.CrossRef
17.
Zurück zum Zitat Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson. 2020;22:1–18.CrossRef Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson. 2020;22:1–18.CrossRef
18.
Zurück zum Zitat Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. PACE. 2005;5:326–8.CrossRef Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. PACE. 2005;5:326–8.CrossRef
19.
Zurück zum Zitat Von Knobelsdorff-Brenkenhoff F, Schulz-Menger J. Role of cardiovascular magnetic resonance in the guidelines of the European Society of Cardiology. J Cardiovasc Magn Reson. 2016;18:1–18. Von Knobelsdorff-Brenkenhoff F, Schulz-Menger J. Role of cardiovascular magnetic resonance in the guidelines of the European Society of Cardiology. J Cardiovasc Magn Reson. 2016;18:1–18.
20.
Zurück zum Zitat Kim J, Hwang J, Choi JH, Choi HI, Kim MS, Jung SH, et al. Frequency and clinical impact of retained implantable cardioverter defibrillator lead materials in heart transplant recipients. PLoS ONE. 2017;12:1–9. Kim J, Hwang J, Choi JH, Choi HI, Kim MS, Jung SH, et al. Frequency and clinical impact of retained implantable cardioverter defibrillator lead materials in heart transplant recipients. PLoS ONE. 2017;12:1–9.
21.
Zurück zum Zitat Vermes E, Pantaléon C, Auvet A, Cazeneuve N, Machet MC, Delhommais A, et al. Cardiovascular magnetic resonance in heart transplant patients: diagnostic value of quantitative tissue markers: T2 mapping and extracellular volume fraction, for acute rejection diagnosis. J Cardiovasc Magn Reson. 2018;20:1–11.CrossRef Vermes E, Pantaléon C, Auvet A, Cazeneuve N, Machet MC, Delhommais A, et al. Cardiovascular magnetic resonance in heart transplant patients: diagnostic value of quantitative tissue markers: T2 mapping and extracellular volume fraction, for acute rejection diagnosis. J Cardiovasc Magn Reson. 2018;20:1–11.CrossRef
22.
Zurück zum Zitat Homsi R, Mellert F, Luechinger R, Thomas D, Doerner J, Luetkens J, et al. Safety and feasibility of magnetic resonance imaging of the brain at 15 tesla in patients with temporary transmyocardial pacing leads. Thorac Cardiovasc Surg. 2019;67:86–91.CrossRef Homsi R, Mellert F, Luechinger R, Thomas D, Doerner J, Luetkens J, et al. Safety and feasibility of magnetic resonance imaging of the brain at 15 tesla in patients with temporary transmyocardial pacing leads. Thorac Cardiovasc Surg. 2019;67:86–91.CrossRef
23.
Zurück zum Zitat Kanal E. Standardized approaches to mr safety assessment of patients with implanted devices. Magn Reson Imaging Clin N Am. 2020;7:89. Kanal E. Standardized approaches to mr safety assessment of patients with implanted devices. Magn Reson Imaging Clin N Am. 2020;7:89.
24.
Zurück zum Zitat Kappus S, Haber T, Ismer B, Laubenberger J. Qualitative in-vitro-investigation of the MRI compatibility of temporary myocardial electrodes and pacemaker leads. Forsch im Fokus. 2015;12:130–3. Kappus S, Haber T, Ismer B, Laubenberger J. Qualitative in-vitro-investigation of the MRI compatibility of temporary myocardial electrodes and pacemaker leads. Forsch im Fokus. 2015;12:130–3.
25.
Zurück zum Zitat Park SM, Kamondetdacha R, Nyenhuis JA. Calculation of MRI-induced heating of an implanted medical lead wire with an electric field transfer function. J Magn Reson Imaging. 2007;26:1278–85.CrossRef Park SM, Kamondetdacha R, Nyenhuis JA. Calculation of MRI-induced heating of an implanted medical lead wire with an electric field transfer function. J Magn Reson Imaging. 2007;26:1278–85.CrossRef
26.
Zurück zum Zitat Bonnemains L, Barbier T, Felblinger J. Metal wires should not be abandoned inside implantable cardioverter-defibrillators leads during heart transplantation! Transpl Int. 2016;29:1136–8.CrossRef Bonnemains L, Barbier T, Felblinger J. Metal wires should not be abandoned inside implantable cardioverter-defibrillators leads during heart transplantation! Transpl Int. 2016;29:1136–8.CrossRef
27.
Zurück zum Zitat Wirtz S, Schulte HD, Winter J, Godehardt E, Kunert J. Reliability of different temporary myocardial pacing leads. Thorac Cardiovasc Surg. 1989;37:163–8.CrossRef Wirtz S, Schulte HD, Winter J, Godehardt E, Kunert J. Reliability of different temporary myocardial pacing leads. Thorac Cardiovasc Surg. 1989;37:163–8.CrossRef
28.
Zurück zum Zitat Peden CJ, Collins AG, Butson PC, Whitwam JG, Young IR. Induction of microcurrents in critically ill patients in magnetic resonance systems. Crit Care Med US. 1993;21:1923–8.CrossRef Peden CJ, Collins AG, Butson PC, Whitwam JG, Young IR. Induction of microcurrents in critically ill patients in magnetic resonance systems. Crit Care Med US. 1993;21:1923–8.CrossRef
29.
Zurück zum Zitat Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD, Kanal E, et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: An American heart association scientific statement from the committee on diagnostic and interventional cardiac catheterization, council on clinical cardiology, and the council o. Circulation. 2007;116:2878–91.CrossRef Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD, Kanal E, et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: An American heart association scientific statement from the committee on diagnostic and interventional cardiac catheterization, council on clinical cardiology, and the council o. Circulation. 2007;116:2878–91.CrossRef
30.
Zurück zum Zitat Murphy KJ, Cohan RH, Ellis JH. MR imaging in patients with epicardial pacemaker wires. Am J Roentgenol. 1999;172:727–8.CrossRef Murphy KJ, Cohan RH, Ellis JH. MR imaging in patients with epicardial pacemaker wires. Am J Roentgenol. 1999;172:727–8.CrossRef
31.
Zurück zum Zitat Shaikhrezai K, Khorsandi M, Patronis M, Prasad S. Is it safe to cut pacing wires flush with the skin instead of removing them? Interact Cardiovasc Thorac Surg. 2012;15:1047–51.CrossRef Shaikhrezai K, Khorsandi M, Patronis M, Prasad S. Is it safe to cut pacing wires flush with the skin instead of removing them? Interact Cardiovasc Thorac Surg. 2012;15:1047–51.CrossRef
32.
Zurück zum Zitat Horng GS, Ashley E, Balsam L, Reitz B, Zamanian RT. Progressive dyspnea after CABG: complication of retained epicardial pacing wires. Ann Thorac Surg. 2008;86:1352–4.CrossRef Horng GS, Ashley E, Balsam L, Reitz B, Zamanian RT. Progressive dyspnea after CABG: complication of retained epicardial pacing wires. Ann Thorac Surg. 2008;86:1352–4.CrossRef
33.
Zurück zum Zitat Jafferjee N, Li B, Mijangos S, Thu Ma AM, Manghisi S. Migrated abandoned epicardial pacing wire resulting in a breast abscess. Breast J. 2014;20:200–1.CrossRef Jafferjee N, Li B, Mijangos S, Thu Ma AM, Manghisi S. Migrated abandoned epicardial pacing wire resulting in a breast abscess. Breast J. 2014;20:200–1.CrossRef
34.
Zurück zum Zitat Klein-Wiele O, Garmer M, Busch M, Mateiescu S, Urbien R, Barbone G, et al. Cardiovascular magnetic resonance in patients with magnetic resonance conditional pacemaker systems at 1.5 T: influence of pacemaker related artifacts on image quality including first pass perfusion, aortic and mitral valve assessment, flow measurement. Int J Cardiovasc Imaging. 2017;33:383–94.CrossRef Klein-Wiele O, Garmer M, Busch M, Mateiescu S, Urbien R, Barbone G, et al. Cardiovascular magnetic resonance in patients with magnetic resonance conditional pacemaker systems at 1.5 T: influence of pacemaker related artifacts on image quality including first pass perfusion, aortic and mitral valve assessment, flow measurement. Int J Cardiovasc Imaging. 2017;33:383–94.CrossRef
35.
Zurück zum Zitat Hilbert S, Jahnke C, Loebe S, Oebel S, Weber A, Spampinato R, et al. Cardiovascular magnetic resonance imaging in patients with cardiac implantable electronic devices: A device-dependent imaging strategy for improved image quality. Eur Heart J Cardiovasc Imaging. 2018;19:1051–61.CrossRef Hilbert S, Jahnke C, Loebe S, Oebel S, Weber A, Spampinato R, et al. Cardiovascular magnetic resonance imaging in patients with cardiac implantable electronic devices: A device-dependent imaging strategy for improved image quality. Eur Heart J Cardiovasc Imaging. 2018;19:1051–61.CrossRef
Metadaten
Titel
Safety and image quality of cardiovascular magnetic resonance imaging in patients with retained epicardial pacing wires after heart transplantation
verfasst von
Constantin Gatterer
Marie-Elisabeth Stelzmüller
Andreas Kammerlander
Andreas Zuckermann
Martin Krššák
Christian Loewe
Dietrich Beitzke
Publikationsdatum
01.12.2021
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2021
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-021-00728-1

Weitere Artikel der Ausgabe 1/2021

Journal of Cardiovascular Magnetic Resonance 1/2021 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.