Protocol analysis of dual-energy CT for optimization of kidney stone detection in virtual non-contrast reconstructions

verfasst von: Matthias Lazar, Helmut Ringl, Pascal Baltzer, Daniel Toth, Christian Seitz, Bernhard Krauss, Ewald Unger, Stephan Polanec, Dietmar Tamandl, Christian J. Herold, Michael Toepker

Erschienen in: European Radiology | Ausgabe 8/2020

Einloggen, um Zugang zu erhalten

Abstract

Objectives

Previous studies have shown that split-bolus protocols in virtual non-contrast (VNC) reconstructions of dual-energy computed tomography (DE-CT) significantly decrease radiation dose in patients with urinary stone disease. To evaluate the impact on kidney stone detection rate of stone composition, size, tube voltage, and iodine concentration for VNC reconstructions of DE-CT.

Methods

In this prospective study, 16 kidney stones of different sizes (1.2–4.5 mm) and compositions (struvite, cystine, whewellite, brushite) were placed within a kidney phantom. Seventy-two scans with nine different iodine contrast agents/saline solutions with increasing attenuation (0–1400 HU) and different kilovoltage settings (70 kV/150 kV; 80 kV/150 kV; 90 kV/150 kV; 100 kV/150 kV) were performed. Two experienced radiologists independently rated the images for the presence and absence of stones. Multivariate classification tree analysis and descriptive statistics were used to evaluate the diagnostic performance.

Results

Classification tree analysis revealed a higher detection rate of renal calculi > 2 mm in size compared with that of renal calculi < 2 mm (84.7%; 12.7%; p < 0.001). For stones with a diameter > 2 mm, the best results were found at 70 kV/Sn 150 kV and 80 kV/Sn 150 kV in scans with contrast media attenuation of 600 HU or less, with sensitivity of 99.6% and 96.0%, respectively. A higher luminal attenuation (> 600 HU) resulted in a significantly decreased detection rate (91.8%, 0–600 HU; 70.7%, 900–1400 HU; p < 0.001). In our study setup, the detection rates were best for cystine stones.

Conclusion

Scan protocols in DE-CT with lower tube current and lower contrast medium attenuation show excellent results in VNC for stones larger than 2 mm but have limitations for small stones.

Key Points

• The detection rate of virtual non-contrast reconstructions is highly dependent on the surrounding contrast medium attenuation at the renal pelvis and should be kept as low as possible, as at an attenuation higher than 600 HU the VNC reconstructions are susceptible to masking ureteral stones.
• Protocols with lower tube voltages (70 kV/Sn 150 kV and 80 kV/Sn 150 kV) improve the detection rate of kidney stones in VNC reconstructions.
• The visibility of renal stones in virtual non-contrast of dual-energy CT is highly associated with the size, and results in a significantly lower detection rate in stones below 2 mm.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
3.
Zurück zum Zitat Worster A, Preyra I, Weaver B, Haines T (2002) The accuracy of noncontrast helical computed tomography versus intravenous pyelography in the diagnosis of suspected acute urolithiasis: a meta-analysis. Ann Emerg Med 40:280–286CrossRef Worster A, Preyra I, Weaver B, Haines T (2002) The accuracy of noncontrast helical computed tomography versus intravenous pyelography in the diagnosis of suspected acute urolithiasis: a meta-analysis. Ann Emerg Med 40:280–286CrossRef
4.
Zurück zum Zitat Graser A, Johnson TR, Hecht EM et al (2009) Dual-energy CT in patients suspected of having renal masses: can virtual nonenhanced images replace true nonenhanced images? Radiology 252:433–440CrossRef Graser A, Johnson TR, Hecht EM et al (2009) Dual-energy CT in patients suspected of having renal masses: can virtual nonenhanced images replace true nonenhanced images? Radiology 252:433–440CrossRef
5.
Zurück zum Zitat Toepker M, Moritz T, Krauss B et al (2012) Virtual non-contrast in second-generation, dual-energy computed tomography: reliability of attenuation values. Eur J Radiol 81:e398–e405CrossRef Toepker M, Moritz T, Krauss B et al (2012) Virtual non-contrast in second-generation, dual-energy computed tomography: reliability of attenuation values. Eur J Radiol 81:e398–e405CrossRef
6.
Zurück zum Zitat Mangold S, Thomas C, Fenchel M et al (2012) Virtual nonenhanced dual-energy CT urography with tin-filter technology: determinants of detection of urinary calculi in the renal collecting system. Radiology 264:119–125CrossRef Mangold S, Thomas C, Fenchel M et al (2012) Virtual nonenhanced dual-energy CT urography with tin-filter technology: determinants of detection of urinary calculi in the renal collecting system. Radiology 264:119–125CrossRef
7.
Zurück zum Zitat Scheffel H, Stolzmann P, Frauenfelder T et al (2007) Dual-energy contrast-enhanced computed tomography for the detection of urinary stone disease. Invest Radiol 42:823–829CrossRef Scheffel H, Stolzmann P, Frauenfelder T et al (2007) Dual-energy contrast-enhanced computed tomography for the detection of urinary stone disease. Invest Radiol 42:823–829CrossRef
8.
Zurück zum Zitat Takahashi N, Vrtiska TJ, Kawashima A et al (2010) Detectability of urinary stones on virtual nonenhanced images generated at pyelographic-phase dual-energy CT. Radiology 256:184–190CrossRef Takahashi N, Vrtiska TJ, Kawashima A et al (2010) Detectability of urinary stones on virtual nonenhanced images generated at pyelographic-phase dual-energy CT. Radiology 256:184–190CrossRef
9.
Zurück zum Zitat Toepker M, Kuehas F, Kienzl D et al (2014) Dual energy computerized tomography with a split bolus-a 1-stop shop for patients with suspected urinary stones? J Urol 191:792–797CrossRef Toepker M, Kuehas F, Kienzl D et al (2014) Dual energy computerized tomography with a split bolus-a 1-stop shop for patients with suspected urinary stones? J Urol 191:792–797CrossRef
10.
Zurück zum Zitat Smith-Bindman R, Lipson J, Marcus R et al (2009) Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med 169:2078–2086CrossRef Smith-Bindman R, Lipson J, Marcus R et al (2009) Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med 169:2078–2086CrossRef
11.
Zurück zum Zitat Yeo YJ, Kim SH, Kim MJ, Kim YH, Cho SH, Lee EJ (2015) Diagnostic efficiency of split-bolus dual-energy computed tomography for patients with suspected urinary stones. J Comput Assist Tomogr 39:25–31CrossRef Yeo YJ, Kim SH, Kim MJ, Kim YH, Cho SH, Lee EJ (2015) Diagnostic efficiency of split-bolus dual-energy computed tomography for patients with suspected urinary stones. J Comput Assist Tomogr 39:25–31CrossRef
12.
Zurück zum Zitat Toepker M, Euller G, Unger E et al (2013) Stenosis quantification of coronary arteries in coronary vessel phantoms with second-generation dual-source CT: influence of measurement parameters and limitations. AJR Am J Roentgenol 201:W227–W234CrossRef Toepker M, Euller G, Unger E et al (2013) Stenosis quantification of coronary arteries in coronary vessel phantoms with second-generation dual-source CT: influence of measurement parameters and limitations. AJR Am J Roentgenol 201:W227–W234CrossRef
13.
Zurück zum Zitat Hesse A, Kruse R, Geilenkeuser WJ, Schmidt M (2005) Quality control in urinary stone analysis: results of 44 ring trials (1980-2001). Clin Chem Lab Med 43:298–303CrossRef Hesse A, Kruse R, Geilenkeuser WJ, Schmidt M (2005) Quality control in urinary stone analysis: results of 44 ring trials (1980-2001). Clin Chem Lab Med 43:298–303CrossRef
14.
Zurück zum Zitat Zeng G, Zhao Z, Wu W, Ou L, Liang Y, Yuan J (2014) Interconversion of stone composition profiles from two recurrent stone episodes in stone formers. Clin Chem Lab Med 52:1019–1024PubMed Zeng G, Zhao Z, Wu W, Ou L, Liang Y, Yuan J (2014) Interconversion of stone composition profiles from two recurrent stone episodes in stone formers. Clin Chem Lab Med 52:1019–1024PubMed
15.
Zurück zum Zitat Karlo CA, Gnannt R, Winklehner A et al (2013) Split-bolus dual-energy CT urography: protocol optimization and diagnostic performance for the detection of urinary stones. Abdom Imaging 38:1136–1143CrossRef Karlo CA, Gnannt R, Winklehner A et al (2013) Split-bolus dual-energy CT urography: protocol optimization and diagnostic performance for the detection of urinary stones. Abdom Imaging 38:1136–1143CrossRef
16.
Zurück zum Zitat Chen CY, Hsu JS, Jaw TS et al (2015) Split-bolus portal venous phase dual-energy CT urography: protocol design, image quality, and dose reduction. AJR Am J Roentgenol 205:W492–W501CrossRef Chen CY, Hsu JS, Jaw TS et al (2015) Split-bolus portal venous phase dual-energy CT urography: protocol design, image quality, and dose reduction. AJR Am J Roentgenol 205:W492–W501CrossRef
17.
Zurück zum Zitat Takahashi N, Hartman RP, Vrtiska TJ et al (2008) Dual-energy CT iodine-subtraction virtual unenhanced technique to detect urinary stones in an iodine-filled collecting system: a phantom study. AJR Am J Roentgenol 190:1169–1173CrossRef Takahashi N, Hartman RP, Vrtiska TJ et al (2008) Dual-energy CT iodine-subtraction virtual unenhanced technique to detect urinary stones in an iodine-filled collecting system: a phantom study. AJR Am J Roentgenol 190:1169–1173CrossRef
18.
Zurück zum Zitat Preminger GM, Tiselius HG, Assimos DG et al (2007) 2007 guideline for the management of ureteral calculi. Eur Urol 52:1610–1631CrossRef Preminger GM, Tiselius HG, Assimos DG et al (2007) 2007 guideline for the management of ureteral calculi. Eur Urol 52:1610–1631CrossRef
19.
Zurück zum Zitat Coursey CA, Nelson RC, Boll DT et al (2010) Dual-energy multidetector CT: how does it work, what can it tell us, and when can we use it in abdominopelvic imaging? Radiographics 30:1037–1055CrossRef Coursey CA, Nelson RC, Boll DT et al (2010) Dual-energy multidetector CT: how does it work, what can it tell us, and when can we use it in abdominopelvic imaging? Radiographics 30:1037–1055CrossRef
20.
Zurück zum Zitat Johnson TR, Krauss B, Sedlmair M et al (2007) Material differentiation by dual energy CT: initial experience. Eur Radiol 17:1510–1517CrossRef Johnson TR, Krauss B, Sedlmair M et al (2007) Material differentiation by dual energy CT: initial experience. Eur Radiol 17:1510–1517CrossRef
21.
Zurück zum Zitat Yu L, Primak AN, Liu X, McCollough CH (2009) Image quality optimization and evaluation of linearly mixed images in dual-source, dual-energy CT. Med Phys 36:1019–1024CrossRef Yu L, Primak AN, Liu X, McCollough CH (2009) Image quality optimization and evaluation of linearly mixed images in dual-source, dual-energy CT. Med Phys 36:1019–1024CrossRef
22.
Zurück zum Zitat Bellin MF, Renard-Penna R, Conort P et al (2004) Helical CT evaluation of the chemical composition of urinary tract calculi with a discriminant analysis of CT-attenuation values and density. Eur Radiol 14:2134–2140CrossRef Bellin MF, Renard-Penna R, Conort P et al (2004) Helical CT evaluation of the chemical composition of urinary tract calculi with a discriminant analysis of CT-attenuation values and density. Eur Radiol 14:2134–2140CrossRef
24.
Zurück zum Zitat Hidas G, Eliahou R, Duvdevani M et al (2010) Determination of renal stone composition with dual-energy CT: in vivo analysis and comparison with X-ray diffraction. Radiology 257:394–401CrossRef Hidas G, Eliahou R, Duvdevani M et al (2010) Determination of renal stone composition with dual-energy CT: in vivo analysis and comparison with X-ray diffraction. Radiology 257:394–401CrossRef
25.
Zurück zum Zitat Nakada SY, Hoff DG, Attai S, Heisey D, Blankenbaker D, Pozniak M (2000) Determination of stone composition by noncontrast spiral computed tomography in the clinical setting. Urology 55:816–819CrossRef Nakada SY, Hoff DG, Attai S, Heisey D, Blankenbaker D, Pozniak M (2000) Determination of stone composition by noncontrast spiral computed tomography in the clinical setting. Urology 55:816–819CrossRef
26.
Zurück zum Zitat Khan SR, Pearle MS, Robertson WG et al (2017) Kidney stones. Nat Rev Dis Primers 3:17001CrossRef Khan SR, Pearle MS, Robertson WG et al (2017) Kidney stones. Nat Rev Dis Primers 3:17001CrossRef
27.
Zurück zum Zitat Knoll T, Bach T, Humke U et al (2016) S2k guidelines on diagnostics, therapy and metaphylaxis of urolithiasis (AWMF 043/025) : Compendium. Urologe A 55:904–922CrossRef Knoll T, Bach T, Humke U et al (2016) S2k guidelines on diagnostics, therapy and metaphylaxis of urolithiasis (AWMF 043/025) : Compendium. Urologe A 55:904–922CrossRef
29.
Zurück zum Zitat Grosse Hokamp N, Lennartz S, Salem J et al (2020) Dose independent characterization of renal stones by means of dual energy computed tomography and machine learning: an ex-vivo study. Eur Radiol 30:1397–1404CrossRef Grosse Hokamp N, Lennartz S, Salem J et al (2020) Dose independent characterization of renal stones by means of dual energy computed tomography and machine learning: an ex-vivo study. Eur Radiol 30:1397–1404CrossRef
30.
Zurück zum Zitat Edwards TJ, Dickinson AJ, Natale S, Gosling J, McGrath JS (2006) A prospective analysis of the diagnostic yield resulting from the attendance of 4020 patients at a protocol-driven haematuria clinic. BJU Int 97:301–305 discussion 305CrossRef Edwards TJ, Dickinson AJ, Natale S, Gosling J, McGrath JS (2006) A prospective analysis of the diagnostic yield resulting from the attendance of 4020 patients at a protocol-driven haematuria clinic. BJU Int 97:301–305 discussion 305CrossRef
Metadaten
Titel
Protocol analysis of dual-energy CT for optimization of kidney stone detection in virtual non-contrast reconstructions
verfasst von
Matthias Lazar
Helmut Ringl
Pascal Baltzer
Daniel Toth
Christian Seitz
Bernhard Krauss
Ewald Unger
Stephan Polanec
Dietmar Tamandl
Christian J. Herold
Michael Toepker
Publikationsdatum
03.04.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 8/2020
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-06806-9

Weitere Artikel der Ausgabe 8/2020

European Radiology 8/2020 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.