Myocardial late gadolinium enhancement using delayed 3D IR-FLASH in the pediatric population: feasibility and diagnostic performance compared to single-shot PSIR-bSSFP

verfasst von: Ankavipar Saprungruang, Julien Aguet, Navjot Gill, Vivian P. Tassos, Afsaneh Amirabadi, Mike Seed, Shi-Joon Yoo, Christopher Z. Lam

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2023

Abstract

Background

This study compares three-dimensional (3D) high-resolution (HR) late gadolinium enhancement (LGE; 3D HR-LGE) imaging using a respiratory navigated, electrocardiographically-gated inversion recovery gradient echo sequence with conventional LGE imaging using a single-shot phase-sensitive inversion recovery (PSIR) balanced steady-state free precession (bSSFP; PSIR-bSSFP) sequence for routine clinical use in the pediatric population.

Methods

Pediatric patients (0–18 years) who underwent clinical cardiovascular magnetic resonance (CMR) with both 3D HR-LGE and single-shot PSIR-bSSFP LGE between January 2018 and June 2020 were included. Image quality (0–4) and detection of LGE in the left ventricle (LV) (per 17 segments), in the right ventricle (RV) (per 3 segments), as endocardial fibroelastosis (EFE), at the hinge points, and at the papillary muscles was analyzed by two blinded readers for each sequence. Ratios of the mean signal intensity of LGE to normal myocardium (LGE:Myo) and to LV blood pool (LGE:Blood) were recorded. Data is presented as median (1st–3rd quartiles). Wilcoxon signed rank test and chi-square analyses were used as appropriate. Inter-rater agreement was analyzed using weighted κ-statistics.

Results

102 patients were included with median age at CMR of 8 (1–13) years-old and 44% of exams performed under general anesthesia. LGE was detected in 55% of cases. 3D HR LGE compared to single-shot PSIR-bSSFP had longer scan time [4:30 (3:35–5:34) vs 1:11 (0:47–1:32) minutes, p < 0.001], higher image quality ratings [3 (3–4) vs 2 (2–3), p < 0.001], higher LGE:Myo [23.7 (16.9–31.2) vs 5.0 (2.9–9.0), p < 0.001], detected more segments of LGE in both the LV [4 (2–8) vs 3 (1–7), p = 0.045] and RV [1 (1–1) vs 1 (0–1), p < 0.001], and also detected more cases of LGE with 13/56 (23%) of patients with LGE only detectable by 3D HR LGE (p < 0.001). 3D HR LGE specifically detected a greater proportion of RV LGE (27/27 vs 17/27, p < 0.001), EFE (11/11 vs 5/11, p = 0.004), and papillary muscle LGE (14/15 vs 4/15, p < 0.001). Inter-rater agreement for the recorded variables ranged from 0.42 to 1.00.

Conclusions

3D HR LGE achieves greater image quality and detects more LGE than conventional single-shot PSIR-bSSFP LGE imaging, and should be considered an alternative to conventional LGE sequences for routine clinical use in the pediatric population.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s12968-023-00917-0.
Ankavipar Saprungruang and Julien Aguet are co-first authors

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
bSSFP
Balanced steady-state free precession
CHD
Congenital heart disease
CMR
Cardiovascular magnetic resonance
ECG
Electrocardiogram
EFE
Endocardial fibroelastosis
FLASH
Fast low angle shot
FOV
Field of view
GBCA
Gadolinium-based contrast agent
GRE
Gradient echo
HR
High resolution
IR
Inversion recovery
LGE
Late gadolinium enhancement
LV
Left ventricle/left ventricular
PSIR
Phase-sensitive inversion recovery
RV
Right ventricle/right ventricular
TE
Echo time
TI
Inversion time
TR
Repetition time

Background

Late gadolinium enhancement (LGE) cardiovascular magnetic resonance (CMR) evaluation of myocardial injury and fibrosis provides essential diagnostic and prognostic information in a variety of congenital and pediatric heart conditions [14]. One of the conventional methods of performing LGE imaging is by using a two-dimensional single-shot phase-sensitive inversion recovery (PSIR) balanced steady-state free precession (bSSFP; PSIR-bSSFP) sequence [1]. Recently, a three-dimensional (3D) high-resolution (HR) LGE (3D HR-LGE) option using a respiratory navigated, electrocardiographically (ECG)-gated inversion recovery gradient echo sequence has been described [57]. This sequence, that is widely used for contrast-enhanced CMR angiography [8, 9], has also been investigated for routine clinical use for LGE imaging in adult patients for selected conditions, such as ischemic cardiomyopathy, atrial and ventricular arrhythmias, and tetralogy of Fallot [7].
Advantages of 3D HR-LGE include higher spatial resolution, increased signal to noise, better fat saturation, and ability to perform multiplanar reconstructions. Disadvantages include a long scan time and potential for compromised quality due to inconsistent breathing, motion, significant heart rate variation, arrhythmia, or inappropriate inversion time [7]. The performance of 3D HR-LGE in the pediatric population has not yet been investigated.
The purpose of this study was to evaluate the use of a 3D HR-LGE sequence compared to a conventional single-shot PSIR-bSSFP LGE sequence for detection of LGE in routine pediatric clinical CMR.

Materials and methods

Study population

This single institution retrospective study received institutional Research Ethics Board approval with waiver of informed consent. Patients between 0 and 18 years of age that underwent a clinical CMR that included LGE imaging performed between January 2018 and June 2020 were included. CMRs that did not have both 3D HR-LGE and single-shot PSIR-bSSFP LGE were excluded. This was related to variable adoption of 3D HR-LGE early in the inclusion period along with standardization of our clinical practice by using conventional single-shot PSIR-bSSFP LGE exclusively as opposed to occasional alternative use of conventional segmented PSIR-FLASH LGE. Non-diagnostic exams related to technical factors were also excluded, which mostly were related to variable attempts at optimization of technique parameters early in the adoption period. Note that exams with excessive patient motion artifact were not excluded.

CMR acquisition

Studies were performed on a 1.5 T CMR system (Avanto Fit, Siemens, Healthineers, Erlangen, Germany). Conventional LGE was performed using a two-dimensional single-shot PSIR-bSSFP sequence acquired in axial and short-axis planes 10 min following intravenous injection of 0.2 mmol/kg of gadobutrol (Gadovist; Bayer Healthcare, Berlin, Germany). 3D HR-LGE was performed using a 3D ECG-gated, respiratory-navigated inversion recovery-prepared (IR) fast low angle shot (FLASH) pulse sequence with fat saturation acquired in coronal plane, and obtained 15–20 min following intravenous administration of gadolinium-based contrast agent. Respiratory navigation was performed using pencil-beam navigation. As exact nulling of myocardium is not necessary for this sequence, an empiric inversion time of 220 ms was used with quiescent interval chosen in systole. However, when the Look-Locker acquired before conventional LGE gave a myocardial nulling inversion time ≥ 300 ms, an empiric inversion time of 280 ms was used for the 3D HR-LGE sequence with quiescent interval in diastole. Other specific sequence parameters are shown in Table 1. Isotropic axial and ventricular short axis planes in the native resolution were reconstructed for the 3D HR-LGE sequence for purposes of analysis.
Table 1
Technical parameters of LGE sequences
CMR parameter
Conventional LGE
3D HR-LGE
Base sequence
Single-shot PSIR-bSSFP
3D IR-FLASH
ECG gating
Mid-late diastole
Cardiac quiescence in systole or diastole
Parallel imaging × acceleration factor
GRAPPA × 2
GRAPPA × 2
Field of view
250–350 mm
250–350 mm
Pixel size
1.8 × 1.8 mm
1.1 × 1.1 to 1.4 × 1.4 mm
Slice thickness
6 mm (no gap)
1.1 to 1.4 mm
Flip angle
40˚
18˚
TR/TE
2.57/1.09 ms
330/1.4 ms
Number of averages
1–3a
1
Fat saturation
Yes
Yes
Acquisition plane(s)
Axial, short axis
Coronal
ECG electrocardiogram; FLASH fast low angle shot; GRAPPA generalized autocalibrating partial parallel acquisition; IR inversion recovery; PSIR phase-sensitive inversion recovery; bSSFP balanced steady-state free precession; TE time to echo; TR repetition time
aSmaller patients and those performed under general anesthesia were performed free breathing with more averages to reduce motion artifact and increase signal

CMR analysis

Each sequence was analyzed by two independent, blinded readers (A.S and J.A.). Presence or absence of LGE per myocardial segment was recorded for each sequence. The left ventricle (LV) was divided into 17 myocardial segments according to standardized American Heart Association LV segmentation [10]. The right ventricle (RV) was divided into three segments: inlet, apical and outlet. In addition, presence of endocardial fibroelastosis (EFE), subendocardial LGE, mid-myocardial LGE, subepicardial LGE, or LGE at the ventricular hinge points and papillary muscles were also recorded. For cases where there was disagreement between presence and absence of LGE, consensus was reached for final analysis via a third reader (C.Z.L.).
Image quality of each sequence was rated on a four-point ordinal scale: (1) poor with no clear distinction of myocardial margins; (2) fair with blurred myocardial edges; (3) good with clear myocardial margins; (4) very good with sharp myocardial margins (Fig. 1).
When myocardial LGE was detected on both sequences, the mean signal intensity of LGE, normal myocardium, and LV blood pool were measured using manually drawn regions of interest. LGE to myocardium (LGE:Myo) and LGE to LV blood pool (LGE:Blood) signal intensity ratios were calculated and compared.

Statistical analysis

Continuous variables are summarized as median (1st–3rd quartiles) and categorical variables are summarized as frequency (percentage). κ-statistics were used for inter-rater agreement of LGE detection and image quality ratings between the two sequences. Wilcoxon signed rank test and chi-square analyses were used to compare conventional LGE and 3D HR LGE sequences. All statistical tests were two-sided and considered statistically significant if p < 0.05. Analyses were performed using SPSS for Windows (version 28.0.0.0, Statistical Package for the Social Sciences, International Business Machines, Inc., Armonk, New York, USA).

Results

Patient selection and demographics

One-hundred and sixty-six patients underwent clinical CMR that included LGE imaging during the study period. Thirty-eight patients were excluded due to the lack of concomitantly performed 3D HR-LGE and conventional single-shot PSIR-bSSFP LGE sequences. One patient was excluded due to extra-cardiac artifact obscuring the heart. Twenty-six (20%) of the remaining patients were excluded due to non-diagnostic images related to inappropriate sequence timing or inversion time, the majority as part of sequence testing and optimization early in the adoption period. This left a final study population of 102 patients (53% male; median age at CMR was 8 (1–13) years) with 45/102 (44%) of cases performed under general anesthesia. Demographic data are summarized in Table 2.
Table 2
Patient demographics
Characteristic
Total cases (n = 102)
Cases with LGE (n = 56)
Age (years)
8 (1–13)
10 (2–15)
Weight (kg)
26 (10–50)
39 (12–56)
Body surface area (m2)
0.96 (0.41–1.47)
1.21 (0.48–1.57)
Diagnosis
  
 Congenital heart disease
64 (63%)
41 (73%)
 Cardiomyopathy
16 (16%)
9 (16%)
 Systemic vasculitis
16 (16%)
5 (9%)
 Systemic vasculopathy
4 (4%)
1 (2%)
 Coronary anomaly
2 (2%)
0 (0%)
Data are presented as median (1st–3rd quartile) or n (%)

LGE sequence quality and signal intensity ratios

Median 3D HR-LGE scan time was 4:30 (3:35–5:34) minutes and was shorter when performed under general anesthesia [3:46 (3:01–4:26) vs 5:18 (4:31–6:17) minutes, p < 0.001]. Ninety-eight (96%) of 3D HR-LGE sequences were performed in systole. Median combined short-axis and axial single-shot PSIR-bSSFP total scan time was 1:11 (0:47–1:32) minutes and also was shorter when performed under general anesthesia [1:02 (0:42–1:23) vs 1:18 (0:54–1:35) minutes, p = 0.026]. Image quality ratings were higher for 3D HR-LGE compared with single-shot PSIR-bSSFP at 3 (3–4) vs 2 (2–3) respectively (p < 0.001). Image quality ratings were not different between sedated or awake 3D HR-LGE and PSIR-bSSFP cases (p = 0.495 and 0.468, respectively). LGE:Myo signal intensity was higher for 3D HR-LGE compared with single-shot PSIR-bSSFP at 23.7 (16.9–31.2) vs 5.0 (2.9–9.0) (p < 0.001), while there were no difference in LGE:Blood signal intensity at 1.2 (0.9–1.8) vs 1.2 (0.6–1.5) (p = 0.284). Boxplots of these data are depicted in Fig. 2.

Detection of LGE

Fifty-six (55%) of exams demonstrated myocardial LGE. Specifically, 24 (24%) had LV LGE, 27 (26%) had RV LGE, 8 (8%) had LV EFE, 3 (3%) had RV EFE, 34 (33%) had subendocardial LGE, 41 (40%) had mid-myocardial LGE, 26 (25%) had subepicardial LGE, 16 (16%) had hinge point LGE, and 15 (15%) had papillary muscle LGE. The 3D HR-LGE sequence detected a greater extent of LGE than single-shot PSIR-bSSFP, with median number of positive LV segments of 4 (2–8) vs 3 (1–7) (p = 0.045), and median positive RV segments of 1 (1–1) vs 1 (0–1) (p < 0.001), respectively (Fig. 3). Moreover, LGE was detected only on the 3D HR-LGE sequence in 4/24 (17%) cases with LV LGE, 10/27 (37%) with RV LGE, 4/8 (50%) with LV EFE, 2/3 (67%) with RV EFE, 12/34 (35%) with subendocardial LGE, 12/41 (29%) with mid-myocardial LGE, 9/26 (35%) with subepicardial LGE, 6/16 (38%) with hinge point LGE, and 11/15 (73%) with papillary muscle LGE. One/24 (4%) cases with LV LGE, 1/41 (2%) with mid-myocardial LGE, 2/16 (13%) with hinge point LGE, and 1/15 (7%) with papillary muscle LGE were only seen on conventional single-shot PSIR-bSSFP LGE. Further review of these cases in conjunction with all other CMR sequences revealed that the cases of LV LGE, mid-myocardial LGE, and papillary muscle LGE were false-positives (these were left as positive cases for purposes of statistical analysis). This data are summarized in Table 3. Representative side-by-side comparison examples are depicted in Figs. 4, 5, 67.
Table 3
Detection of LGE using 3D HR-LGE and conventional single-shot PSIR-bSSFP LGE sequences in 56 patients with positive LGE
Location of LGE
LGE on 3D HR- LGE only (%)
LGE on PSIR-bSSFP LGE only (%)
p-value
LGE in any location (n = 56)
13 (23)
0 (0)
 < 0.001
LV LGE (n = 24)
4 (17)
1 (4)a
0.346
RV LGE (n = 27)
10 (37)
0 (0)
 < 0.001
LV or RV EFE (n = 11)
6 (55)
0 (0)
0.004
Subendocardial LGE (n = 34)
12 (35)
0 (0)
 < 0.001
Mid-myocardial LGE (n = 41)
12 (29)
1 (2)a
0.002
Subepicardial LGE (n = 26)
9 (35)
0 (0)
0.002
Hinge point LGE (n = 16)
6 (38)
2 (13)
0.220
Papillary muscle LGE (n = 15)
11 (73)
1 (7)a
 < 0.001
aFalse positives after re-review
3D 3-dimensional, EFE endocardial fibroelastosis, HR high resolution, LGE late gadolinium enhancement, LV left ventricle, PSIR phase-sensitive inversion recovery, R right ventricle, bSSFP balanced steady-state free precession

Inter-rater agreement

Inter-rater agreement via weighted κ-statistics was 0.86 [95% CI 0.79–0.93] for 3D HR LGE image quality ratings, 0.60 [95% CI 0.46–0.74] for single-shot PSIR-bSSFP LGE image quality ratings, and ranged from 0.42 to 1.00 for detection of LGE depending on myocardial location and sequence. Inter-rater agreement was not clearly different between the two techniques, with overlapping 95% confidence intervals. This data is summarized in Additional file 1: Table S1.

Discussion

In this study, a 3D HR-LGE sequence was compared with conventional single-shot PSIR-bSSFP LGE imaging as a clinical routine in the pediatric population. The key findings of this work are that 3D HR-LGE is feasible for routine use in pediatric clinical CMR, and compared to single-shot PSIR-bSSFP LGE, achieves greater image quality and detects more LGE, particularly for the RV, EFE, and papillary muscles.
Some of the disadvantages of a 3D HR-LGE sequence relate to the long scan time that may result in patient motion and suboptimal image quality. In this cohort, 3D HR-LGE was performed with a median scan time of 4.5 min, which is shorter than reported in adults with an average of 10 min [7]. This may be related to faster heart rates, faster respiratory rates, and smaller size of children. Conversely, single-shot PSIR-bSSFP sequences suffer from low signal in small children with fast heart rates. 3D HR-LGE may thus be well suited for the pediatric population, particularly when the study is performed under general anesthesia or sedation where respirations are better controlled. Indeed, in this study we confirm that 3D HR-LGE shows greater image quality and a four to fivefold higher LGE:myo signal intensity compared with conventional LGE.
This translated to superior diagnostic performance with 3D HR-LGE detecting both more patients with LGE, and greater extent of LGE when present. In fact, in 23% of patients that had LGE, the LGE was not detectable by single-shot PSIR-bSSFP. The performance of 3D HR-LGE was particularly superior for LGE of small or thin structures, comparable to the use of 3D HR-LGE for the atrium in adults [11]. In our population, this improved performance was most evident for RV LGE, EFE, and papillary muscle LGE, where single-shot PSIR-bSSFP missed up to 37% of RV LGE, 55% of EFE, and 38% of papillary muscle LGE. Some of the superior performance for RV LGE may be related to the fact 3D HR-LGE was primarily performed in systole whereas single-shot PSIR-bSSFP was performed in diastole, however, the improved spatial resolution and signal of the 3D HR-LGE sequence was also likely beneficial. The fact that 3D HR-LGE was superior to PSIR-bSSFP for detection of LGE in all myocardial layers, along with the similar LGE:blood signal intensity, suggests that improved contrast between subendocardial LGE and blood pool was not necessarily the primary driver of superior performance in this study. Therefore, this technique may be adjunctive to proposed dark-blood or gray-blood LGE strategies that aim to optimize LGE:blood contrast [12]. Importantly, 3D HR-LGE did not miss any clinically significant LGE in our study (only two cases of hinge point LGE that was related to patient motion on the 3D HR-LGE sequence), and in fact, helped to show apparent LGE on single-shot PSIR-bSSFP sequences were false positives in 2 cases. Our data therefore supports using 3D HR-LGE as a clinical standard in the pediatric population.
Of practical considerations for implementation, this sequence is widely available and can be repeated for both angiography and LGE imaging in the same exam. It does not preclude perfusion imaging, conventional LGE, or T1/ECV mapping. At our institution we no longer perform conventional LGE sequences for non-cardiomyopathy patients.

Limitations

Our study is limited by its retrospective design using clinical cases that did not allow for strict control of technical parameters. The true failure rate of the 3D HR-LGE sequence could not be determined as we included examinations performed during our testing and adoption period, which produced some non-diagnostic exams due to varying of CMR parameters. However, with subsequent experience and optimization, recent internal quality improvement efforts analyzing 100 consecutive patients showed that our current non-diagnostic rate is only 4/100 (4%), 3 related to excessive patient motion (poor patient cooperation or inconsistent breathing pattern) and 1 related to technical artifact. Moreover, our results may be confounded by heterogeneity in exams including comparisons of 3D HR-LGE acquisitions in systole versus single-shot PSIR-bSSFP in diastole. Nonetheless, we feel our results remain valid, showing that 3D HR-LGE as we performed in clinical routine is superior to the conventional method of acquiring single-shot PSIR-bSSFP. Finally, this study was limited by the lack of pathologic correlation or gold-standard for presence of LGE.

Conclusion

3D HR-LGE is feasible for routine use in pediatric clinical CMR, and compared to single-shot PSIR-bSSFP LGE, achieves greater image quality and detects more LGE. Given wide-spread availability and superior performance, 3D HR-LGE can be used as the primary standard LGE sequence in the pediatric population, with conventional LGE sequences reserved only for situations when proper ECG gating and/or respiration navigation are not possible.

Acknowledgements

None.

Declarations

This study received approval with waiver of informed consent from the institutional Research Ethics Board at the Hospital for Sick Children, Toronto, Canada.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Kellman P, Arai AE. Cardiac imaging techniques for physicians: late enhancement. J Magn Reson Imaging. 2012;36(3):529–42.CrossRef Kellman P, Arai AE. Cardiac imaging techniques for physicians: late enhancement. J Magn Reson Imaging. 2012;36(3):529–42.CrossRef
2.
Zurück zum Zitat Axelsson Raja A, Farhad H, Valente AM, Couce JP, Jefferies JL, Bundgaard H, et al. Prevalence and progression of late gadolinium enhancement in children and adolescents with hypertrophic cardiomyopathy. Circulation. 2018;138(8):782–92.CrossRef Axelsson Raja A, Farhad H, Valente AM, Couce JP, Jefferies JL, Bundgaard H, et al. Prevalence and progression of late gadolinium enhancement in children and adolescents with hypertrophic cardiomyopathy. Circulation. 2018;138(8):782–92.CrossRef
3.
Zurück zum Zitat Muscogiuri G, Ciliberti P, Mastrodicasa D, Chinali M, Rinelli G, Santangelo TP, et al. Results of late gadolinium enhancement in children affected by dilated cardiomyopathy. Front Pediatr. 2017;5:13.CrossRef Muscogiuri G, Ciliberti P, Mastrodicasa D, Chinali M, Rinelli G, Santangelo TP, et al. Results of late gadolinium enhancement in children affected by dilated cardiomyopathy. Front Pediatr. 2017;5:13.CrossRef
4.
Zurück zum Zitat Etesami M, Gilkeson RC, Rajiah P. Utility of late gadolinium enhancement in pediatric cardiac MRI. Pediatr Radiol. 2016;46(8):1096–113.CrossRef Etesami M, Gilkeson RC, Rajiah P. Utility of late gadolinium enhancement in pediatric cardiac MRI. Pediatr Radiol. 2016;46(8):1096–113.CrossRef
5.
Zurück zum Zitat Oakes RS, Badger TJ, Kholmovski EG, Akoum N, Burgon NS, Fish EN, et al. Detection and quantification of left atrial structural remodeling with delayed-enhancement magnetic resonance imaging in patients with atrial fibrillation. Circulation. 2009;119(13):1758–67.CrossRef Oakes RS, Badger TJ, Kholmovski EG, Akoum N, Burgon NS, Fish EN, et al. Detection and quantification of left atrial structural remodeling with delayed-enhancement magnetic resonance imaging in patients with atrial fibrillation. Circulation. 2009;119(13):1758–67.CrossRef
6.
Zurück zum Zitat Lintingre PF, Nivet H, Clément-Guinaudeau S, Camaioni C, Sridi S, Corneloup O, et al. High-resolution late gadolinium enhancement magnetic resonance for the diagnosis of myocardial infarction with nonobstructed coronary arteries. JACC Cardiovasc Imaging. 2020;13(5):1135–48.CrossRef Lintingre PF, Nivet H, Clément-Guinaudeau S, Camaioni C, Sridi S, Corneloup O, et al. High-resolution late gadolinium enhancement magnetic resonance for the diagnosis of myocardial infarction with nonobstructed coronary arteries. JACC Cardiovasc Imaging. 2020;13(5):1135–48.CrossRef
7.
Zurück zum Zitat Toupin S, Pezel T, Bustin A, Cochet H. Whole-heart high-resolution late gadolinium enhancement: techniques and clinical applications. J Magn Reson Imaging. 2022;55(4):967–87.CrossRef Toupin S, Pezel T, Bustin A, Cochet H. Whole-heart high-resolution late gadolinium enhancement: techniques and clinical applications. J Magn Reson Imaging. 2022;55(4):967–87.CrossRef
8.
Zurück zum Zitat Aguet J, Gill N, Tassos VP, Chavhan GB, Lam CZ. Contrast-enhanced body magnetic resonance angiography: how we do it. Pediatr Radiol. 2022;52(2):262–70.CrossRef Aguet J, Gill N, Tassos VP, Chavhan GB, Lam CZ. Contrast-enhanced body magnetic resonance angiography: how we do it. Pediatr Radiol. 2022;52(2):262–70.CrossRef
9.
Zurück zum Zitat Lam CZ, Pagano JJ, Gill N, Vidarsson L, de la Mora R, Seed M, et al. Dual phase infusion with bolus tracking: technical innovation for cardiac and respiratory navigated magnetic resonance angiography using extracellular contrast. Pediatr Radiol. 2019;49(3):399–406.CrossRef Lam CZ, Pagano JJ, Gill N, Vidarsson L, de la Mora R, Seed M, et al. Dual phase infusion with bolus tracking: technical innovation for cardiac and respiratory navigated magnetic resonance angiography using extracellular contrast. Pediatr Radiol. 2019;49(3):399–406.CrossRef
10.
Zurück zum Zitat Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Int J Cardiovasc Imaging. 2002;18(1):539–42. Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. Int J Cardiovasc Imaging. 2002;18(1):539–42.
11.
Zurück zum Zitat Siebermair J, Kholmovski EG, Marrouche N. Assessment of left atrial fibrosis by late gadolinium enhancement magnetic resonance imaging: methodology and clinical implications. JACC Clin Electrophysiol. 2017;3(8):791–802.CrossRef Siebermair J, Kholmovski EG, Marrouche N. Assessment of left atrial fibrosis by late gadolinium enhancement magnetic resonance imaging: methodology and clinical implications. JACC Clin Electrophysiol. 2017;3(8):791–802.CrossRef
12.
Zurück zum Zitat Holtackers RJ, Van De Heyning CM, Chiribiri A, Wildberger JE, Botnar RM, Kooi ME. Dark-blood late gadolinium enhancement cardiovascular magnetic resonance for improved detection of subendocardial scar: a review of current techniques. J Cardiovasc Magn Reason. 2021;23(1):96.CrossRef Holtackers RJ, Van De Heyning CM, Chiribiri A, Wildberger JE, Botnar RM, Kooi ME. Dark-blood late gadolinium enhancement cardiovascular magnetic resonance for improved detection of subendocardial scar: a review of current techniques. J Cardiovasc Magn Reason. 2021;23(1):96.CrossRef
Metadaten
Titel
Myocardial late gadolinium enhancement using delayed 3D IR-FLASH in the pediatric population: feasibility and diagnostic performance compared to single-shot PSIR-bSSFP
verfasst von
Ankavipar Saprungruang
Julien Aguet
Navjot Gill
Vivian P. Tassos
Afsaneh Amirabadi
Mike Seed
Shi-Joon Yoo
Christopher Z. Lam
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2023
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-023-00917-0

Weitere Artikel der Ausgabe 1/2023

Journal of Cardiovascular Magnetic Resonance 1/2023 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.