Skip to main content
Erschienen in: Journal of Cardiovascular Magnetic Resonance 1/2019

Open Access 01.12.2019 | Research

Fast self-navigated wall shear stress measurements in the murine aortic arch using radial 4D-phase contrast cardiovascular magnetic resonance at 17.6 T

verfasst von: Patrick Winter, Kristina Andelovic, Thomas Kampf, Fabian Tobias Gutjahr, Julius Heidenreich, Alma Zernecke, Wolfgang Rudolf Bauer, Peter Michael Jakob, Volker Herold

Erschienen in: Journal of Cardiovascular Magnetic Resonance | Ausgabe 1/2019

Abstract

Purpose

4D flow cardiovascular magnetic resonance (CMR) and the assessment of wall shear stress (WSS) are non-invasive tools to study cardiovascular risks in vivo. Major limitations of conventional triggered methods are the long measurement times needed for high-resolution data sets and the necessity of stable electrocardiographic (ECG) triggering. In this work an ECG-free retrospectively synchronized method is presented that enables accelerated high-resolution measurements of 4D flow and WSS in the aortic arch of mice.

Methods

4D flow and WSS were measured in the aortic arch of 12-week-old wildtype C57BL/6 J mice (n = 7) with a radial 4D-phase-contrast (PC)-CMR sequence, which was validated in a flow phantom. Cardiac and respiratory motion signals were extracted from the radial CMR signal and were used for the reconstruction of 4D-flow data. Rigid motion correction and a first order B0 correction was used to improve the robustness of magnitude and velocity data.
The aortic lumen was segmented semi-automatically. Temporally averaged and time-resolved WSS and oscillatory shear index (OSI) were calculated from the spatial velocity gradients at the lumen surface at 14 locations along the aortic arch. Reproducibility was tested in 3 animals and the influence of subsampling was investigated.

Results

Volume flow, cross-sectional areas, WSS and the OSI were determined in a measurement time of only 32 min. Longitudinal and circumferential WSS and radial stress were assessed at 14 analysis planes along the aortic arch. The average longitudinal, circumferential and radial stress values were 1.52 ± 0.29 N/m2, 0.28 ± 0.24 N/m2 and  0.21 ± 0.19 N/m2, respectively. Good reproducibility of WSS values was observed.

Conclusion

This work presents a robust measurement of 4D flow and WSS in mice without the need of ECG trigger signals. The retrospective approach provides fast flow quantification within 35 min and a flexible reconstruction framework.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1186/​s12968-019-0566-z) contains supplementary material, which is available to authorized users.
Patrick Winter and Kristina Andelovic contributed equally to this work.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ApoE
Apolipoprotein E
CMR
Cardiovascular magnetic resonance
CSA
Cross-sectional area
ECG
Electrocardiogram
FLASH
Fast Low Angle Shot
MRI
Magnetic Resonance Imaging
NUFFT
Nonuniform Fast Fourier Transform
OSI
Oscillatory Shear Index
PC
Phase contrast
ROI
Region of interest
SNR
Signal to noise ratio
TEM
Transmit-receive electromagnetic
TOF
Time-of-flight
UTE
Ultrashort echo time
WSS
Wall Shear Stress
WT
Wild-type

Background

Cardiovascular diseases such as aortic valve disease, aneurysms and atherosclerosis are responsible for almost 25% of deaths in the US [1]. The hemodynamic environment is described by parameters such as flow and wall shear stress (WSS) and plays an important role in the development of these diseases [2, 3], since the complex vascular geometry and the pulsatile flow in the arterial system lead to regionally different flow characteristics and thus spatial and temporal changes in shear forces acting on the vessel wall [4]. The WSS is proportional to the spatial velocity gradient at the vessel wall and can therefore be assessed from the measured velocity fields. It is a vector quantity, which is expressed in N/m2 [5]. Shear stress can be dissected into a longitudinal part in parallel to the vessel and along the preferred flow direction and a circumferential component, which is perpendicular to the longitudinal component. In addition, normal stress components can occur when blood flow components pointing towards the surface normal of the vessel wall are present [6]. As the normal component is pointing in radial direction towards the center of the vessel [7], this component will be referred to as “radial stress” in the following. A further important parameter associated with the WSS is the oscillatory shear index (OSI), which describes the temporal variability of the WSS waveform. Both low WSS and high OSI values are potential markers for the formation of plaques in the aorta and other vessels [8]. WSS and OSI can be estimated non-invasively by measuring the flow velocities with phase contrast (PC) - cine cardiovascular magnetic resonance (CMR). Two-dimensional acquisitions provide localized information regarding hemodynamic forces and flow rates [9]. However, the difficulty in localizing a slice perpendicular to the flow direction in curved vessels such as the aortic arch can lead to experimental errors, especially near the aortic branches. Thus, a three-dimensional acquisition is needed to characterize the complete hemodynamic environment throughout the aortic arch. Conventional electrocardiogram (ECG)-triggered 3D measurements of flow and WSS however, are limited in spatiotemporal resolution due to long measurement times [7, 10]. This usually leads to an underestimation of WSS [5]. Measurements in mice are particulary challenging due to the small scale of the murine aorta and the high demands on animal handling, as unstable heart rates can impede the assessment of accurate flow waveforms. This becomes even more problematic at ultra-high-field strengths, where ECG-based navigation can become unreliable due to magnetohydrodynamic effects and interferences with the fast switching of the imaging gradients [11, 12]. In recent studies, the use of radial acquisitions for flow measurements in mice have been investigated [9, 13]. Radial trajectories are less prone to flow artifacts and provide an intrinsic cardiac motion signal, which can be used for self-gated ECG-free measurements [14]. The use of self-navigation for 3D flow measurements in the murine heart has already been investigated [15]. However, a measurement time of approximately 2 h was still needed for a complete dataset.
Time-of-flight (TOF) CMR techniques achieve a high blood-tissue contrast by exciting small image volumes using excitation pulses with large flip angles and small repetition times. This leads to a strong signal enhancement for the inflowing blood and a strong suppression of the static background. As a result, large undersampling factors are possible with 3D radial acquisitions [16].
In this work we propose an advanced method based on a self-navigated 3D radial PC-cine FLASH (Fast Low Angle SHot) acquisition, which exploits the inflow effect in order to achieve high resolution (isotropic 100 μm) flow measurements in the murine aortic arch in only 32 min. With this method, 3D flow velocities, aortic cross sectional areas, longitudinal, circumferential and radial stress components and the oscillatory shear index were determined.

Methods

Animal handling

All studies were conducted according to a protocol approved by the Institutional Animal Care and Use Committee. Female wild-type (WT) C57BL/6 J mice (n = 7) were obtained from Charles River Laboratories (Sulzfeld, Germany) and were studied at an age of 12 weeks. Mice were anesthetized with 4% isoflurane in 2.0 Vol.% oxygen (2 L/min), applied by a nose cone and were positioned vertically (head up). For cardiac and respiratory monitoring during the measurement, a pressure- sensitive pneumatic balloon (Graseby Medical Limited, Watford, United Kingdom) was placed between the inner radio frequency (RF) resonator wall and the murine thorax. The pressure signal from the balloon was transformed into an electrical signal by a pressure transducer (24PCEFA6 D, Honeywell S&C, Golden Valley, Minnesota, USA) and was amplified and processed in real-time by a custom-built ECG unit [17]. Due to the small inner diameter of the gradient insert and the RF coil, core body temperature could be maintained at physiological 37 °C during CMR-measurements by adjusting the temperature of the gradient cooling unit.

Data acquisition

Measurements were performed with a 17.6 T vertical-bore small animal MR system (Bruker Avance 750 WB, Bruker BioSpin MRI GmbH, Rheinstetten, Germany, operated with Paravision 4.0) with a 1 T/m gradient system (diameter: 40 mm) and a custom-built single-channel transmit-receive electromagnetic (TEM) resonator (inner diameter: 24 mm). To localize the position of the aortic arch, balloon-triggered axial and longitudinal 2D-cine FLASH measurements were acquired. Subsequently, retrospective flow measurements were performed with a radial PC-FLASH-sequence (see Fig. 1B) in a 3D slab perpendicular to the aorta (image volume: 25 × 25 × 4 mm3, see Fig. 1A). Spatial encoding was performed with a 3D radial trajectory with an angular density optimized for the anisotropic field of view [18], which was calculated with an open source tool box [19]. For flow-encoding a balanced 4-point flow-encoding scheme [20] with an encoding velocity of vENC = 125 cm/s was used. Each flow-encoding step consists of a read-out with 1.6 x 105 radial projections (140 read-out points, TR = 3 ms) covering a 3D-sphere in k-space (Fig. 1C). A flip angle of 15° was used in order to achieve high blood-tissue contrast. To minimize artifacts caused by off-resonances and signal dephasing due to accelerated flow, the echo time was set to 1.1 ms. In order to guarantee such a short echo time, it was necessary to design the amplitudes of the dephase gradients in a way that the gradient echo occurs at tE = 0.× tacq, where tacq is the acquisition time (Fig. 1B). To further increase robustness of data acquisition, the measurement was segmented into 10 subsets, each consisting of 1.× 104 projections (each covering a full 3D sphere in k-space) and 4 flow-encoding steps (Fig. 1D), which were acquired one at a time.
This kind of segmentation allows repetition of corrupted data sets in case of disturbances, e.g. caused by instabilities of the heart rate. Acquisition time of one subset was 3.2 min, leading to a total measurement time of 32 min for a full 4D flow protocol.

Phantom measurements

The stability of the 4D flow-encoding sequence was tested in a phantom consisting of a flow pump (MPC-Z V1.10, ISMATEC, Cole-Partner GmbH, Wertheim, Germany) with constant flow and adjustable flow values (max flow: 50 ml/s) and a silicone tube (ø = 6 mm). Flow was measured at 10 different flow values (15.00 ml/s - 26.25 ml/s in equidistant steps) with the protocol described above using only one subset per measurement (scan time: 3.2 min per subset) and the same encoding velocity as the in vivo measurements vENC = 125 cm/s. Mean flow values and standard deviations were calculated over 10 slices. To prevent artifacts due to phase aliasing, phase unwrapping was applied when necessary [21]. For comparison, flow was also quantified by measuring volumetrically in liters. The reference measurement was repeated 8 times and mean values and standard deviations were determined.

Self-navigation

All signal processing was performed with MATLAB (The Mathworks, Inc., Natick, Massachusetts, USA). For retrospective self-navigation, the magnitude value of the center k-space signal (k = 0) was used. First, high-frequency disturbances were removed by using a matched filter for low-pass filtering [22]. The matched filter can be interpreted as a convolution of the noisy navigator signal with a conjugated time-reversed small portion of the signal [23]. After filtering a baseline subtraction [24] was used in order to eliminate low-frequency modulations caused by respiratory motion and by the transient to the steady state.
Trigger points and breath gating intervals were determined with variable thresholds (Fig. 2A-C). Using a linear assignment, each read-out was afterwards allocated to a value between 0 and 1, corresponding to a phase in the cardiac cycle (Fig. 2D). For removal of corrupted data points due to respiratory motion, the time average of the trigger point intervals (i.e., the mean cardiac period) was calculated for all 40 subsets, respectively. Only trigger point intervals lying in a ± 4 × TR interval (±12 ms) window around the temporal average were accepted for reconstruction. For respiratory gating data points during inspiration were assigned to a cardiac phase value of −1. For reconstruction, the read-outs were divided into 20 selection intervals, corresponding to 20 cardiac phase intervals. For each selection interval, the associated projections were combined and an image was reconstructed using a nonuniform fast Fourier transform (NUFFT) with an open source software toolbox [25, 26]. In this manner a set of four 3D-cines (one flow-compensated cine and 3 flow-encoded cines) with 20 frames and an isotropic spatial resolution of 100 μm, respectively, was reconstructed.

Off-resonance correction

At high magnetic field strengths, B0 offsets and field gradients cause deviations of the radial trajectory, which can lead to severe blurring artifacts in the reconstructed images. To remove these artifacts, an additional flow compensated radial 3D FLASH measurement with two different echo times was performed in the same field of view (FOV) (tE1 = 1.3 ms, tE2 = 2.3 ms total measurement time: 3.2 min, spatial resolution: isotropic 100 μm). A 3D off-resonance map was calculated from the phase differences between the two images [27]:
$$ \Delta f\left(x,y,z\right)=\frac{\phi_2-{\phi}_1}{2\pi \cdot \left({t}_{E1}-{t}_{E2}\right)}, $$
(1)
where tE1,2 and φ1,2 are the echo times and phases of the two images. Assuming only a global field offset and constant field gradients and neglecting local field inhomogeneities, the spatial dependent off-resonance frequencies can be approximated as:
$$ \Delta f\left(x,y,z\right)\approx \Delta {f}_0+\alpha \cdot x+\beta \cdot y+\delta \cdot z. $$
(2)
To remove blurring artifacts induced by a global frequency offset ∆f0, the frequency value from the center of the FOV (x = y = z = 0) was taken and used for a phase correction of the MR signal:
$$ {S}^{\prime }(t)=S(t)\cdot {e}^{i2\pi \cdot \Delta {f}_0t}, $$
(3)
where S(t) denotes the raw uncorrected MR signal. To also correct deviations of the k-space trajectory caused by global field gradients, the corrected trajectory \( {\overrightarrow{k}}^{\prime }(t) \) used for re-gridding was calculated with:
$$ {\overrightarrow{k}}^{\prime }(t)=\overrightarrow{k}(t)+\left(\begin{array}{c}\alpha \\ {}\beta \\ {}\delta \end{array}\right)t, $$
(4)
where \( {\overrightarrow{k}}^{\prime }(t) \) denotes the undistorted radial trajectory. α, β and δ were determined through linear fits of the off-resonance map.

Rigid motion correction

Due to the vertical setup of the MR scanner, a slight slipping and shifting of the mouse was sometimes observed during the measurement. Since this would lead into undesired motion artifacts such as blurring and phase subtraction errors, a rigid motion correction was applied prior to the cine reconstructions.
First, the 10 measurement subsets described above were used for the reconstruction of 40 time-averaged 3D-images (4 encoders times 10 measurement subsets). Using the first image I1 as reference, the shifts x, y and z were calculated for each subsequent image In in order to minimize the error between these images (Fig. 3A):
$$ \Psi =\underset{x,y,z}{argmin}{\left\Vert {I}_1-{T}_{x,y,z}\cdot {I}_n\right\Vert}_2^2\kern2em n=2,3,...\mathrm{40.} $$
(5)
Hereby Tx,y,z denotes the translation operator in respect to the image coordinates x,y and z, which needs to be applied for minimization of the error between the first and the n − th image. The algorithm yields shift values on a time base of 48 s (Fig. 3B). This information was used for a phase correction of the signal in k-space using the Fourier Shift Theorem [28] prior to the reconstruction.

Image processing and segmentation

Depending on the slice orientation and the alignment of the aortic arch within the image volume, the phase accumulations induced by flow encoding can cause slight deformations of the waveforms of the self-gating signal. Due to these distortions, the cines of the 3 flow-encoders are sometimes temporally shifted against the flow-compensated cine. In order to correct these temporal shifts, the time-dependent image intensities averaged over one slice were compared against each other. By using cross-correlation [29], the temporal shifts were determined for each encoder and the cines were synchronized.
For segmentation of the aortic arch, an adapted version of the previously described semi-automatic segmentation technique [30] was used. This technique assumes that the segmentation of all 4 cines should in principle lead to the same number of identified pixels. By evaluating a cost function, an optimum threshold value corresponding to a minimum deviation between the flow encoding measurements can be derived. Using this technique, each slice (in z-direction) of the 3D cine was segmented independently. Slices near the aortic root were excluded due to strong signal cancellations induced by accelerated flow. Subsequently, the three velocity components (vx, vy, vz) were calculated from the phase differences between the cines. Using the segmentation data, velocity was afterwards zeroed outside the aorta and filtered with a spatial median filter with a 3-connectivity neighborhood inside the lumen [31]. The spatial median filter removes outliers of velocity values due to segmentation errors near the lumen boundaries but leaves velocity data within smooth regions inside the vessel untouched [7].

Calculation of WSS and OSI

Assuming a Newtonian and incompressible fluid, the general form of the WSS →τ can be written as [6]:
$$ \overrightarrow{\tau}=2\eta \overset{\cdot }{\varepsilon}\cdot \hat{n}, $$
(6)
where η denotes the viscosity of blood, \( \hat{n} \) the inward unit normal of the lumen surface and \( \overset{\cdot }{\varepsilon } \) the deformation tensor:
$$ {\overset{\cdot }{\varepsilon}}_{ij}=\frac{1}{2}\left(\frac{\partial {v}_j}{\partial {x}_i}+\frac{\partial {v}_i}{\partial {x}_j}\right),\kern2em i,j=1,2,3. $$
(7)
Hereby xi,j denotes the spatial coordinates and vi,j the velocity components.
To calculate the WSS and radial stress, the PC and segmentation data were imported into Ensight (CEI systems, USA). The velocity derivatives and the surface normals were calculated directly from the 3D velocities and the isosurface of the lumen segmentation using a custom-built Python script. For the blood viscosity, a value of η = 0.04 Pas was assumed [9].
To separate the two components of the WSS and the radial stress, a centerline of the aortic arch was calculated, as described by [7] (Fig. 4A). Ring segments perpendicular to the centerline were afterwards generated at 14 different locations of the aorta (Fig. 4B). The isosurface of the segmented lumen imported to Ensight consists of a grid with approx. 5 · 103 nodes. For each node of the surface grid, the longitudinal (\( \hat{l} \): in parallel to centerline), radial (\( \hat{r} \): pointing towards the centerline) and circumferential \( \left(\hat{c}=\hat{l}\times \hat{r}\ \right) \) unit vector was calculated. The WSS and radial stress can afterwards be separated using:
$$ {\tau}_l=\overrightarrow{\tau}\cdot \hat{l},\kern2em {\tau}_c=\overrightarrow{\tau}\cdot \hat{c},\kern2em {\tau}_r=\overrightarrow{\tau}\cdot \hat{r}. $$
(8)
Mean and median values of the three components were calculated for each ring segment and cardiac phase. In addition, temporal averaged WSS values \( \left(\overline{\tau \to \left(r,\to \right)}\right) \) were derived using:
$$ \overline{\tau \to \left(r,\to \right)}\mid =\frac{1}{T_{RR}}{\int}_0^{T_{RR}}\overrightarrow{\tau}\left(\overrightarrow{r},t\right) dt, $$
(9)
where \( \overrightarrow{\tau}\left(\overrightarrow{r},t\right)=\left[{\tau}_l(t)\kern0.5em {\tau}_c(t)\kern0.5em {\tau}_r(t)\right] \)are the time-dependent WSS and radial stress components and TRR is the cardiac period. To also measure the temporal variability of the WSS waveforms and the degree of oscillatory flow, the OSI was calculated from the time dependent stress values using [5]:
$$ \mathrm{OSI}=\frac{1}{2}\left(1-\frac{\mid {\int}_0^{T_{RR}}\overrightarrow{\tau}\left(\overrightarrow{r},t\right) dt\mid }{\int_0^{T_{RR}}\mid \overrightarrow{\tau}\left(\overrightarrow{r},t\right)\mid dt}\right). $$
(10)
No changes in the direction of the stress over time result in a minimal OSI value (OSI = 0). In contrast, when strong periodic variations and sign changes occur, e.g. caused by recirculative flow during the diastolic cardic phase, the integral value nears itself to the limit \( {\int}_0^{T_{RR}}\overrightarrow{\tau}(t) dt\to 0 \) and the OSI approximates its maximal value (OSI = 0.5).

Results

Phantom measurements

Figure 5A shows a correlation plot between the CMR flow measurement and the reference values of the flow phantom measurements, indicating a high correlation (r2 = 0.996) between both methods. The Bland-Altman Plot in Fig. 5B reveals only a small bias of the CMR measurement relative to the reference ( 0.15 ml/s).

Stability of self-navigation

Figure 6A shows an exemplary section of the self-navigation signal, measured in a representative mouse. The strong signal modulations due to cardiac and respiratory motion were used to calculate trigger points and breath-gating windows. To quantify the variance of the extracted trigger signals, mean values and variations of the cardiac periods were calculated for all individual measurements (Fig. 6B). No significant variations in quality between the 4 flow-encoding measurements were observed. The results of all 7 WT mice are displayed in Table 1. All animals showed a slight drift towards shorter cardiac periods during the measurement similar to Fig. 6B, which may be attributed to the adaption of the animal to the ambient temperature within the scanner and the anesthesia. However, the mean variation of the cardiac periods (\( \overline{\sigma} \) ≤ 4.3 ms) indicates a high stability of the self-navigation signal during the whole measurement in all mice. Depending on the heart and respiratory rate, 15–20% of the data was discarded from the measurement due to respiration. Approximately 13,000–15,000 heart beats were detected in each measurement, providing about 6000–7500 spokes for each cine frame in the retrospective cine reconstruction.
Table 1
Range of cardiac periods (rrmin-rrmax) detected by the algorithm, mean variance and number of detected heart beats N for all 7 wild-type mice
Animal
rrmin-rrmax [ms]
\( \overline{\sigma} \) [ms]
N
Mouse 1
99–147
3.3
15,120
Mouse 2
93–129
3.4
15,680
Mouse 3
84–123
4.3
15,261
Mouse 4
96–141
2.7
13,328
Mouse 5
87–123
3.4
13,896
Mouse 6
99–135
4.2
13,057
Mouse 7
99–138
4.3
12,851

Off-resonance correction

To demonstrate the effect of the off-resonance correction described above, magnitude images and velocity maps corresponding to the systolic cardiac phase are presented in Fig. 7. Uncorrected datasets result in severe blurring artifacts and distortions of the velocity data are recognizable (left), impeding further analysis of this data set. Using off-resonance correction, these artifacts can be removed (right).

Flow and cross-sectional areas

To visualize the measured flow through the aortic arch, a streamline representation of velocities corresponding to the systolic cardiac phase is shown for an exemplary mouse in Fig. 8A. Due to the geometry of the arch, a strong asymmetric distribution of velocities is recognizable with the highest values near the outer radius. For further analysis, the maximum cross-sectional area (CSA) of the lumen, the peak volume flow and the net flow were assessed at 14 analysis planes along the aorta (Table 2). The results indicate a narrowing of the lumen relative to the ascending aorta in the bifurcation region and the descending aorta. CSA values of more than 2.1 mm2 were found in the ascending aorta. This corresponds to approximately 210 voxels and, given a circular geometry, to 16 voxels across the diameter. In the descending aorta, the area reduces to 1.2 mm2, which corresponds to 12 voxels across the diameter. Regarding the volume flow, a decline with increasing distance to the aortic root was observed, likely due to the partial outflow into the major branches.
Table 2
Peak cross-sectional areas (CSA), differences between maximum and minimum CSA, peak volume flow and netflow for 14 analysis planes (see Fig. 4B). All data are presented as average values over all 7 wild-type mice
Region
peak CSA [mm2]
CSAmax-CSAmin [mm2]
peak flow [ml/s]
net flow [ml/s]
Ascending aorta
 1
1.96 ± 0.30
0.49 ± 0.08
0.88 ± 0.11
4.09 ± 0.77
 2
2.00 ± 0.24
0.51 ± 0.07
0.90 ± 0.11
4.22 ± 0.66
 3
2.06 ± 0.20
0.56 ± 0.06
0.90 ± 0.10
4.20 ± 0.58
 4
2.08 ± 0.14
0.55 ± 0.04
0.87 ± 0.10
4.06 ± 0.45
 5
2.10 ± 0.11
0.53 ± 0.05
0.83 ± 0.09
4.01 ± 0.36
 6
2.14 ± 0.10
0.52 ± 0.09
0.80 ± 0.09
3.92 ± 0.41
Bifurcation area
 7
1.60 ± 0.19
0.32 ± 0.05
0.58 ± 0.06
2.83 ± 0.35
 8
1.23 ± 0.10
0.33 ± 0.05
0.54 ± 0.03
2.37 ± 0.22
Descending aorta
 9
1.16 ± 0.07
0.30 ± 0.05
0.49 ± 0.02
1.91 ± 0.16
 10
1.18 ± 0.07
0.32 ± 0.07
0.47 ± 0.03
1.80 ± 0.18
 11
1.23 ± 0.06
0.37 ± 0.09
0.48 ± 0.02
1.78 ± 0.16
 12
1.28 ± 0.11
0.36 ± 0.14
0.47 ± 0.01
1.75 ± 0.13
 13
1.28 ± 0.10
0.34 ± 0.10
0.48 ± 0.03
1.76 ± 0.12
 14
1.30 ± 0.16
0.31 ± 0.08
0.50 ± 0.07
1.84 ± 0.18

Temporally averaged WSS

Figure 8B shows a map of the temporally averaged WSS magnitude values in a representative mouse. In Fig. 9, Bullseye-plots are displayed to illustrate the distribution of the time-averaged longitudinal, circumferential and radial stress values over the vessel cross-section. As expected, the strong velocity gradient due to the aortic geometry leads to much larger longitudinal WSS values near the outer radius in comparison to the inner radius. Furthermore, radial stress has its maximum magnitude values near the outer radius of the ascending aorta. In Fig. 10 the profiles of mean and median values of the temporally averaged longitudinal, circumferential and radial stress components are shown for 14 locations along the aorta (see also Fig. 4B and Table 3). The mean longitudinal WSS increases with rising distance to the aortic root, reaching its maximum values at the top of the aorta (with bold emphasis in Table 3). This effect is even more pronounced for the median values (with bold emphasis in Table 3), indicating a strong asymmetric distribution of longitudinal WSS values in the ascending aorta. In the descending aorta, the longitudinal WSS declines, reaching its minimum value at plane 12–13. For the circumferential WSS, the maximum values were found at analysis plane 5 in the ascending aorta (with bold emphasis in Table 3). The circumferential WSS remains at high levels until analysis plane 8. In the descending aorta, a drop and even a sign change (planes 11–14) is observable. For the radial stress, the highest values were found in the ascending aorta near analysis plane 4. An aligned behavior was observed for the magnitude of radial stress values and the maximum changes of CSAs occurring during the cardiac cycle (see Table 2 and Fig. 11A).
Table 3
Mean and median values of the temporally averaged longitudinal, circumferential and radial stress for 14 locations along the aortic arch (see Fig. 4B). All data are presented as average values over all 7 wild-type mice
Region
long. WSS
[N/m2]
circ. WSS
[N/m2]
rad. Stress
[N/m2]
Ascending aorta
mean
median
mean
median
mean
median
 1
1.44 ± 0.21
1.40 ± 0.23
0.28 ± 0.11
0.28 ± 0.11
 0.35 ± 0.25
 0.33 ± 0.22
 2
1.54 ± 0.15
1.54 ± 0.17
0.37 ± 0.11
0.35 ± 0.14
 0.40 ± 0.22
 0.35 ± 0.24
 3
1.60 ± 0.15
1.70 ± 0.22
0.45 ± 0.12
0.40 ± 0.08
 0.43 ± 0.21
 0.36 ± 0.24
 4
1.61 ± 0.17
1.73 ± 0.26
0.54 ± 0.10
0.52 ± 0.11
− 0.51 ± 0.20
− 0.48 ± 0.36
 5
1.65 ± 0.21
1.82 ± 0.32
0.59 ± 0.13
0.58 ± 0.19
 0.50 ± 0.21
 0.48 ± 0.23
 6
1.65 ± 0.15
1.83 ± 0.21
0.55 ± 0.17
0.55 ± 0.22
 0.46 ± 0.12
 0.39 ± 0.15
Bifurcation. area
mean
median
mean
median
mean
median
 7
1.67 ± 0.16
1.86 ± 0.12
0.51 ± 0.23
0.52 ± 0.18
 0.46 ± 0.25
 0.36 ± 0.31
 8
1.85 ± 0.17
1.87 ± 0.18
0.44 ± 0.19
0.45 ± 0.18
0.02 ± 0.14
0.02 ± 0.20
Descending aorta
mean
median
mean
median
mean
median
 9
1.60 ± 0.24
1.55 ± 0.26
0.15 ± 0.15
0.21 ± 0.19
−0.23 ± 0.17
0.04 ± 0.20
 10
1.36 ± 0.24
1.34 ± 0.29
0.03 ± 0.19
0.07 ± 0.21
0.33 ± 0.16
−0.03 ± 0.7
 11
1.18 ± 0.23
1.18 ± 0.26
−0.01 ± 0.20
0.02 ± 0.22
−0.06 ± 0.18
0.03 ± 0.15
 12
1.11 ± 0.18
1.13 ± 0.21
−0.03 ± 0.16
−0.03 ± 0.15
0.17 ± 0.15
0.08 ± 0.15
 13
1.05 ± 0.19
1.15 ± 0.18
−0.02 ± 0.10
−0.03 ± 0.10
0.15 ± 0.16
0.02 ± 0.23
 14
1.11 ± 0.22
1.19 ± 0.15
−0.02 ± 0.06
−0.02 ± 0.10
0.18 ± 0.10
0.12 ± 0.20

Time resolved WSS and OSI

In Figs. 12 and 13 the time resolved mean longitudinal, circumferential and radial stress values are displayed for all 14 regions along the aorta. All three components show a strong pulsatile behavior during the systolic cardiac phase. In Table 4 the peak stress and OSI values are shown. The peak longitudinal WSS reaches its highest value at analysis plane 8 at the top region of the aorta, while the largest radial and circumferential WSS values can be found in plane 5 and 7, respectively (see with bold emphasis in Table 4). A drop of peak circumferential and radial stress values in the descending aorta is apparent. Regarding the OSI, the largest values were found near the inner radius of the ascending aorta where the longitudinal WSS has the lowest values (Fig. 9). Larger values were detected in the descending aorta (8.9–12.3%) compared to the ascending aorta (4.2–5.8%) (see Fig. 11B).
Table 4
Peak longitudinal, circumferential and radial stress and oscillatory shear index (OSI) for all 14 analysis planes (see Fig. 4B). All data are presented as average values over all 7 wild-type mice
Region
peak long. WSS
[N/m2]
peak circ. WSS
[N/m2]
peak rad. Stress
[N/m2]
OSI
[%]
Ascending aorta
 1
6.± 1.1
1.13 ± 0.47
2.± 1.1
5.± 3.1
 2
6.59 ± 0.93
1.39 ± 0.49
2.± 1.2
4.± 2.6
 3
6.87 ± 0.83
1.71 ± 0.48
2.± 1.2
5.± 2.5
 4
6.90 ± 0.66
2.08 ± 0.47
 3.20 ± 0.92
4.± 1.7
 5
6.82 ± 0.68
2.01 ± 0.60
− 3.54 ± 0.96
4.± 1.2
 6
6.91 ± 0.71
2.01 ± 0.57
3.± 1.0
4.± 0.83
Bifurcation area
 7
7.93 ± 0.71
2.38 ± 0.49
2.± 1.0
5.± 2.6
 8
8.62 ± 0.61
2.10 ± 0.76
0.85 ± 0.30
7.± 1.8
Descending aorta
 9
8.18 ± 0.81
1.36 ± 0.61
 1.87 ± 0.80
8.± 3.3
 10
7.95 ± 0.75
1.00 ± 0.73
 1.11 ± 0.68
10.± 4.0
 11
7.66 ± 0.58
0.80 ± 0.46
 1.40 ± 0.63
10.± 2.9
 12
7.20 ± 0.45
0.63 ± 0.38
 1.40 ± 0.22
11.± 3.1
 13
7.09 ± 0.78
0.78 ± 0.26
 1.24 ± 0.41
12.± 2.9
 14
7.± 1.5
0.79 ± 0.40
 1.17 ± 0.72
12.± 2.8

Reproducibility

To test the reproducibility of the introduced method, 3 mice were measured twice within 7 days. Mean values and standard deviations (STD) of temporally averaged longitudinal and circumferential WSS as well as radial stress and OSI were calculated in 12 regions of interest (inner, outer, anterior and posterior radius in the ascending aorta, bifurcation area and descending aorta). Correlations between measurement 1 and 2, bias (mean of differences between measurement 1 and 2) and distribution (1.96 × STD of difference values) were determined with linear fitting and Bland-Altman plots (Table 5 and Additional file 1: Figures S1–S3). Reproducibility was found in particular for the longitudinal (r2 = 0.73–0.84) and circumferential (r2 = 0.63–0.74) WSS measurements.
Table 5
Reproducibility. Three mice were measured twice within 7 days. Longitudinal WSS, circumferential WSS and radial stress as well as OSI values were determined in 3 regions (ascending, bifurcation area, descending) and 4 sections, respectively. Correlation coefficient r2 (between measurement 1 and 2), bias (mean of difference between measurement 1 and 2) and scattering (1.96 SD of difference between measurement 1 and 2) were determined for all 4 measured variables. The corresponding plots can be found in Additional file 1: Figure S1–S3
Mouse
1
2
3
Longitudinal WSS
 Correlation r2
0.73
0.84
0.82
 Bias [N/m2]
0.004
0.27
0.38
 1.96· STD [N/m2]
0.807
0.65
0.53
Circumferential WSS
 Correlation r2
0.63
0.74
0.66
 Bias [N/m2]
0.08
0.02
0.03
 1.96· STD [N/m2]
0.43
0.65
0.50
Radial Stress
 Correlation r2
0.60
0.29
0.37
 Bias [N/m2]
0.11
0.21
0.006
 1.96· STD [N/m2]
0.82
0.65
1.38
OSI
 Correlation r2
0.72
0.22
0.15
 Bias
0.02
0.04
0.02
 1.96· STD
0.07
0.15
0.10

Subsampling

To investigate the influence of subsampling, one exemplary measurement (mouse 4 in Table 1) was reconstructed using 1, 2, 3 ... 10 subsets. Semiautomatic segmentation was performed for all 10 reconstructions. Mean values and standard deviations of temporally averaged longitudinal and circumferential WSS, radial stress and OSI were calculated for 12 regions of interest (see above) and all 10 reconstructions. The results are displayed in Additional file 1: Figures S4–S6. The strongest deviations relative to a full dataset are found when less than 6 subsets are used for reconstruction. To further assess the stability of the measurement and the measurement error, five 4D flow datasets were generated using different subsets (reconstruction 1: subsets 1–6, reconstruction 2: subsets 2–7, reconstruction 3: subsets 3–8, reconstruction 4: subsets 4–9, reconstruction 5: subsets 5–10). New lumen segmentations were generated by randomly combining the previous segmentations. Mean values and standard deviations were determined over 12 ROIs and the 5 datasets for all variables. The results are displayed in Table 6.
Table 6
Subsampling: Mean values and standard deviations, determined in 12 ROIs (ascending, top, descending with inner radius, outer radius, anterior radius, posterior radius, respectively) over 5 individual reconstructions in a representative mouse (mouse 4 in Table 1). Each 4D flow image was reconstructed from 6 subsets (reconstruction 1: subsets 1–6. reconstruction 2: subsets 2–7
Region
Longitudinal WSS
[N/m2]
Circumferential WSS
[N/m2]
Radial stress
[N/m2]
OSI
[%]
Ascending aorta
 inner radius
0.45 ± 0.15
0.15 ± 0.11
0.14 ± 0.10
27.05 ± 1.89
 outer radius
2.48 ± 0.09
0.51 ± 0.03
1.13 ± 0.19
5.74 ± 0.90
 anterior radius
1.01 ± 0.04
0.24 ± 0.06
0.63 ± 0.14
5.22 ± 4.09
 posterior radius
1.97 ± 0.08
0.19 ± 0.09
0.12 ± 0.09
7.92 ± 4.39
Bifurcation area
 inner radius
1.26 ± 0.07
0.05 ± 0.04
0.55 ± 0.10
10.38 ± 4.32
 outer radius
2.31 ± 0.14
0.94 ± 0.15
0.97 ± 0.23
7.44 ± 2.48
 anterior radius
2.32 ± 0.08
0.52 ± 0.12
0.73 ± 0.10
3.41 ± 2, 97
 posterior radius
1.39 ± 0.11
0.21 ± 0.06
1.17 ± 0.29
10.89 ± 6.69
Descending aorta
 inner radius
0.47 ± 0.06
0.15 ± 0.05
0.43 ± 0.04
26.07 ± 2.83
 outer radius
1.50 ± 0.07
0.17 ± 0.06
0.93 ± 0.09
8.35 ± 1.62
 anterior radius
1.50 ± 0.07
0.21 ± 0.05
0.30 ± 0.05
16.05 ± 2.00
 posterior radius
1.11 ± 0.02
0.01 ± 0.02
0.05 ± 0.05
13.35 ± 4.97

Discussion

Measurement time

In this work we introduce a robust self-navigated technique for fast measurements of flow and wall shear stress in mice using radial PC-cine CMR. The acquisition time needed for a dataset with isotropic 100 μm resolution, including the additional B0 measurement, was only 35 min. This corresponds to an effective acceleration of almost 3 compared to a cartesian ECG-triggered 3D flow-encoding. Our investigations of the influence of subsampling indicate that an even further reduction of measurement time to approx. 20 min using only 6 subsets might be achievable. However, to guarantee the highest sampling density available for all measurements, reconstructions were always performed with 10 subsets in this study. The shorter scan time and the use of self-navigation instead of external trigger signals facilitates animal handling. The extra time gained due to the acceleration could be spent to expand the imaging protocol, e.g. for additional vessel wall morphology measurements. Further acceleration might be possible when combining the radial acquisition with k − t acceleration techniques [30]. In this study, images were reconstructed at a high spatial resolution (100 μm3) and a moderate temporal resolution (20 frames/cardiac cycle). The cine reconstruction framework can easily be adapted to also investigate high dynamic flow variations, e.g. by sacrificing spatial resolution while increasing the frame rate.
A radial 4D flow-encoding technique based on UTE-sampling and self-navigation was previously proposed in [15]. The main difference between the UTE-based flow- encoding technique and our proposed method is the longer measurement time and the smaller spatial resolution. Coverage of the full murine heart at an isotropic spatial resolution of 160 μm required a measurement time of 1 h 58 min. The longer scan protocol was due to the fact that global excitation was used in order to guarantee short echo times. The technique presented in this work, in contrast, uses a slice-selective excitation pulse and benefits from the strong signal suppression of the static tissue. Less data acquisition is required, hence the measurement time can be significantly reduced.

Limitations

The proposed method strongly benefits from the ultrahigh magnetic field (17.6 T) and the high gradient strength (1 T/m), which allow high SNR values and short repetition times, both leading to shorter scan times. However, one limitation of the radial trajectory is its vulnerability to off-resonance effects, which can lead to severe blurring and distortion artifacts in the reconstructed magnitude images and velocity maps at ultrahigh field strengths. The first order B0 correction method presented in this work is computationally fast, easy to implement, and yields a significant improvement of image quality. However, in the lower part of the thoracic aorta close to the lung, large local field gradients can occur, which can lead to signal cancellations that cannot be reversed with the current method [30]. These artifacts lead to segmentation and phase errors, which can result in an underestimation of WSS values. However, with improved shimming and more advanced reconstruction techniques [32] it is feasible to reduce these artifacts. One further limitation is the susceptibility to signal cancellations caused by accelerated flow, which are more prominent at ultra high field strengths due to the larger local field gradients and especially present in the aortic root during the systolic cardiac phase. These artifacts can lead to an underestimation of flow values and hence to a possible underestimation of WSS values in the proximal part of the ascending aorta. In the presented method, we reduced flow artifacts by choosing a short TE (1.1 ms). We believe that it should be possible to reduce the vulnerability to flow by using ramp sampling [15] and slice-selective pulses for excitation. Furthermore the use of lower magnetic field strengths (7 T) in combination with cryogenic surface coils [10] should lead to a reduction of off-resonance and flow artifacts while still maintaining high SNR values.

Flow and WSS values

Measurements of volume flow were conducted in a flow phantom and a group of 7 healthy WT mice. Both in vitro and in vivo measurements are in very good accordance with reference measurements and the literature [3335].
WSS was directly derived by calculating the gradients of the measured 3D velocity field at the segmented lumen surface. As shown previously with a 3D spiral sequence in mice [7], a strong asymmetric distribution, yielding the highest values near the outer radius and the lowest values near the inner radius of the aortic arch, could be observed. Longitudinal and circumferential components of the temporally averaged WSS as well as radial stress were calculated at 14 sites along the aorta. The highest longitudinal WSS values were found in the top region of the arch, which is in accordance with the results reported previously [7]. In the descending aorta, the temporal averaged WSS values are in range with the values reported previously for 9-month-old WT mice (0.8–2.1 N/m2, see [9]). The lowest or even negative longitudinal WSS were found near the inner radius of the ascending aorta and the bifurcation area near the aortic branches, indicating low or even recirculative flow near these regions. For the distribution of mean circumferential WSS, a similar behavior was observed as reported for studies in humans [5, 8] and mice [7], yielding the highest values in the upper ascending aorta and the top region of the arch.
Investigations of repeatability showed that the measurement of both WSS components can be well reproduced. Possible causes of deviations might be differences in slice positioning, uncertainties in calculating an accurate centerline and B0 inhomogeneities. Overall, WSS values were larger than reported previously for 6-months old ApoE−/− mice [7] but still lower than reported for measurements using computational fluid dynamics for WSS calculation, most likely due to the still low spatial resolution [36]. Further improvements of the described technique should include the use of interpolation and CFD in order to reduce this error.
In 2017, Braig et al. presented an ECG-triggered cartesian sequence for preclinical measurements of the WSS in the murine aortic arch of 4-weeks-old WT mice [10]. The sequence enables 4D flow measurements within 40 min, however, the native spatial resolution was much lower (300 μm) and the evaluation workflow only considered the longitudinal WSS without taking into account circumferential WSS. Furthermore the method presented in this paper still required an ECG signal for navigation while our method enables completely wireless WSS measurements. Peak flow values were similar to those observed in our study (see Table 2). Regarding the peak longitudinal WSS, we measured higher values (6–8 N/m2) relative to the results presented by Braig et al. (4–6 N/m2), which might be due to the higher spatial resolution and the difference in age. Furthermore, we also observed an increase of the peak longitudinal WSS with rising distance from the aortic root. In accordance with the previous study, the highest peak longitudinal values were measured in the top region between the 2nd and 3rd bifurcation (see Figs. 12 and 13).
As byproduct of the calculation of the stress tensor, radial components pointing towards the vessel wall could be derived. An aligned relationship between the distribution of radial stress and the dilatation of aorta occurring during the cardiac cycle was observed. Since aortic stiffness has a large influence on the degree of aortic dilatation, the investigation of possible correlations between the magnitude of radial stress and the elasticity could be interesting. To our knowledge this is the first time that results for the radial component are reported for mice. However, since the radial stress values could not be as well reproduced, the true benefit of this parameter still needs to be investigated.

OSI

Regarding the OSI, the largest values were measured in the upper part of the descending aorta, indicating a larger amount of oscillatory and recirculative flow in these areas. The OSI has its highest values near the inner radius of the aortic arch, which is in agreement with the results reported for ApoE−/− mice [7]. OSI values were between 4.6% (ascending aorta) and 12.3% (descending aorta). Since the OSI usually has its highest values in regions where the WSS magnitude is low, the measurement of this parameter is more susceptible to SNR and could not be as well reproduced as the longitudinal and circumferential WSS measurements. The results, however, closely correspond with the values found in literature [7].

Conclusion

In summary, a robust accelerated measurement of flow and wall shear stress in the murine aortic arch was presented. The new method does not require ECG triggering and enables easier animal handling. The longitudinal, circumferential and radial component of the WSS and OSI values could be assessed. Future studies should focus on WSS measurements in atherosclerotic mouse models and possible correlations with pulse wave velocities and vessel wall morphology.

Acknowledgements

We thank Prof. Dr. Axel Friebe for his help in organizing wild-type mice for this study and Sabine Voll for the help in organizing the study.

Declaration

All experimental procedures were in accordance with institutional guidelines and were approved by local authorities (Government of Lower Franconia, Würzburg, Germany) to comply with German Animal Protection Law. Reference number of the authorization for animal experiments is 55.2–2531.01-427/17.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Kochanek KD, Murphy SL, Xu J, Tejada-Vera B. Deaths: final data for 2015. Natl Vital Stat Rep. 2016;65(4):1–122.PubMed Kochanek KD, Murphy SL, Xu J, Tejada-Vera B. Deaths: final data for 2015. Natl Vital Stat Rep. 2016;65(4):1–122.PubMed
2.
Zurück zum Zitat Knobelsdorff-Brenkenhoff Fv, Karunaharamoorthy A, Trauzeddel RF, Barker AJ, Blaszczyk D, Markl M, Schulz-Menger J. Evaluation of aortic blood flow and wall shear stress in aortic stenosis and its association with left ventricular remodeling. Circ Cardiovasc Imaging. 2016;9(3):1–22.CrossRef Knobelsdorff-Brenkenhoff Fv, Karunaharamoorthy A, Trauzeddel RF, Barker AJ, Blaszczyk D, Markl M, Schulz-Menger J. Evaluation of aortic blood flow and wall shear stress in aortic stenosis and its association with left ventricular remodeling. Circ Cardiovasc Imaging. 2016;9(3):1–22.CrossRef
3.
Zurück zum Zitat Ooij Pv, Markl M, Collins JD, Carr JC. Aortic valve stenosis alters expression of regional aortic wall shear stress. J AHA. 2017;6(9):1–13. Ooij Pv, Markl M, Collins JD, Carr JC. Aortic valve stenosis alters expression of regional aortic wall shear stress. J AHA. 2017;6(9):1–13.
4.
Zurück zum Zitat Cunningham KS, Gotlieb AI. The role of shear stress in the pathogenesis of atherosclerosis. Nature: Lab Invest. 2004;85:9–23. Cunningham KS, Gotlieb AI. The role of shear stress in the pathogenesis of atherosclerosis. Nature: Lab Invest. 2004;85:9–23.
5.
Zurück zum Zitat Stalder A, Russe M, Frydrychowicz A, Bock J, Hennig J, Markl M. Quantitative 2d and 3d phase contrast mri: optimized analysis of blood flow and vessel wall parameters. Magn Reson Med. 2008;60(5):1218–31.CrossRef Stalder A, Russe M, Frydrychowicz A, Bock J, Hennig J, Markl M. Quantitative 2d and 3d phase contrast mri: optimized analysis of blood flow and vessel wall parameters. Magn Reson Med. 2008;60(5):1218–31.CrossRef
6.
Zurück zum Zitat Emanuel G. Analytical fluid dynamics. 2nd ed. Boca Raton: CRC Press; 2000.CrossRef Emanuel G. Analytical fluid dynamics. 2nd ed. Boca Raton: CRC Press; 2000.CrossRef
7.
Zurück zum Zitat Janiczek RL, Blackman B, Roy R, Meyer CH, Acton S, Epstein FH. Three-dimensional phase contrast angiography of the mouse aortic arch using spiral mri. Magn Reson Med. 2011;66(5):1382–90.CrossRef Janiczek RL, Blackman B, Roy R, Meyer CH, Acton S, Epstein FH. Three-dimensional phase contrast angiography of the mouse aortic arch using spiral mri. Magn Reson Med. 2011;66(5):1382–90.CrossRef
8.
Zurück zum Zitat Frydrychowicz A, Stalder A, Russe M, Bock J, Bauer S, Harloff A, Berger A, Langer M, Hennig J, Markl M. Three-dimensional analysis of segmental wall shear stress in the aorta by flow-sensitive four-dimensional-mri. J Magn Reson Imaging. 2009;30(1):77–84.CrossRef Frydrychowicz A, Stalder A, Russe M, Bock J, Bauer S, Harloff A, Berger A, Langer M, Hennig J, Markl M. Three-dimensional analysis of segmental wall shear stress in the aorta by flow-sensitive four-dimensional-mri. J Magn Reson Imaging. 2009;30(1):77–84.CrossRef
9.
Zurück zum Zitat Zhao X, Pratt R, Wansapura J. Quantification of aortic compliance in mice using radial phase contrast mri. J Magn Reson Imaging. 2009;30(2):286–91.CrossRef Zhao X, Pratt R, Wansapura J. Quantification of aortic compliance in mice using radial phase contrast mri. J Magn Reson Imaging. 2009;30(2):286–91.CrossRef
10.
Zurück zum Zitat Braig M, Leupold J, Menza M, Russe M, Ko C-W, Hennig J, Elverfeldt D. Preclinical 4d-flow magnetic resonance phase contrast imaging of the murine aortic arch. PLoS One. 2017;12(11):1–4.CrossRef Braig M, Leupold J, Menza M, Russe M, Ko C-W, Hennig J, Elverfeldt D. Preclinical 4d-flow magnetic resonance phase contrast imaging of the murine aortic arch. PLoS One. 2017;12(11):1–4.CrossRef
11.
Zurück zum Zitat Polson MJR, Barker AT, Gardiner S. The effect of rapid rise-time magnetic fields on the ecg of the rat. Clin Phys Physiol Meas. 1982;3:231–4.CrossRef Polson MJR, Barker AT, Gardiner S. The effect of rapid rise-time magnetic fields on the ecg of the rat. Clin Phys Physiol Meas. 1982;3:231–4.CrossRef
12.
Zurück zum Zitat Li W, Griswold M, Yu Xa. Rapid t1 mapping of mouse myocardium with saturation recovery look-locker method. Magn Reson Med. 2010;64(5):1296–303.CrossRef Li W, Griswold M, Yu Xa. Rapid t1 mapping of mouse myocardium with saturation recovery look-locker method. Magn Reson Med. 2010;64(5):1296–303.CrossRef
13.
Zurück zum Zitat Winter P, Kampf T, Helluy X, Gutjahr FT, Meyer CB, Rommel E, Bauer WR, Jakob PM, Herold V. Fast retrospectively triggered local pulse-wave velocity measurements in mice with cmr-microscopy using a radial trajectory. J Cardiovasc Magn Reson. 2013;15(88):1–11. Winter P, Kampf T, Helluy X, Gutjahr FT, Meyer CB, Rommel E, Bauer WR, Jakob PM, Herold V. Fast retrospectively triggered local pulse-wave velocity measurements in mice with cmr-microscopy using a radial trajectory. J Cardiovasc Magn Reson. 2013;15(88):1–11.
14.
Zurück zum Zitat Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP. Self-gated cardiac cine MRI. Magn Reson Med. 2004;51(1):93–102.CrossRef Larson AC, White RD, Laub G, McVeigh ER, Li D, Simonetti OP. Self-gated cardiac cine MRI. Magn Reson Med. 2004;51(1):93–102.CrossRef
15.
Zurück zum Zitat Krämer M, Motaal A, Herrmann K-H, Löffler B, Reichenbach J, Strijkers GJ, Hoerr V. Cardiac 4d phase-contrast cmr at 9.4 t using self-gated ultra-short echo time (ute) imaging. J Cardiovasc Magn Reson. 2017;19(39):1–13. Krämer M, Motaal A, Herrmann K-H, Löffler B, Reichenbach J, Strijkers GJ, Hoerr V. Cardiac 4d phase-contrast cmr at 9.4 t using self-gated ultra-short echo time (ute) imaging. J Cardiovasc Magn Reson. 2017;19(39):1–13.
16.
Zurück zum Zitat Gu T, Kober F, Bock W, Fain S, Turk Q, Zhou Y, Grist T, Haughton V, Mistretta A. Pc vipr: a high-speed 3d phase-contrast method for flow quantification and high-resolution angiography. Am J Neuroradiol. 2005;26(4):743–9.PubMed Gu T, Kober F, Bock W, Fain S, Turk Q, Zhou Y, Grist T, Haughton V, Mistretta A. Pc vipr: a high-speed 3d phase-contrast method for flow quantification and high-resolution angiography. Am J Neuroradiol. 2005;26(4):743–9.PubMed
17.
Zurück zum Zitat Rommel E, editor. A double trigger unit for ECG and breath triggered mouse heart imaging. In: Proceedings of the Conference of the European Society of Magnetic Resonance in Medicine and Biology, vol. 17. Paris: European Society of Magnetic Resonance in Medicine and Biology; 2000. Rommel E, editor. A double trigger unit for ECG and breath triggered mouse heart imaging. In: Proceedings of the Conference of the European Society of Magnetic Resonance in Medicine and Biology, vol. 17. Paris: European Society of Magnetic Resonance in Medicine and Biology; 2000.
18.
Zurück zum Zitat Larson P, Gurney P, Nishimura D. Anisotropic field-of-views in radial imaging. IEEE Trans Med Imaging. 2008;27(1):47–57.CrossRef Larson P, Gurney P, Nishimura D. Anisotropic field-of-views in radial imaging. IEEE Trans Med Imaging. 2008;27(1):47–57.CrossRef
20.
Zurück zum Zitat Pelc N, Bernstein M, Shimakawa A, Glover G. Encoding strategies for three-direction phase-contrast mr imaging of flow. J Magn Reson Imaging. 1991;1(4):405–13.CrossRef Pelc N, Bernstein M, Shimakawa A, Glover G. Encoding strategies for three-direction phase-contrast mr imaging of flow. J Magn Reson Imaging. 1991;1(4):405–13.CrossRef
21.
Zurück zum Zitat Jenkinson M. Fast, automated, n-dimensional phase-unwrapping algorithm. Magn Reson Med. 2003;49(1):193–7.CrossRef Jenkinson M. Fast, automated, n-dimensional phase-unwrapping algorithm. Magn Reson Med. 2003;49(1):193–7.CrossRef
22.
Zurück zum Zitat Herold V, Winter P, Mörchel P, Gutjahr FT, Jakob PM. Free breathing self-gated pc-mri with pseudo random sampled kt-sparse-sense. Proc Int Soc Magnetic Reson Med Singapore. 2016;24:0466. Herold V, Winter P, Mörchel P, Gutjahr FT, Jakob PM. Free breathing self-gated pc-mri with pseudo random sampled kt-sparse-sense. Proc Int Soc Magnetic Reson Med Singapore. 2016;24:0466.
23.
Zurück zum Zitat Turin G. An introduction to matched filters. IRE Trans Inf Theory. 1960;6(3):311–29.CrossRef Turin G. An introduction to matched filters. IRE Trans Inf Theory. 1960;6(3):311–29.CrossRef
24.
Zurück zum Zitat Mazet V, Carteret C, Brie D, Idier J, Humbert B. Background removal from spectra by designing and minimising a non-quadratic cost function. Chemom Intell Lab Syst. 2005;76(2):121–33.CrossRef Mazet V, Carteret C, Brie D, Idier J, Humbert B. Background removal from spectra by designing and minimising a non-quadratic cost function. Chemom Intell Lab Syst. 2005;76(2):121–33.CrossRef
25.
Zurück zum Zitat Fessler J, Sutton B. Nonuniform fast fourier transform using minmax interpolation. IEEE Trans Signal Process. 2003;51(2):560–74.CrossRef Fessler J, Sutton B. Nonuniform fast fourier transform using minmax interpolation. IEEE Trans Signal Process. 2003;51(2):560–74.CrossRef
27.
Zurück zum Zitat Schneider E, Glover G. Rapid in vivo proton shimming. Magn Reson Med. 1991;18(2):335–47.CrossRef Schneider E, Glover G. Rapid in vivo proton shimming. Magn Reson Med. 1991;18(2):335–47.CrossRef
28.
Zurück zum Zitat Brown R, Haacke E, Cheng Y, Thompson M, Venkatesan R. Magnetic resonance imaging: physical principles and sequence design, vol. 2. Cleveland: Wiley; 2014.CrossRef Brown R, Haacke E, Cheng Y, Thompson M, Venkatesan R. Magnetic resonance imaging: physical principles and sequence design, vol. 2. Cleveland: Wiley; 2014.CrossRef
29.
Zurück zum Zitat Buck J, Daniel M, Singer A. Computer explorations in signals and systems using MATLAB, vol. 1. New Jersey: Prentice Hall; 1997. Buck J, Daniel M, Singer A. Computer explorations in signals and systems using MATLAB, vol. 1. New Jersey: Prentice Hall; 1997.
30.
Zurück zum Zitat Volker H, Herz S, Winter P, Gutjahr F, Andelovic K, Bauer WR, Jakob PM. Assessment of local pulse wave velocity distribution in mice using k-t blast pc-mri with semi-automatic area segmentation. J Cardiovasc Magn Reson. 2017;19(1):1-10. Article number 77. Volker H, Herz S, Winter P, Gutjahr F, Andelovic K, Bauer WR, Jakob PM. Assessment of local pulse wave velocity distribution in mice using k-t blast pc-mri with semi-automatic area segmentation. J Cardiovasc Magn Reson. 2017;19(1):1-10. Article number 77.
31.
Zurück zum Zitat Lim J. Two-dimensional signal and image processing. New Jersey: Prentice Hall; 1990. Lim J. Two-dimensional signal and image processing. New Jersey: Prentice Hall; 1990.
32.
Zurück zum Zitat Fessler J, Sutton B, Noll DC. Fast, iterative image reconstruction for mri in the presence of field inhomogeneities. IEEE Trans Med Imaging. 2003;22(2):178–88.CrossRef Fessler J, Sutton B, Noll DC. Fast, iterative image reconstruction for mri in the presence of field inhomogeneities. IEEE Trans Med Imaging. 2003;22(2):178–88.CrossRef
33.
Zurück zum Zitat Herold V, Parczyk M, Mörchel P, Ziener CH, Klug G, Bauer WR, Rommel E, Jakob PM. In vivo measurement of local aortic pulse-wave velocity in mice with mr microscopy at 17.6 tesla. Magn Reson Med. 2009;61(6):1293–9.CrossRef Herold V, Parczyk M, Mörchel P, Ziener CH, Klug G, Bauer WR, Rommel E, Jakob PM. In vivo measurement of local aortic pulse-wave velocity in mice with mr microscopy at 17.6 tesla. Magn Reson Med. 2009;61(6):1293–9.CrossRef
34.
Zurück zum Zitat Herold V, Wellen J, Ziener CH, Weber T, Hiller KH, Nordbeck P, Rommel E, Haase A, Bauer WR, Jakob PM, Sarkar S. In vivo comparison of atherosclerotic plaque progression with vessel wall strain and blood flow velocity in apoe−/− mice with mr microscopy at 17.6t. Magn Reson Mater Phys. 2009;22(3):159–66.CrossRef Herold V, Wellen J, Ziener CH, Weber T, Hiller KH, Nordbeck P, Rommel E, Haase A, Bauer WR, Jakob PM, Sarkar S. In vivo comparison of atherosclerotic plaque progression with vessel wall strain and blood flow velocity in apoe−/− mice with mr microscopy at 17.6t. Magn Reson Mater Phys. 2009;22(3):159–66.CrossRef
35.
Zurück zum Zitat Gotschy A, Bauer E, Schrodt C, Lykowsky G, Ye YX, Rommel E, Jakob PM, Bauer WR, Herold V. Local arterial stiffening assessed by mri precedes atherosclerotic plaque formation. Circ Cardiovasc Imaging. 2013;6(6):916–23.CrossRef Gotschy A, Bauer E, Schrodt C, Lykowsky G, Ye YX, Rommel E, Jakob PM, Bauer WR, Herold V. Local arterial stiffening assessed by mri precedes atherosclerotic plaque formation. Circ Cardiovasc Imaging. 2013;6(6):916–23.CrossRef
36.
Zurück zum Zitat Greve J, Les A, Tang B, Draney B, Wilson N, Dalman R, Pelc N, Taylor C. Allometric scaling of wall shear stress from mice to humans: quantification using cine phase-contrast mri and computational fluid dynamics. Am J Physiol Heart Circ Physiol. 2006;291(4):1700–8.CrossRef Greve J, Les A, Tang B, Draney B, Wilson N, Dalman R, Pelc N, Taylor C. Allometric scaling of wall shear stress from mice to humans: quantification using cine phase-contrast mri and computational fluid dynamics. Am J Physiol Heart Circ Physiol. 2006;291(4):1700–8.CrossRef
Metadaten
Titel
Fast self-navigated wall shear stress measurements in the murine aortic arch using radial 4D-phase contrast cardiovascular magnetic resonance at 17.6 T
verfasst von
Patrick Winter
Kristina Andelovic
Thomas Kampf
Fabian Tobias Gutjahr
Julius Heidenreich
Alma Zernecke
Wolfgang Rudolf Bauer
Peter Michael Jakob
Volker Herold
Publikationsdatum
01.12.2019
Verlag
BioMed Central
Erschienen in
Journal of Cardiovascular Magnetic Resonance / Ausgabe 1/2019
Elektronische ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-019-0566-z

Weitere Artikel der Ausgabe 1/2019

Journal of Cardiovascular Magnetic Resonance 1/2019 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.