Skip to main content
Erschienen in: The Ultrasound Journal 1/2020

Open Access 01.12.2020 | Letter to the Editor

Evaluation of a novel handheld point-of-care ultrasound device in an African emergency department

verfasst von: Samuel L. Burleson, Jonathan F. Swanson, Erin F. Shufflebarger, Douglas W. Wallace, Matthew A. Heimann, James C. Crosby, David C. Pigott, John P. Gullett, Maxwell A. Thompson, Christopher J. Greene

Erschienen in: The Ultrasound Journal | Ausgabe 1/2020

Abstract

Background

Many point-of-care ultrasound devices are now “pocket-sized” or handheld, allowing easy transport during travel and facilitating use in crowded spaces or in austere low-resource settings. Concerns remain about their durability, image quality, and clinical utility in those environments.

Method

Five emergency physicians with training in point-of-care ultrasound employed the Butterfly iQ, a novel handheld ultrasound device, in routine clinical care in a busy, high-acuity African emergency department over a period of 10 weeks. We retrospectively evaluated the performance of the Butterfly iQ from the perspectives of both the clinicians using the device and expert ultrasound faculty reviewing the images.

Results

We found advantages of the Butterfly iQ in a high-acuity African emergency department include its use of a single probe for multiple functions, small size, ease of transport, relatively low cost, and good image quality in most functions. Disadvantages include large probe footprint, lower, though still adequate, cardiac imaging quality, frequent overheating, and reliance on internet-based cloud storage, but these were surmountable. We also report a wide variety of patient presentations, pathology, and procedures to which the device was used.

Conclusion

We conclude the Butterfly iQ is an effective, though imperfect, point-of-care ultrasound device in a low-resource emergency setting. We will continue to employ the device in clinical emergency care and teaching in this setting.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ED
Emergency department
POCUS
Point-of-care ultrasound
RLS
Resource-limited settings

Introduction

Newly-marketed point-of-care ultrasound (POCUS) devices like the Butterfly iQ (iQ, Butterfly Network, Inc, Guilford, CT, USA) have generated significant excitement over their potential in emergency departments (EDs), critical care units, and resource-limited settings (RLS). Potential benefits of the iQ include small size, lower cost, integration with the user’s mobile phone or tablet, and use of silicon-chip based technology obviating the need for multiple transducers. There has been little critical evaluation of the device itself, particularly in RLS. We describe our use of the Butterfly iQ in routine clinical operations in a busy, high-acuity African ED and review its performance and applicability to RLS.

Personnel and setting

Five emergency physicians from the United States worked alongside the ED staff of a busy referral hospital in rural east Africa treating medical and surgical patients of all ages over a period of 10 consecutive weeks in the fall of 2019. All physicians have surpassed accepted POCUS training guidelines [1]. Two (SLB and JFS) have completed or are enrolled in a Point-of-care Ultrasound in Resource Limited Settings fellowship [2]. Advanced imaging is available, but access is limited by the need for payment prior to testing; in practice, patients often wait hours or days prior to imaging. Patients were selected by clinicians at the bedside if POCUS was indicated as a part of routine clinical care; therefore, informed consent was not required. Patients were scanned with the Butterfly iQ connected to an Apple iPad (iPad 5th Generation, Apple, Cupertino, CA, USA). Images were later reviewed for quality assurance by ultrasound fellowship-trained faculty. Final diagnoses were determined from a combination of the medical record, discussions with inpatient teams, and expert image review.

Results and discussion

The Butterfly iQ performed well and met clinician needs for a POCUS machine in this single RLS. Its advantages over cart-based machines are magnified, where financial resources, floor space, and reliable power may be scarce. Advantages and disadvantages of the Butterfly iQ are summarized in Table 1.
Table 1
Advantages and disadvantages of the butterfly iQ
ADVANTAGES
Single probe replaces multiple traditional transducers, capable of many scan types
Image quality excellent compared to other handhelds
Low cost
User-friendly app
DISADVANTAGES
Cardiac imaging lower quality than other modes
Relatively frequent overheating
Single probe relatively heavy with large footprint, occasionally compressing small structures

Advantages

The combination of flexibility and mobility of a single probe with preset modes replacing multiple transducers is the paramount benefit of the device. We successfully employed the iQ in a wide variety of scans and patient presentations (see Table 2, Figs. 1, 2, 3) and procedural guidance (see Table 3).
Table 2
Point-of-care ultrasound findings using Butterfly iQ in African Emergency Department
Age/sex
Presentation
US exams
US findings
44 yo M
Afib RVR, Cardiogenic shock
Cardiac
Calcified left atrial thrombus
3 yo M
Abdominal pain, Fussy, Bloody stool
GI
Intussusception
28 yo F
Submandibular swelling
MSK
Submandibular abscess
25 yo M
Dyspnea on exertion, Hx of PCE
Cardiac, Lung
Large PCE without tamponade, Bilateral pleural effusions
Young Adult M
Leg pain
MSK
Mid-shaft femur fracture with displacement
23 yo F
Abdominal pain, Hx of abdominal mass
GI
Dermoid cyst (recurrent)
78 yo M
Early satiety, Rectal mass on exam
GI
Novel dx of diffuse hepatosplenic lesions concerning for malignancy
28 yo M
Necrotic finger
MSK
Abscess vs necrosis
3 mo M
Acute respiratory failure
Cardiac, Lung
Novel dx of Atrial Septal Defect
114 yo M
Dyspnea
Cardiac, Lung
Novel dx of HFrEF
13 yo F
Novel Afib and hypoxia, Hx of RHD
Cardiac
Massive MR, LA dilation, Small PCE (Consistent with known RHD)
14 yo M
Dyspnea, Anasarca on exam, Hx of RHD and malaria
Cardiac
HFrEF, TR with RA dilation, MR (Consistent with known RHD)
24 yo M
Left flank pain
GI
Splenic lesion (Subcapsular hematoma vs Infiltration)
1 yo M
Hypoxia, fever, sepsis
Lung
Bilateral B-lines
1 yo F
Hypoxia, Hx of Ventricular Septal Defect
Cardiac, Lung
RV Dilation and hypertrophy, Persistent VSD, Bilateral B-lines
74 yo F
Respiratory arrest
Cardiac
Dilated/poorly contractile RV, Full IVC
2 mo F
Failure to thrive
CNS
Hydrocephalus
Young Adult M
Leg pain
MSK
Mid-shaft tibia fracture with displacement
Elderly Adult M
Recent DVT, Dyspnea
Cardiac
Right heart strain
20 yo F
2-week post-partum, Peritonitis on exam
GI
Pelvic free fluid
28 yo M
"Hematemesis" found to be hemoptysis, Hypoxia
Cardiac, Lung, GI
Splenic lesion, Bilateral B-lines, Normal LV Ejection Fraction
Elderly Adult M
Dyspnea, Anasarca on exam, Hypoxia
Cardiac, Lung
Dilated RA and RV, TR, Bilateral B-lines, Pleural effusions, Ascites
23 yo F
Suicide attempt by drowing, Third trimester pregnancy
OB
Normal Fetal heart rate and Fetal movement
7 yo M
Abdominal pain, vomiting
GI
SBO from worm burden
25 yo M
Dyspnea and chest pain with near-syncope, Novel Afib with RVR
Cardiac
Thickened Mitral Valve, Dilated LA, Massive MR (Suspected RHD)
5 yo M
Abdominal distention
GI
Enlarged bladder with mild hydronephrosis
3 yo F
Constipation, Reports of “worms in stool”
GI
SBO, No parasites visualized
85 yo M
RUQ abdominal pain and jaundice
GI
Intrahepatic biliary dilation, RUQ Mass
35 yo F
Chest pain, Hx of Tuberculosis
Lung
Loculated pleural effusion
37 yo M
Abdominal pain, Constipation
GI
SBO, Dilated gallbladder
76 yo M
Hypoxia and sepsis
Lung
Subpleural consolidations and B-lines
19 yo F
Hx of tamponade on outpatient echocardiogram
Cardiac
Large PCE, No tamponade
25 yo M
Hypoxia, Novel Afib with RVR
Cardiac
Massive LA dilation and MR, Thickened anterior mitral valve leaflet (Suspected RHD)
Afib, Atrial fibrillation; CNS, Central Nervous System; Dx, Diagnosis; HFrEF, Heart Failure with Reduced Ejection Fraction; GI, Gastrointestinal/Genitourinary; Hx, History; IVC, Inferior Vena Cava; LA, Left Atrium; MR, Mitral Regurgitation; MSK, Musculoskeletal/Soft tissue; OB, Obstetric; PCE, Pericardial Effusion; RA, Right Atrium; RHD, Rheumatic Heart Disease; RUQ, Right Upper Quadrant; RV, Right Ventricle; RVR, Rapid Ventricular Response; SBO, Small Bowel Obstruction; TR, Tricuspid Regurgitation
Table 3
Procedures guided by Butterfly iQ in an African Emergency Department
“Easy IJ”—peripheral IV cannula placed in internal jugular vein for short-term resuscitation [4]
Peripheral IV cannulation
Endotracheal tube placement [5]
Foreign body localization and removal
Paracentesis
Thoracentesis
Image quality, particularly in the abdominal and musculoskeletal modalities, was excellent, though not generally up to the standards of a cart-based system. With few exceptions (see below), image quality in all modalities was adequate to answer the clinical question.
The software application is user-friendly, allowing rapid alternation between preset scanning modes, enabling multiple protocols on the same patient with only a few gestures. Some routine calculation functions (such as gestational age) were not available during our experience with the device. We found the increased screen size of an iPad compared to a mobile phone beneficial for most uses.
The cost of the device (approximately US$2000), plus a required $420 annual subscription fee to the cloud-based image storage) places it within reach of some individual clinicians, and many healthcare institutions, even in RLS. Battery life was adequate, usually enough to last an entire 12-h shift on a full charge, though we did not specifically measure continuous scanning time. When necessary, we were able to charge the device from main power on shift. Small, relatively inexpensive, third party solar-powered chargers could also provide additional charging.

Disadvantages

Echocardiography habitually seemed lower quality than other modes. There was an apparent drop in resolution and frame rate, most noticeably associated with the use of color Doppler. Like other pocket-sized devices, the iQ also lacks spectral Doppler. Despite these limitations, we were able to detect significant cardiovascular pathology. All views were generally obtainable, adequate to guide resuscitation, and answer basic clinical questions pertinent to the RLS [3].
Images are stored on a cloud-based server requiring internet access to upload. Until they are uploaded, the images remain in an "outbox", where they can easily be deleted. Without reliable internet access, many images remained in this "outbox” for the duration of our field work, inhibiting our ability to share or review images.
We encountered several challenges related to the device’s hardware. The first was periodic overheating, rendering further scanning impossible until the device cooled. Overheating was not appreciably tied to any specific scanning mode or function. We mitigated this by briefly running water over the waterproof end of the device until cooled. We noticed a small rubber seal loosening near the end of the transducer by the closing of our field experience, without any discernible change in function. We found the cord length (125 cm) slightly short, especially when performing POCUS-guided procedures. The probe itself weighs 0.7 lb (0.3 kg), more than twice most other transducers, and its footprint is larger than a phased array probe, which was occasionally problematic when placing ultrasound-guided peripheral IVs and scanning between ribs, respectively. This may have contributed to the decreased quality of some echocardiographic images.

Limitations

These findings represent a retrospective review of the authors’ personal experiences with the Butterfly iQ device during routine clinical work in an African ED in an attempt to evaluate its performance in RLS. Patients were scanned at the discretion of the clinician at the bedside or because of restricted access to other diagnostics, introducing the possibility of selection bias. No patient-oriented outcomes were assessed and no comparisons between devices were available.

Conclusion

The Butterfly iQ was employed in a wide variety of patient presentations, scanning indications, and procedural guidance in a busy, high acuity RLS ED. Advantages include its use of a single probe for multiple functions, small size, relatively low cost, and good image quality in most functions. Disadvantages include large probe footprint, lower, though adequate, cardiac imaging quality, frequent overheating, and reliance on internet-based cloud storage, but these were surmountable. We believe the iQ is an effective POCUS device for emergency care in the RLS and we will continue to employ it for patient care and clinical teaching.

Acknowledgements

Not applicable.
Not applicable. This manuscript is not a study, but a documentation and evaluation of an-already FDA-approved device in routine clinical care.
Not applicable. Consent for publication was not obtained for individual patients. All ultrasound images were obtained as part of routine clinical care, images are de-identified, no specific details or identifying information from the case is presented, and no photos are included.

Competing interests

No authors have anything to disclose or any financial conflicts of interests. Specifically, no authors have any connection with Butterfly Network, Inc.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
2.
Zurück zum Zitat Burleson SL, Pigott DC, Gullett JP et al (2020) Point-of-care ultrasound in resource-limited settings: the PURLS fellowship. Ultrasound J 12:14CrossRef Burleson SL, Pigott DC, Gullett JP et al (2020) Point-of-care ultrasound in resource-limited settings: the PURLS fellowship. Ultrasound J 12:14CrossRef
3.
Zurück zum Zitat Huson MAM, Kaminstein D, Kahn D et al (2019) Cardiac ultrasound in resource-limited settings (CURLS): towards a wider use of basic echo applications in Africa. Ultrasound J 11:34CrossRef Huson MAM, Kaminstein D, Kahn D et al (2019) Cardiac ultrasound in resource-limited settings (CURLS): towards a wider use of basic echo applications in Africa. Ultrasound J 11:34CrossRef
4.
Zurück zum Zitat Zitek T, Buby E, Hudson H et al (2018) Ultrasound-guided placement of single-lumen peripheral intravenous catheters in the internal jugular vein. West J Emerg Med 19(5):808–812CrossRef Zitek T, Buby E, Hudson H et al (2018) Ultrasound-guided placement of single-lumen peripheral intravenous catheters in the internal jugular vein. West J Emerg Med 19(5):808–812CrossRef
Metadaten
Titel
Evaluation of a novel handheld point-of-care ultrasound device in an African emergency department
verfasst von
Samuel L. Burleson
Jonathan F. Swanson
Erin F. Shufflebarger
Douglas W. Wallace
Matthew A. Heimann
James C. Crosby
David C. Pigott
John P. Gullett
Maxwell A. Thompson
Christopher J. Greene
Publikationsdatum
01.12.2020
Verlag
Springer International Publishing
Erschienen in
The Ultrasound Journal / Ausgabe 1/2020
Print ISSN: 2036-3176
Elektronische ISSN: 2524-8987
DOI
https://doi.org/10.1186/s13089-020-00200-8

Weitere Artikel der Ausgabe 1/2020

The Ultrasound Journal 1/2020 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Bei seelischem Stress sind Checkpoint-Hemmer weniger wirksam

03.06.2024 NSCLC Nachrichten

Wie stark Menschen mit fortgeschrittenem NSCLC von einer Therapie mit Immun-Checkpoint-Hemmern profitieren, hängt offenbar auch davon ab, wie sehr die Diagnose ihre psychische Verfassung erschüttert

Antikörper mobilisiert Neutrophile gegen Krebs

03.06.2024 Onkologische Immuntherapie Nachrichten

Ein bispezifischer Antikörper formiert gezielt eine Armee neutrophiler Granulozyten gegen Krebszellen. An den Antikörper gekoppeltes TNF-alpha soll die Zellen zudem tief in solide Tumoren hineinführen.

Erhebliches Risiko für Kehlkopfkrebs bei mäßiger Dysplasie

29.05.2024 Larynxkarzinom Nachrichten

Fast ein Viertel der Personen mit mäßig dysplastischen Stimmlippenläsionen entwickelt einen Kehlkopftumor. Solche Personen benötigen daher eine besonders enge ärztliche Überwachung.

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.