Background
Hypertrophic cardiomyopathy (HCM), characterized by an unexplained increase in left ventricular (LV) wall thickness, is the most common genetic cardiac disorder, with a prevalence of approximately 1 in 500; this prevalence may be as high as 1 in 200 when accounting for both genotype-positive/phenotype-positive and genotype-negative/phenotype-positive individuals [
1]. Typical pathologic findings of HCM include cardiomyocyte hypertrophy and disarray, as well as focal or diffuse interstitial fibrosis [
2]. In recent years, cardiovascular magnetic resonance (CMR) has been used to characterize and quantify myocardial fibrosis. Increased fibrosis, seen as late gadolinium enhancement (LGE), has been identified as a risk factor for sudden cardiac death and heart failure in this population [
3]. T1 mapping and extracellular volume (ECV) quantification through CMR have also been correlated with increased risk of cardiovascular events [
4,
5]. However, not all HCM patients will go on to have an event; LGE has a high prevalence (as high as 70%) in this population [
6,
7] but a low specificity for the prediction of future cardiovascular events, limiting its negative predictive value [
8]. Additionally, gadolinium-based contrast agents (GBCAs) confer a risk of nephrogenic systemic fibrosis in patients with renal disease, and additionally are deposited in brain tissue [
9,
10]. Accordingly, there is interest in the development and validation of more specific and non-contrast methods for myocardial characterization in HCM patients.
T1ρ CMR is an endogenous contrast method for tissue characterization that does not require GBCAs and is distinct from both T1 and T2 contrast. It utilizes a low power radiofrequency pulse, also called a spin-lock pulse, to enable measurement of longitudinal relaxation in the rotating frame (T1ρ). The spin lock pulse mitigates the loss of transverse magnetization, suppressing contributions to relaxation from chemical exchange and water diffusion through magnetic field gradients [
11]. Its ability to detect myocardial fibrosis has been validated in animal models of ischemia and reperfusion [
12‐
14] as well as in explanted hearts from patients with dilated cardiomyopathy [
15]. Despite its mechanistic relevance to HCM pathophysiology, few studies have investigated the value of T1ρ in this population. Thus, we sought to evaluate and characterize the role of T1ρ in HCM patients by comparing it to conventional LGE and native T1.
Methods
Study population
We prospectively enrolled HCM patients between August 10, 2015 and July 10, 2017 as part of the Novel Markers of Prognosis in Hypertrophic Cardiomyopathy (HCMR) study. Detailed trial inclusion and exclusion criteria have been previously published [
16]. In brief, key inclusion criteria were patients aged 18–65 years with an established HCM diagnosis defined as unexplained myocardial hypertrophy of ≥ 15 mm without cavity dilation, etiologies such as hypertension and aortic stenosis, or other infiltrative cardiomyopathies such as amyloidosis and sarcoidosis. Additional exclusion criteria were: (1) prior septal myectomy or alcohol septal ablation, (2) prior myocardial infarction or coronary artery disease, (3) incessant ventricular arrhythmias, (4) inability to lie flat, (5) contraindications to CMR including pacemakers, defibrillators, intraocular metal, certain types of intracranial aneurysm clips, severe claustrophobia, and stage IV/V chronic kidney disease with estimated glomerular filtration rate < 30 mL/min/1.73 m
2, (6) diabetes mellitus with end organ damage, (7) pregnancy, and (8) inability to provide informed consent. In addition, we recruited 10 healthy subjects without cardiovascular risk factors or diseases and on no medications to serve as a control group. The study protocol was approved by the Institutional Review Board of the University of Pennsylvania and all subjects gave written informed consent prior to enrollment.
CMR imaging
CMR was performed using a 1.5 T CMR scanner (Avanto; Siemens Healthineers; Erlangen, Germany), equipped with 18 channel anterior and posterior array coils. Retrospectively gated, short axis, multi-slice cine CMR was performed with a temporal resolution = 34–40 ms, flip angle = 70°, bandwidth = 940 Hz/pixel, spatial resolution = 1.8 × 1.8 mm2, slice thickness = 8 mm.
2D T1ρ breath-held single-shot balanced steady-state free precession (bSSFP) sequences were performed at 3 short axis slice positions for HCM patients (apical, mid, and basal) in systole and 2 short axis slice positions for controls (mid and basal) using a motion- and heart rate-corrected spin echo, spin lock (SL) T1ρ pulse cluster (90
x—SL
y—180
y—SL
-y—90
-x) at end-systole [
17‐
19]. T1ρ images were acquired with different SL times (TSL) using the following parameters: TSL = 2, 10, 18, 26, 34, 42, 50 ms, B
1 = 400–500 Hz, spatial resolution = 1.4 × 1.4 mm
2, slice thickness = 8 mm, flip angle = 70°, echo time (TE) = 1.45 ms, repetition time (TR) = 2.9 ms, number of segments (N
Seg) = 55, bandwidth = 900 Hz/pixel, linear k-space phase encoding ordering, parallel imaging with acceleration factor = 2, 34 reference k-space lines obtained in a separate heartbeat, and allowing 1 additional heartbeat for T1 relaxation between shots. The T1ρ pulse amplitude was set at the highest available within scanner specific absorption rate (SAR) limits (B
1 = 400–500 Hz). Motion correction was used to reduce residual cardiac and respiratory motion between T1ρ images (Equation [
1]). The relaxation rate
\(R1\rho =\frac{1}{T1\rho }\) and intercept
\(B\) were estimated by two-parameter fit
$$\min _{{R1\rho ,B}} \left\| {\ln (S_{i} ) - B + R1\rho \cdot TSL_{i}} \right\|_{{2}}^{2}$$
(1)
where
\({S}_{i}\) is the magnitude signal at each spin lock duration
\({TSL}_{i}\). Motion correction and parametric mapping (Eq [
1]) were implemented using custom C + + software on the CMR scanner [
17].
2D T1 images were obtained with a breath-held shortened modified Look-Locker inversion recovery (ShMOLLI) [
20] sequence at 3 short axis slice positions matched to T1ρ at mid-end-diastole [
21]. Other parameters were: spatial resolution = 1.4 × 1.4 mm
2, slice thickness = 8 mm, flip angle = 35°, TE = 1.2, TR = 2.4 ms, N
Seg = 57, bandwidth = 1080 Hz/pixel, linear k-space encoding, parallel imaging acceleration factor = 2, 34 reference k-space lines obtained in a separate heartbeat. These images were prospectively electrocardiogram gated.
A 0.15 mmol/kg intravenous injection of gadolinium-based contrast was used for LGE imaging (Magnevist; Bayer Schering Pharma; Leverkusen, Germany). Imaging was performed 15–20 min after injection of contrast agent using an inversion time (TI) scout sequence to determine the TI to null myocardial tissue signal. LGE CMR was obtained using a 2D segmented phase-sensitive inversion recovery (PSIR) sequence at spatial resolution = 1.2 × 1.2 mm
2, flip angle = 50°, TE = 1.6 ms, TR = 3.2 ms, slice thickness = 8 mm, and parallel imaging acceleration factor = 2 [
22].
Image analysis
Cardiac function
Cardiac volumes and functional data were analyzed on the short-axis cine images using a commercially available software (Suiteheart, Neosoft, Pewaukee, Wisconsin, USA) The endocardium and epicardium were automatically traced at end-diastole and end-systole and manually adjusted following Society for Cardiovascular Magnetic Resonance guidelines [
23]. Papillary muscles were included in the ventricular volume.
Presence of enhancement on LGE
All LGE images were anonymized, shuffled, and presented to 2 blinded expert readers (B.D. and H.L., each with > 10 years of CMR experience), who labeled each slice as showing positive visible enhancement or not. Slices were labeled as showing positive (++) or negative enhancement (–) if both experts agreed, and otherwise were labeled equivocal (+ −).
Determination of myocardial relaxation times, scar size, and ECV
Relaxation times were measured in pre-contrast T1, post-contrast T1, and T1ρ images by manual contouring of the LV myocardium using QMass (Medis, Leiden, Netherlands). In LGE, T1, and T1ρ images, enhancement area was quantified using full width at half maximum (FWHM) thresholding and reported as the ratio of enhanced to total LV area (%). ECV was calculated per Equation [
2] using blood and entire myocardial T1 values, and hematocrit (Hct) obtained within 24 h of CMR [
24].
$$ECV = 100\% \times \left( {1 - Hct} \right) \times \frac{{1/Myocardial\,T1_{post - contrast} - 1/Myocardial\,T1_{pre - contrast} }}{{1/Blood\,T1_{post - contrast} - 1/Blood\,T1_{pre - contrast} }}$$
(2)
Statistical analysis
Statistical analysis was performed using R 3.6.1 (R Foundation for Statistical Computing, Vienna, Austria) and MATLAB R2019b (The MathWorks Inc., Natick, Massachusetts, USA). Categorical variables are expressed as N (%); continuous variables are expressed as mean ± SD or median [interquartile range (IQR)] depending on the distribution of the data. Normality testing was performed using the Shapiro–Wilk test. If the data were normally distributed, parametric methods were used, otherwise non-parametric methods were used. Student’s t-test, Wilcoxon Signed Rank test, one-way analysis of variance (ANOVA), and Kruskal–Wallis test (with post-hoc Dunn test adjusted with the Benjamini–Hochberg method) were used as appropriate based upon the variables and data distribution. To compare proportions of categorical variables, Chi-square test and Fisher’s exact test were used, as appropriate. The correlation between T1ρ and other parameters was assessed using Pearson’s and Spearman’s correlation coefficients, as appropriate. p values less than 0.05 were considered statistically significant.
Discussion
In our study characterizing the role of endogenous T1ρ imaging in the assessment of patients with HCM, we found that (1) percent area enhancement as measured by T1 and T1ρ at FWHM were moderately correlated with LGE area enhancement, (2) HCM short-axis slices categorized as LGE + +, LGE + −, and LGE −− each demonstrated elevated pre-contrast T1, T1ρ, and ECV compared to controls, and (3) ECV was significantly different between images rated LGE + + compared to LGE −−.
Both T1- and T2-weighted imaging have been used to demonstrate elevations in HCM patient myocardial relaxation times relative to normal patients [
5,
25‐
29]. Cardiac T2 mapping may be sensitive to several different mechanisms of relaxation in vivo. Some of these mechanisms may be considered ‘undesired’ because they suppress ∆T2 between diseased and healthy myocardium. Since each mechanism of relaxation is additive to the overall relaxation rate (i.e.,
\({R}_{2}={R}_{2,a}+{R}_{2,b}+\dots\), where
\(a\),
\(b\), and so on refer to a different relaxation mechanism), eliminating these ‘undesired’ sources of relaxation could increase the difference in the net transverse relaxation. While the ‘unwanted’ contributions to T2 in myocardium are not fully elucidated at present, their effect is to dephase magnetization irreversibly. Potential ‘undesired’ mechanisms of relaxation may include diffusion through background magnetic fields, chemical exchange, among others. By using a sufficiently strong SLk pulse, it is possible to prevent these unwanted mechanisms of relaxation [
11]. Using a moderate amplitude (> 400 Hz) SL pulse, we have found that there is a significantly larger ∆T1ρ than ∆T2 in these regions [
30]. The net effect of this is an increase in the contrast between normal and diseased myocardium.
Patchy fibrosis occurs in the majority of HCM patients. This is observed primarily as replacement fibrosis, but may also take the form of interstitial fibrosis, which can be imaged and quantified by T1 mapping and subsequent ECV calculation [
31,
32]. Most studies of fibrosis in HCM patients have focused on LGE imaging, which allows visualization of replacement fibrosis and has demonstrated associations with adverse outcomes [
3]. However, fibrosis accumulates throughout the course of HCM, and additionally, LGE has limited specificity for the prediction of events such as sudden cardiac death and heart failure [
8]. It is therefore of both clinical and research interest to investigate new contrast mechanisms such as T1ρ in the HCM population.
To date, only one study has measured T1ρ in human patients with HCM; Wang et al. compared visually-assessed LGE area with 2–6 standard deviation-thresholding of T1ρ in 18 HCM patients, finding high correlation (Pearson’s r ranging from 0.81 to 0.88) of percent fibrosis between these modalities [
33]. In our cohort, we found a lower correlation of T1ρ with LGE-assessed enhancement area using Spearman’s rho, which may be due to several reasons. Our cohort is larger with 40 HCM patients and is more heterogenous with both genotype-positive and -negative patients. Additionally, our group applied FWHM thresholding to LGE images, rather than manual measurement of enhancement area, decreasing observer bias. The use of FWHM thresholding therefore increases the robustness of our measurements, allowing for direct comparison in future studies. An additional study of T1ρ in a mouse model of cardiac hypertrophy [
34] examined T1ρ at several timepoints after transverse aortic constriction and verified fibrosis ex vivo using Masson’s trichrome staining [
34]. Similarly, their findings showed that T1ρ increased over time and was highly correlated with fibrotic areas [
34].
Our study brings to light several interesting findings. We show moderate correlations between LGE and T1 and T1ρ-assessed percent enhancement area, and mild correlation between T1 and T1ρ. Variations in the enhancement areas calculated by each method may reflect a physiologic difference in the way that LGE, T1, and T1ρ assess healthy and abnormal tissue. Our results indicate that LGE, T1, and T1ρ may each give different and additive information that one method alone cannot provide, a finding that warrants further study. Additionally, we demonstrate that HCM patients showed elevations in non-contrast quantitative MR measurements (pre-contrast T1 and T1ρ) regardless of LGE status. The significance of T1ρ imaging and its added value will need to be prospectively evaluated.
Limitations
Several limitations to our study should be acknowledged. Our cohort was small; thus our findings require validation and further investigation in larger groups of patients. Given the low annual cardiovascular event rate in patients with HCM, longer term follow-up will be needed to understand the utility of T1ρ in the assessment of patients with HCM.
Conclusions
T1 and T1ρ relaxation time moderately correlate with LGE percent enhancement area using FWHM thresholding. Additionally, T1, T1ρ, and ECV distinguish HCM patients from healthy controls, irrespective of whether the patient’s myocardium demonstrated positive LGE, showing potential value as a noninvasive biomarker. Further study is needed to elucidate the role of T1ρ in risk prediction for HCM patients.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit
http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (
http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.