Skip to main content
Erschienen in: Journal of Robotic Surgery 1/2024

Open Access 01.12.2024 | Research

Effectiveness and safety of SR-ENS-600endoscopic surgical system in benign and malignant gynecological diseases: a prospective, multicenter, clinical trial with 63 cases

verfasst von: Ren Chang, Duan Ping, Shen Yang, Wang Yongjun, Zhang Wei, Zheng Ying, Li Xingming, Zhang Kexin, Sun Dawei

Erschienen in: Journal of Robotic Surgery | Ausgabe 1/2024

Abstract

Single-port laparoscopy has gained more attention, but inherent technical challenges hinder its wider use. To overcome the disadvantage of traditional single-port surgery, robotic laparoendoscopic single-site surgery system was designed and clinically utilized. This multi-center single-arm trial was aimed to present the clinical outcomes of the SHURUI robotic endoscopic single-site surgery system. 63 women with ovary cysts, myoma, cervical epithelial neoplasm, or endometrial carcinoma were recruited at 6 academic medical centers in different districts of China. The trial was registered on September 5, 2023, with the register number: ChiCTR2300075431, retrospectively registered. Patients underwent robotic LESS surgery with the SHURUI endoscopic surgical system from January 17 to May 26, 2023. Demographic information, perioperative parameters, complications, scar healing, and operator satisfaction scores were recorded. Patients were followed up for 30 ± 4 days. Average operative time and estimated blood loss were 157.03 ± 75.24 min and 63.86 ± 98.33 ml, respectively, for all surgeries. Average anal exhaust time and hospitalization stay were 30.99 ± 14.25 h and 3.63 ± 1.59 days, respectively. Patients’ postoperative rehabilitation assessment showed satisfactory results on the day of discharge and 30 ± 4 days after surgery. The surgery achieved good cosmetic benefits and was surgeon friendly. There were no conversions to alternative surgical modalities, complications, or readmissions. The SHURUI endoscopic surgical system showed both the technical feasibility and safety of this surgical modality for gynecologic patients. Further randomized studies comparing this modality with traditional LESS surgery are suggested.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Within the last 10 years, significant advances in minimal-invasive surgery have been reported. Due to the advantages of improved cosmesis, fewer complications, less pain, shorter hospital stays, and lower health care costs, single-port laparoscopy has gained increasing attention. Sixty-four percent of gynecological patients preferred the appearance of a single-port laparoscopic scar when surveyed [1]. However, the operation is challenging for surgeons owing to technical and ergonomic difficulties, including loss of instrumental triangulation, reduced operative working place, blocked visualization, instrumental crowding, and clashing [2].
A robotic platform was thus developed to help restore intra-abdominal triangulation while maintaining the maximum degree of freedom for precise maneuvers. Significant advances have been achieved in the field of robotic laparoendoscopic single-site (R-LESS) surgery since the first reported clinical series in 2009. The specially designed da Vinci single-port surgical robot system (SP) produced by the Intuitive Surgical was approved for clinical application by the United States Food & Drug Administration (FDA) in 2018. Its use in gynecological surgery was relatively recent compared to urology and general surgery; moreover, it has not yet been allowed to be exported to China. The endoscopic surgery system was developed by Beijing Surgerii Robotics Co., Ltd. and is a single-port surgical robot system independently developed in China.
To explore the effectiveness and the safety of the SHURUI endoscopic surgical system (SR-ENS-600) in gynecological diseases, a prospective, multicenter, single-arm clinical trial was carried out, and the results are reported below.

Materials and methods

Ethics statement and registration

This prospective, multicenter, single-arm clinical trial was conducted between January 17 and May 26, 2023. The institutional review board approved the study (K2022566) on September 8, 2022. The trial was registered at https://​www.​chictr.​org.​cn/​ (registration number: ChiCTR2300075431).

Patient population

Eligible patients with ovary cysts, myoma, cervical epithelial neoplasms, and endometrial carcinoma were recruited according to the following inclusion and exclusion criteria.
Inclusion Criteria
(1)
Females, aged between 18 and 75 years;
 
(2)
Patients with indications for elective gynecological surgery;
 
(3)
Patients who can tolerate surgery and anesthesia;
 
(4)
Patients without a history of complicated gynecological diseases and serious diseases;
 
(5)
Patients with no history of mental illness;
 
(6)
Patients with body mass index (BMI) ≤ 32 kg/m2;
 
(7)
Patients who agreed to the study, had good compliance, and voluntarily signed an informed consent form.
 
Exclusion criteria.
(1)
Patients with a history of more than two pelvic or abdominal surgeries or severe pelvic adhesions;
 
(2)
Patients with ruptured ectopic pregnancy, massive bleeding and unstable vital signs;
 
(3)
Patients with acute genital tract or systemic infection;
 
(4)
Patients with coagulation dysfunction or treated with long-term anticoagulant drugs;
 
(5)
Patients with severe heart and lung function diseases, liver and kidney insufficiency, or unable to tolerate anesthesia;
 
(6)
Patients with abdominal wall or diaphragmatic hernia, abnormal umbilical development, or history of umbilical surgery;
 
(7)
Patients who refuse to undergo laparoscopic surgery;
 
(8)
Patients who had participated in clinical trials of other drugs or devices within 3 months before surgery.
 

Study design

The sample size was calculated before the study commenced. According to clinical experience, P0 for this experimental device was set as 90%, and the expected surgical success rate PT was 99%. The power (1-β) was 80% and α was 0.025 (unilateral); 56 women would be needed to observe a significant difference in primary outcome according to calculation. The sample size was increased by 10% to account for potential dropout; thus, the final sample size was 63 women.
Written informed consent was obtained from all participants before their enrollment, and all practices were performed according to related regulations. Medical history, vital signs, general and gynecological physical examination, laboratory tests, and imaging examination were performed. All surgical procedures were performed by one gynecologist who was both experienced in traditional single-port surgery and robotic endoscopic multiport surgery system in each center. Simply put, the SHURUI endoscopic surgical system consists of a surgeon console and a surgical trolley, a 3D endoscope and several surgical instruments. A trans-umbilical incision of approximately 25 mm (as measured by a sterile ruler) was made through an open Hasson approach before a four-channel trocar was inserted (see Fig. 1A–C). The patient was placed in a steep Trendelenburg position, and the surgical trolley was subsequently parked beside the patient-side bed and the robot was docked on the patient’s left side. The surgical instruments and the 3D endoscope were then loaded on the surgical trolley and steered through the access channels in the trocar to enter the patient’s abdomen (see Fig. 1D). The surgeon then accomplished the surgery by maneuvering the haptic devices on the surgeon console to tele-operate the surgical instruments or the 3D endoscope under the guidance of stereo visual feedback of the surgical site (see Fig. 1E). During the teleoperation, the surgical instruments were controlled in an intuitive manner to reduce the mental work of the manipulation. The incision was closed with purse-string suture of both peritoneum and rectus sheath, and then continuous suture of subcutaneous tissue and intradermic suture (see Fig. 1F).
Demographic information, perioperative parameters, and complications—including adjacent organ injury, fever, surgical field infection, pelvic hematoma, thromboembolism, and postoperative intestinal obstruction—as well as scar healing and operator satisfaction scores were recorded.

Outcomes

The primary outcome was the surgical success rate. Surgical success is defined as no conversion to open surgery or other conventional endoscopic surgeries for any reason. Surgical success rate = number of subjects operated on as planned/total number of included subjects × 100%.
Secondary evaluation parameters included operation time, bleeding volume, hospitalization stay, anal exhaust time, incision healing grade (evaluated at discharge day and 30 ± 4 days after surgery) (see Table 1), incision scar satisfaction score evaluated with The Scar Cosmesis Assessment and Rating (SCAR) scale [3] (30 ± 4 days after surgery) (see Table 2), subjective postoperative rehabilitation score (discharge day, 30 ± 4 days after surgery), surgeon satisfaction score and safety evaluation parameters: laboratory examination, vital signs, reoperation rate, perioperative complications, and AE/SAE status. Scar-related parameters were measured by trained nurses, and the subjective postoperative rehabilitation score, surgeon satisfaction score, and safety evaluation were self-reported using a standard questionnaire.
Table 1
Incision healing grade
Incision healing grade
Grade parameter
A
Excellent healing, no redness and infection at the incision, linear healing
B
Healing is poor, the incision has an inflammatory response but is not suppurated, and the incision is partially linear
C
The incision skin is large and suppurated, requiring incision drainage
Table 2
The Scar Cosmesis Assessment and Rating (SCAR) scale
Parameter
Descriptor
Score
Clinician questions
0
Scar spread
None/near invisible
0
Pencil-thin line
1
Mild spread, noticeable on close inspection
2
Moderate spread, obvious scarring
3
Severe spread
4
Erythema
None
0
Light pink, some telangiectasias may be present
1
Red, many telangiectasias may be present
2
Deep red or purple
3
Dyspigmentation
Absent
0
Present
1
Suture marks
Absent
0
Present
1
Hypertrophy/atrophy
None
0
Mild: palpable, barely visible hypertrophy or atrophy
1
Moderate: clearly visible hypertrophy or atrophy
2
Severe: marked hypertrophy or atrophy or keloid formation
3
Overall impression
Desirable scar
0
Undesirable scar
1
Patient questions
Itch
No
0
Yes
1
Pain
No
0
Yes
1
Subjective postoperative rehabilitation scores were evaluated with the following questions scaled between 0 and 10 according to severity: 1. breathing well, 2. good appetite, 3. feeling relaxed, 4. sleeping well, 5. able to use the toilet and take care of personal hygiene by herself, 6. able to communicate with family or friends, 7. able to communicate with doctors and nurses and get their support and encouragement, 8. able to engage in work or family activities, 9. feeling comfortable and self-controlling, 10. feeling healthy overall, 11. moderate pain, 12. intense pain, 13. nausea and vomiting, 14. feeling nervous and anxious, and 15. feeling sad and depressed within the past 24 h. The postoperative rehabilitation score consists of two parts: the higher the comfort score (questions 1–10), the better, with a maximum score of 100; the lower the scores for pain and anxiety (questions 11–15), the better, 0 means no above symptoms present, with a total score of 50.
The Surgeon Satisfaction Questionnaire included two parts: system performance (questions 1–12) and operator comfort-related scores (questions 13–20), with 1–5 points from most uncomfortable/difficult to most comfortable/easy for each question and a full score of 100 points.

Statistical analysis

Statistical analysis was carried out using SPSS Statistics for Windows (version 25.0; SPSS Inc., Chicago, IL, USA), and the sample size calculation adopted PASS 13 (NCSS, LLC). The curative effect evaluation included the analysis of the PPS and FAS sets, and the conclusion was mainly based on the FAS set. The safety evaluation adopted SS set analysis. The descriptive statistical analysis of quantitative indicators was reported as the number of cases, mean, standard deviation, median, lower quartile (Q1), and upper quartile (Q3), and the statistical description of categorical variables included frequency and percentage. A one-sample t test was used to compare the difference between the results from the present study and the literature using the da Vinci SP system performing same surgery as in our study. The results were deemed significantly different from each other when the two-sided P value was less than 0.05.

Quality control

Before the study began, relevant training was provided to all researchers participating in the study to ensure that they fully understood the process. All investigators authorized to perform surgeries in participating research centers had relevant surgical experience to reduce the bias caused by the researcher’s skills on the trial results. All independent evaluators received unified training to reduce the impact of differences in evaluators' evaluations on the trial results. When the clinical study was completed, data storage and organization were ensured. The data administrator checked and confirmed the data through the data clarification form to avoid recording errors.

Results

Participant disposition and demographics

A total of 64 cases were enrolled in 6 hospitals in different districts of China, with 11 cases enrolled in center 1, 16 cases in center 2, 10 cases in center 3, 7 cases in center 4, 10 cases in center 5, and 12 cases in center 6. One subject voluntarily withdrew from the trial due to psychological stress and preoperative anxiety and thus was excluded from all datasets (FAS, PPS, and SS), with 63 cases included in the final datasets (Fig. 2).
Sixty (95.2%) patients were of Han nationality, and 37 (58.7%) patients had a previous history of surgery. Nineteen (30.2%) patients were postmenopausal for 7.00(4.00, 13.00) years. Fourteen (31.8%) patients had a history of dysmenorrhea. Gestation and parturition for all patients were 3.00 (1.00, 4.00) and 1.00 (0.00, 2.00) respectively. The patient withdrawn was 33 years old and her BMI was 20.3 kg/m2, she suffered from dysmenorrhea and had one abortion before. Nine (14.3%) patients received cystectomy/adnexectomy, 10 (15.9%) underwent myomectomy, 35 (55.6%) received hysterectomy, and 9 (14.3%) patients underwent staging surgery for endometrial carcinoma. Other demographic information of the patients compared with the literature [4, 5] is shown in Table 3 and Fig. 3.
Table 3
Demographic information of the patients compared to the literature [4, 5]
Variablesa
SRC
SRM
SRH
SRES
All
 
S (n = 9)
L4(n = 108)
p value
S (n = 10)
L4 (n = 56)
p value
S (n = 35)
L4 (n = 35)
p value
S (n = 9)
S (n = 63)
L5 (n = 31)
p value
Age (year)
30.56 ± 7.25
31.79 ± 10.31
0.623
39.50 ± 7.52
36.98 ± 7.14
0.317
51.26 ± 7.27
49.71 ± 7.86
0.217
51.56 ± 10.32
46.48 ± 10.91
47.7 ± 12.8
0.377
BMI (kg/m2)
23.20 ± 4.60
22.86 ± 4.39
0.830
22.39 ± 3.37
23.09 ± 4.19
0.527
23.24 ± 2.50
24.40 ± 3.744
0.010b
25.49 ± 2.37
23.42 ± 3.06
22.7 ± 3.1
0.065
aData are presented as the mean ± standard deviation, median (Q1, Q3), or number (percentage). S stands for results of present study, L stands for results from the literature, W stands for results of withdrawn patient. bstands for results showing differences between the present and literature

Surgical and safety profiles of patients

The surgical success rate was 100.0% (95% CI: 94.3–100.0%).
Perioperative indicators of the patients compared to the literature [4, 5] are shown in Table 4 and Fig. 4.
Table 4
Perioperative indicators of the patients compared to the literature [4, 5]
Variables
SRC
SRM
SRH
SRES
All
 
S(n = 9)
L4 (n = 108)
p value
S (n = 10)
L4 (n = 56)
p value
S (n = 35)
L4 (n = 35)
p value
S (n = 9)
S (n = 63)
L5 (n = 31)
p value
Operation time (min)
84.33 ± 31.95
81.90 ± 45.07
0.825
145.60 ± 47.38
134.55 ± 63.39
0.480
167.66 ± 79.65
114.71 ± 44.20
0.000a
200.00 ± 69.33
156.87 ± 75.34
126.3 ± 61.6
0.002a
Bleeding volume (ml)
12.00 ± 9.58
  
51.50 ± 49.44
  
80.86 ± 124.77
  
63.33 ± 38.08
63.86 ± 98.33
93.9 ± 77.0
0.018a
Hemoglobin change (g/l)
6.00 ± 4.03
1.59 ± 1.084
0.011a
4.60 ± 7.28
1.75 ± 1.20
0.247
−1.21 ± 13.18
1.53 ± 1.04
0.227
-3.28 ± 19.52
0.45 ± 12.85
2.4 ± 0.8
0.232
Anal exhaust time (hr)
21.86 ± 9.96
  
23.76 ± 6.35
  
35.01 ± 14.81
  
32.49 ± 16.24
30.99 ± 14.25
34.1 ± 15.5
0.088
Hospitalization stay(day)
2.00 ± 0.71
4.55 ± 1.86
0.000a
2.60 ± 1.58
4.57 ± 1.12
0.003a
3.97 ± 1.27
4.54 ± 1.01
0.012a
5.11 ± 1.45
3.63 ± 1.59
4.6 ± 0.7
0.000a
SRC stands for single-port robotic cystectomy/adnexectomy, SRM stands for single-port robotic myomectomy, SRH stands for single-port robotic hysterectomy, SRES stands for single-port robotic endometrial carcinoma staging. S represents the results of the present study, and L represents the results from the literature. a indicates results showing differences between the present study and the literature
All patients experienced grade A incision healing on the discharge day and D30 ± 4 after surgery. The incision scar satisfaction score evaluated with the Scar Cosmesis Assessment and Rating (SCAR) scale was no more than 1 point for all patients, with over 55.6% of patients scoring 0 for all items except for scar spread after myomectomy.
On the day of discharge and D30 ± 4 after surgery, the average postoperative rehabilitation assessment scores were 88.71 ± 10.64 points and 94.87 ± 8.16 points, respectively, and the pain and anxiety scores were 3.38 ± 3.18 points and 0.87 ± 2.88 points, respectively.
Regarding the surgeon satisfaction score, the average score related to system performance was 51.35 ± 6.73 points, the average score related to operator comfort was 35.29 ± 4.03 points, and the average total score was 86.63 ± 10.59 points. Approximately 85% of surgeons gave 3 ~ 5 points for each item.
No patients suffered from intraoperative or postoperative complications, and no patients experienced reoperation on D30 ± 4. There were 79 mild adverse events, 9 moderate adverse events (including 3 cases of nausea and vomiting, 2 cases of anemia, 2 cases of acute respiratory infection, 1 case of vaginal bleeding, and 1 case of insomnia) and 1 severe adverse event (vaginal stump bleeding after discharge needing resuturing) reported, but none were trial-related. Compared with rigid links, the continuum had softer gripping power in pulling out fibroids.

Discussion

Minimally invasive surgery has revolutionized the treatment of gynecologic diseases over the last 30 years. The advantages of laparoscopic surgery over abdominal surgery have been well established [6, 7]. However, there are several limitations to traditional laparoscopy, including counterintuitive hand movements, accentuated tremors and ergonomically challenging positions, which may result in fatigue and even injuries to the surgeon with lengthy practice. With the advent of the da Vinci robotic system (DRS), many surgeries are now being performed with robotic procedures. The rate of robotic surgeries performed in 2012 was significantly increased by 50% compared to 27.9% performed in 2007 [8].
In the meantime, laparoendoscopic single-site surgery has shown great attraction owing to its cosmetic benefits; however, it has surgical limitations in angulation and device manipulation [9]. Robotic single-port surgery, on the other hand, combines the advantages of robotics with the esthetic result of a single incision. In the literature, compared with LESS hysterectomy, R-LESS had a longer operating time but had less blood loss, a decreased length of hospital stays, and a favorable learning curve [10, 11]. In the SHURUI endoscopic surgical system, both the robotic instrument and the 3D endoscopy are composed of a so-called dual continuum mechanism. Unlike conventional instruments that are composed of joints and rigid links, the continuum mechanism achieves motion via continuous deformation of the elastic structure and is designed for enhanced payload capability. As shown in Fig. 5, the dual continuum mechanism consists of proximal segments, distal segments and guiding cannula. The bending of the proximal segment is coupled to that of the distal segment via the translations (pulling and pushing) of all the structural backbones made of super-elastic Ni–Ti alloy (Fig. 5). The failure of several backbones does not affect the functionality of the mechanism. And the backbones could collaboratively balance the external load applied on the segment. Therefore, the dual continuum mechanism in the SHURUI endoscopic surgical system can help realize a payload-enhanced multi-DoF surgical instrument with a tight bending wrist for intra-abdomen dexterity. The advantages of such structure include flexible arms, compact size, high accessibility with sufficient triangulation and motion degree of freedom [12].
In the present trial, the demographic information of patients was similar to that of the other two studies using the da Vinci SP [4, 5], except that the BMI of patients who received hysterectomy in our study was lower than that of Kim’s study (23.24 ± 2.50 vs. 24.40 ± 3.74, p = 0.010) [4].
In regard to perioperative parameters, the operating time showed no difference for cystectomy/adnexectomy and myomectomy but was longer for hysterectomy in our study (167.66 ± 79.65 vs. 114.71 ± 44.20 min, p < 0.001) [4]. The possible reason for this was because one center included patients with multiple-myoma, especially one nullipara patient with both adenomyosis and myoma whose uterus was irregularly enlarged, parametrium was severely shortened and lacked mobility, with severe uterosacral ligament adhesion. The whole surgery took 472.00 min, and the bleeding volume was 700 ml. This might account for the prolonged mean operating time for hysterectomy. This also led to a longer mean operating time in all surgeries compared with that in Shin’s study (156.87 ± 75.34 vs. 126.3 ± 61.6 min, p = 0.002) [5]. After cystectomy/adnexectomy, hemoglobin was reduced more than that reported by Kim [4]. Almost all surgeons who started with cystectomy/adnexectomy for practice might account for this result, and we noticed that as the trial went on, there was less bleeding with cystectomy/adnexectomy. We also found that there was less bleeding in the whole group than in the Shin group (63.86 ± 98.33 vs. 93.9 ± 77.0 ml, p = 0.018) [5], but the hemoglobin change showed no difference. The anal exhaust time for all surgeries including gynecologic tumors was 30.99 ± 14.25 h in the present trial compared to 34.1 ± 15.5 h reported by Shin, but there was no significant difference (p = 0.088) [5]. The most notable difference between our results and the literature was a much shorter hospital stay for all surgery types (p < 0.015), which was an important parameter for postoperative rehabilitation [4, 5].
There was no conversion to multiport laparoscopy or laparotomy or perioperative complications in our and other studies, except that Kwak reported 1 case (1.3%) of superficial bowel laceration due to severe adhesion from prior surgery during endometrial cancer staging and 1 case (1.3%) of wound complication after discharge [4, 5, 1315].
The SHURUI endoscopic surgical system was shown to be efficacious and safe in cystectomy/adnexectomy, myomectomy, hysterectomy, or endometrial carcinoma staging under the conditions of this study.
Since this study was conducted under careful quality control, the results suggest that ovary cysts, myoma, cervical epithelial neoplasms, and endometrial carcinoma may be appropriate indication for the SHURUI endoscopic surgical system. Randomized controlled trials are needed to determine whether the SHURUI endoscopic surgical system is effective for a wider range of gynecological indications.
The present trial has several notable features. To our knowledge, this is the first report of gynecological results using the SHURUI endoscopic surgical system and the first study to include both benign and malignant gynecologic disorders. To explore its benefits as a minimally invasive surgery, we evaluated the results of incision healing and scar cosmesis assessment, which revealed that all patients experienced grade A healing, and over 65% of patients showed no scar on D30 ± 4. With 100% of patients scoring no more than 1 point, when concealed in the umbilicus, the incision successfully satisfied patients’ demand for cosmesis. Furthermore, an ideal single-port robotic surgery system should allow good visualization, prevent hand tremors, provide three-dimensional (3D) high-definition images, and allow the surgeon to perform lengthy surgery in a more comfortable position. Thus, our study evaluated the surgeon’s satisfaction score and found that the average scores related to system performance and operator comfort were 51.35 ± 6.73 out of 60 points and 35.29 ± 4.03 points out of 40 points, respectively, with a total score of 86.63 ± 10.59 out of 100 points, which showed good experience with the surgeons. The prospective, multicenter nature of this study avoided the possible limitations of a single research institution and restricted population, making its results more generalizable. Careful sample size calculation and strict quality control ensured the reliability of the results.
The greatest limitation of this study was its single-arm design, which hinders direct comparison with traditional and multiport robotic laparoscopic surgery. Further randomized controlled trials are needed to confirm the present results. Another limitation was that one-sample t test was not appropriate to calculate variables such as percentage of dysmenorrhea or numbers of gestation and parturition, and thus we could not identify if this demographic information was comparable to literature.

Conclusions

The clinical outcomes of single-port laparoscopic surgery using the SHURUI endoscopic surgical system for gynecological disorders, including cystectomy/adnexectomy, myomectomy, hysterectomy, and endometrial carcinoma staging, were comparable to those performed with the da Vinci robotic single-port system. There was no need for conversion to multiport or open surgery, nor was there any operative complications. It achieved good cosmetic benefits and was surgeon-friendly.
Therefore, this study suggests that the SHURUI endoscopic surgical system can be proposed as a relatively safe and feasible surgical method for patients who desire to undergo single-port surgery.

Acknowledgements

The authors acknowledge participating principal investigators and clinical centers in the study. Furthermore, the authors thank the participants and clinical center staff for their contributions to this study.

Declarations

Conflict of interest

The trial received fund from single-port laparoendoscopic surgical system industry-university-research cooperation project sponsored by Beijing Surgerii Robotics Co., Ltd.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Urologie

Kombi-Abonnement

Mit e.Med Urologie erhalten Sie Zugang zu den urologischen CME-Fortbildungen und Premium-Inhalten der urologischen Fachzeitschriften.

Weitere Produktempfehlungen anzeigen
Literatur
1.
Zurück zum Zitat Goebel K, Goldberg JM (2014) Women’s preference of cosmetic results after gynecologic surgery. J Minim Invasive Gynecol 21(1):64–67CrossRefPubMed Goebel K, Goldberg JM (2014) Women’s preference of cosmetic results after gynecologic surgery. J Minim Invasive Gynecol 21(1):64–67CrossRefPubMed
2.
Zurück zum Zitat Jung YW, Lee M, Yim GW, Lee SH, Paek JH, Kwon HY, Nam EJ, Kim SW, Kim YT (2011) A randomized prospective study of single-port and four-port approaches for hysterectomy in terms of postoperative pain. Surg Endosc 25(8):2462–2469CrossRefPubMed Jung YW, Lee M, Yim GW, Lee SH, Paek JH, Kwon HY, Nam EJ, Kim SW, Kim YT (2011) A randomized prospective study of single-port and four-port approaches for hysterectomy in terms of postoperative pain. Surg Endosc 25(8):2462–2469CrossRefPubMed
3.
Zurück zum Zitat Kantor J (2017) Reliability and photographic equivalency of the scar cosmesis assessment and rating (scar) scale, an outcome measure for postoperative scars. JAMA Dermatol 153(1):55–60CrossRefPubMed Kantor J (2017) Reliability and photographic equivalency of the scar cosmesis assessment and rating (scar) scale, an outcome measure for postoperative scars. JAMA Dermatol 153(1):55–60CrossRefPubMed
4.
Zurück zum Zitat Kim JM, Lee SM, Seol A, Song JY, Ryu KJ, Lee S, Park HT, Cho HW, Min KJ, Hong JH, Lee JK, Lee NW (2023) Comparison of surgical outcomes between single-port laparoscopic surgery and da vinci single-port robotic surgery. J Pers Med 13(2):205CrossRefPubMedPubMedCentral Kim JM, Lee SM, Seol A, Song JY, Ryu KJ, Lee S, Park HT, Cho HW, Min KJ, Hong JH, Lee JK, Lee NW (2023) Comparison of surgical outcomes between single-port laparoscopic surgery and da vinci single-port robotic surgery. J Pers Med 13(2):205CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Shin HJ, Yoo HK, Lee JH, Lee SR, Jeong K, Moon HS (2020) Robotic single-port surgery using the da Vinci SP® surgical system for benign gynecologic disease: a preliminary report. Taiwan J Obstet Gynecol 59(2):243–247CrossRefPubMed Shin HJ, Yoo HK, Lee JH, Lee SR, Jeong K, Moon HS (2020) Robotic single-port surgery using the da Vinci SP® surgical system for benign gynecologic disease: a preliminary report. Taiwan J Obstet Gynecol 59(2):243–247CrossRefPubMed
6.
Zurück zum Zitat Paraiso MF, Walters MD, Rackley RR, Melek S, Hugney C (2005) Laparoscopic and abdominal sacral colpopexies: a comparative cohort study. Am J Obstet Gynecol 192(5):1752–1758CrossRefPubMed Paraiso MF, Walters MD, Rackley RR, Melek S, Hugney C (2005) Laparoscopic and abdominal sacral colpopexies: a comparative cohort study. Am J Obstet Gynecol 192(5):1752–1758CrossRefPubMed
7.
Zurück zum Zitat Mais V, Ajossa S, Guerriero S, Mascia M, Solla E, Melis GB (1996) Laparoscopic versus abdominal myomectomy: a prospective, randomized trial to evaluate benefits in early outcome. Am J Obstet Gynecol 174(2):654–658CrossRefPubMed Mais V, Ajossa S, Guerriero S, Mascia M, Solla E, Melis GB (1996) Laparoscopic versus abdominal myomectomy: a prospective, randomized trial to evaluate benefits in early outcome. Am J Obstet Gynecol 174(2):654–658CrossRefPubMed
8.
Zurück zum Zitat Conrad LB, Ramirez PT, Burke W, Naumann RW, Ring KL, Munsell MF, Frumovitz M (2015) Role of minimally invasive surgery in gynecologic oncology: an updated survey of members of the society of gynecologic oncology. Int J Gynecol Cancer 25(6):1121–1127CrossRefPubMedPubMedCentral Conrad LB, Ramirez PT, Burke W, Naumann RW, Ring KL, Munsell MF, Frumovitz M (2015) Role of minimally invasive surgery in gynecologic oncology: an updated survey of members of the society of gynecologic oncology. Int J Gynecol Cancer 25(6):1121–1127CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Lopez S, Mulla ZD, Hernandez L, Garza DM, Payne TN, Farnam RW (2016) A comparison of outcomes between robotic-assisted, single-site laparoscopy versus laparoendoscopic single site for benign hysterectomy. J Minim Invasive Gynecol 23(1):84–88CrossRefPubMed Lopez S, Mulla ZD, Hernandez L, Garza DM, Payne TN, Farnam RW (2016) A comparison of outcomes between robotic-assisted, single-site laparoscopy versus laparoendoscopic single site for benign hysterectomy. J Minim Invasive Gynecol 23(1):84–88CrossRefPubMed
11.
Zurück zum Zitat Paek J, Lee JD, Kong TW, Chang SJ, Ryu HS (2016) Robotic single-site versus laparoendoscopic single-site hysterectomy: a propensity score matching study. Surg Endosc 30(3):1043–1050CrossRefPubMed Paek J, Lee JD, Kong TW, Chang SJ, Ryu HS (2016) Robotic single-site versus laparoendoscopic single-site hysterectomy: a propensity score matching study. Surg Endosc 30(3):1043–1050CrossRefPubMed
12.
Zurück zum Zitat Zhang X, Li WB, Chiu PWY, Li Z (2020) A novel flexible robotic endoscope with constrained tendon-driven continuum mechanism. IEEE Robot Autom Lett 5:1366–1372CrossRef Zhang X, Li WB, Chiu PWY, Li Z (2020) A novel flexible robotic endoscope with constrained tendon-driven continuum mechanism. IEEE Robot Autom Lett 5:1366–1372CrossRef
13.
Zurück zum Zitat Lee JH, Yoo HK, Park SY, Moon HS (2022) Robotic single-port myomectomy using the da Vinci SP surgical system: a pilot study. J Obstet Gynaecol Res 48(1):200–206CrossRefPubMed Lee JH, Yoo HK, Park SY, Moon HS (2022) Robotic single-port myomectomy using the da Vinci SP surgical system: a pilot study. J Obstet Gynaecol Res 48(1):200–206CrossRefPubMed
14.
Zurück zum Zitat Misal M, Magtibay PM, Yi J (2021) Robotic LESS and reduced-port hysterectomy using the da vinci sp surgical system: a single-institution case series. J Minim Invasive Gynecol 28(5):1095–1100CrossRefPubMed Misal M, Magtibay PM, Yi J (2021) Robotic LESS and reduced-port hysterectomy using the da vinci sp surgical system: a single-institution case series. J Minim Invasive Gynecol 28(5):1095–1100CrossRefPubMed
15.
Zurück zum Zitat Kwak YH, Lee H, Seon K, Lee YJ, Lee YJ, Kim SW (2022) Da vinci SP single-port robotic surgery in gynecologic tumors: single surgeon’s initial experience with 100 cases. Yonsei Med J 63(2):179–186CrossRefPubMedPubMedCentral Kwak YH, Lee H, Seon K, Lee YJ, Lee YJ, Kim SW (2022) Da vinci SP single-port robotic surgery in gynecologic tumors: single surgeon’s initial experience with 100 cases. Yonsei Med J 63(2):179–186CrossRefPubMedPubMedCentral
Metadaten
Titel
Effectiveness and safety of SR-ENS-600endoscopic surgical system in benign and malignant gynecological diseases: a prospective, multicenter, clinical trial with 63 cases
verfasst von
Ren Chang
Duan Ping
Shen Yang
Wang Yongjun
Zhang Wei
Zheng Ying
Li Xingming
Zhang Kexin
Sun Dawei
Publikationsdatum
01.12.2024
Verlag
Springer London
Erschienen in
Journal of Robotic Surgery / Ausgabe 1/2024
Print ISSN: 1863-2483
Elektronische ISSN: 1863-2491
DOI
https://doi.org/10.1007/s11701-024-01871-4

Weitere Artikel der Ausgabe 1/2024

Journal of Robotic Surgery 1/2024 Zur Ausgabe

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.